Cascade TEK Tek TFO-1 Instruction Manual

Cascade forced air and vacuum ovens user's manual

Advertisement

Cascade TEK
D ivision of Cascade Technical Sciences, Inc.
TH
⋅ ℡
1800 NE 25
Avenue
Hillsboro, OR 97124
(503) 648-1818
FAX: (503) 648-1798

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Tek TFO-1 and is the answer not in the manual?

Questions and answers

Summary of Contents for Cascade TEK Tek TFO-1

  • Page 1 Cascade TEK D ivision of Cascade Technical Sciences, Inc. ⋅ ⋅ ℡ ⋅ 1800 NE 25 Avenue Hillsboro, OR 97124 (503) 648-1818 FAX: (503) 648-1798...
  • Page 2 This instruction manual applies to Models TFO-1, TFO-3, TFO-5, TFO-10, TFO-28, TVO-1, TVO-2, and TVO-5 SECTION 1.0 SECTION 2.0 SECTION 3.0 SECTION 4.0 SECTION 5.0 SECTION 6.0 SECTION 7.0 SECTION 8.0 SECTION 9.0 SECTION 10.0 TABLE OF CONTENTS RECEIVING AND INSPECTION WARNING POWER SUPPLY INSTALLATION...
  • Page 3: Receiving And Inspection

    This oven is not designed for use in Class I, II, or III locations as described in the National Electrical Code. It is not intended, nor shall it be used as a patient-connected device as described in Article 517 of the National Electrical Code.
  • Page 4 Vacuum Inlet Valve: SMALL KNOB located to the left of the vacuum gauge. This knob allows is for venting of the oven or use with an inert gas backfill. The connection for this valve is at the back of the oven and is the smaller ¼”...
  • Page 5: Preventive Maintenance

    Using the exhaust damper: If you want moderate ventilation in the oven, open the vent on top. The slide damper controls volume of ventilation. Maximum flow will slightly increase heat up and recovery time.
  • Page 6: Troubleshooting

    A heater element has failed. The Controller Calls For Heat Constantly: The controller's Thermocouple temperature sensor has failed. Oven temperature is now controlled by the overtemperature safety. A solid state relay has failed closed. Temperature is now controlled by the overtemperature safety.
  • Page 7: Parts List

    9.0 PARTS LIST FORCED AIR OVENS PART NO. DESCRIPTION TFO-1 1800516 Power Cord 9570762 Element Assy. 9600565 Blower Motor Assy. 100056 Blower Wheel 6 x 2 4880512 Blower Motor 600002 Gasket 10 ft 5120689 Shelf 200129 Adjustable Feet TFO-5 101990 Power Cord 9570665 Element Assy.
  • Page 8: Limited Warranty

    VACUUM OVENS TVO-1 9570739 Element Assy., top / bottom 5620506 Shelf - Top 2 5620519 Shelf - Bottom 1 TVO-2 9570742 Element Assy. 5620518 Shelf - Top 2 5620520 Shelf - Bottom 1 TVO-5 9570765 Element Assy., top / bottom 5620523 Shelf 5220580...
  • Page 9: Wiring Diagram

    North America Cascade TEK agrees to correct for the original user of this product, either by repair or at our election, by replacement, any defect in material or workmanship which develop within twelve (12) months after delivery of this product to the original user. In the event of replacement, replacement unit will be warranted for the original twelve (12) month period or ninety (90) days, whichever is longer.
  • Page 10 TFO-1, TFO-3...
  • Page 11 WIRING DIAGRAM TFO-5...
  • Page 12 WIRING DIAGRAM TFO-10...
  • Page 13 WIRING DIAGRAM TFO-28...
  • Page 14 WIRING DIAGRAM TVO-1, TVO-2, TVO-5...

Table of Contents