Electrical connection
18
Current output configuration version
4
Current output, active, 4 to 20 mA
A
Plug (measuring device)
B
Socket (customer side)
L+
Supply voltage +
L-
Supply voltage -
The current flows from the output to L-. The maximum load may not exceed 500 Ω. A
bigger load distorts the output signal.
Voltage output configuration version
5
Voltage output, active, 2 to 10 V
A
Plug (measuring device)
B
Socket (customer side)
L+
Supply voltage +
L-
Supply voltage -
The voltage from the output applies to L-. The load must be at least 600 Ω. The output is
protected against short-circuiting and reverse polarity.
Digital input configuration version
• 15 V (switch-on threshold)
• 5 V (switch-off threshold)
6
Digital input
A
Plug (measuring device)
B
Socket (customer side)
L+
Supply voltage +
L-
Supply voltage -
A
B
L+
1
2
4
A
A
3
L-
A
B
L+
1
2
4
V
V
3
L-
A
B
L+
1
2
4
=
=
3
L-
Picomag IO-Link
A0046581
A0046582
A0033008
Endress+Hauser
Need help?
Do you have a question about the Picomag IO-Link and is the answer not in the manual?
Questions and answers