Settings
50
Menu item
Measurement /
Measuring cell /
Type
Measurement /
Measuring cell /
Cell const. man
Measurement /
Measuring cell /
Cell const. LR325/
01
Measurement /
Temp. comp. (TC)
/Method
Measurement /
Temp. comp. (TC)
/ Linear coeff.
Possible
Description
setting
Measuring cell used
Cal
Measuring cells whose cell constant
is determined by calibration in the
KCL control standard solution.
Calibration ranges:
0.450 ... 0.500 cm
0.800 ... 0.880 cm
The currently valid cell constant is
displayed in the status line.
LR325/01
LR 325/01 measuring cell,
nominal cell constant
0.100 cm
The cell constant can be adjusted in
the range 0.090 ... 0.110 cm
LR325/001
LR 325/001 measuring cell,
nominal cell constant
0.010 cm
The cell constant is permanently set.
man
Any measuring cells with freely
adjustable cell constants in the range
0.250 ... 25.000 cm
0.250 ...
Display and setting option of the cell
0.475
constant of any measuring cells
-1
25.000 cm
(man).
0.090 ...
Display and setting option of the cell
0.100 ...
constant of the LR 325/01 measuring
-1
0.110 cm
cell.
nLF
Procedure for temperature compen-
Lin
sation (see section 8.1.7 T
Off
TURE COMPENSATION
This setting is only available for the
measured parameters, ϰ and ρ.
0.000 ...
Coefficient of the linear temperature
2.000 ...
compensation.
10.000 %/K
This setting is only available when the
linear temperature compensation is
set.
pH/Cond 3320
-1
and
-1
-1
.
-1
.
-1
.
-1
.
EMPERA
, page 51).
ba77054d01
-
02/2014