Page 1
Model QS-550 Free-Standing Tubestand Service Manual Manual Part No. DC30-009 Revision B...
Page 2
This manual is copyrighted and all rights are reserved. No portion of this document may be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine readable form without prior consent in writing from Quantum Medical Imaging, LLC.
Page 3
Revision History REVISION DATE TYPE OF MODIFICATION 12/7/00 Initial Release. 2/28/01 Minor modification made to installation of tube stand on floor track procedure. Page Page Page Number Number Number i - x 1 - 47 Model QS-550 - Service Manual...
Page 4
Revision History THIS PAGE INTENTIONALLY LEFT BLANK Model QS-550 - Service Manual...
Table of Contents Table of Contents Chapter 1, Specifications Chapter 2, Assembly and Installation Chapter 3, Theory of Operation Chapter 4, Service Maintenance Chapter 5, Diagrams Model QS-550 - Service Manual...
Page 6
Table of Contents THIS PAGE INTENTIONALLY LEFT BLANK Model QS-550 - Service Manual...
±5 inches (127 mm) from centerline of table • Transverse Travel Detent: Center position • X-ray Tube Angulation: 240° (±120°) • X-ray Tube Angulation Detents: 0°, 90°(CW), 90°(CCW) • Column Rotation: 180° (±90°) • Column Rotation Detents: 0°, 90°(L), 90°(R) Model QS-550 - Service Manual...
Page 10
-30.5 to +3048 meters relative to sea level COMPATIBILITY STATEMENT The Quantum Medical Imaging, LLC. tubestand is compatible with all Quantum Medi- cal Imaging, LLC. manufactured tables, wall stands, and high-voltage x-ray genera- tors, and with other manufacturer’s equipment having equivalent means for indication of SID and perpendicularity.
Page 11
Chapter 1 Specifications Figure 1. Model QS-550 Tubestand Configuration Diagram Model QS-550 - Service Manual...
Page 12
Chapter 1 Specifications THIS PAGE INTENTIONALLY LEFT BLANK Model QS-550 - Service Manual...
Chapter 2 Assembly & Installation OVERVIEW This chapter describes the steps required to assemble and install the Model QS-550 Tubestand, including making all required electrical connections, and performing all required alignments and calibrations. ROOM CONSTRUCTION A conventional radiographic room construction should be used in the area where the tubestand is to be installed.
Page 16
Do not dispose of packing NOTE material until packing list is matched with actual parts received. Should there be a shortage or damage, notify Quantum Medical Imaging, LLC.’s customer service Examine all car- department immediately. The manufacturer is relieved of any responsibility for...
Page 17
Chapter 2 Assembly & Installation MOUNTING HOLES (5X) Figure 2. Floor Track Installation RETAINING RING Figure 3. Installing the Column Retaining Ring Model QS-550 - Service Manual...
Page 18
4. Securel the Front and Rear Bearing Brackets to the Base Assembly (see Fig- ures 4 and 5). FRONT BEARING BRACKET Figure 4. Bearing Bracket Locations BEARING W/ ECCENTRIC BUSHING REAR BEARING Figure 5. Vertical Column Angle Adjustment Model QS-550 - Service Manual...
Page 19
1. Align mounting holes in the Transverse Arm with holes in Vertical Column Trolley and install four (4) 3/8-16 x 1-1/2" hex head bolts and lockwashers (see Figure 6). LEVELING SET SCREW (4X) 3/8-16 X 1-1/2" HEX HEAD BOLT (4X) Figure 6. Installing the Transverse Arm Model QS-550 - Service Manual...
Page 20
(4) 1/4-20 x 2" hex head bolts and lockwashers (see Figure 7). TUBE SUPPORT BRACKET TRUNNION ASSEMBLY COLLIMATOR 1/4-20 X 2" HEX MOUNTING HEAD BOLT PLATE AND LOCK- WASHER (4X) Figure 7. Mounting the Tube and Trunnion Assembly Model QS-550 - Service Manual...
Page 21
8. Mount Tube Support Bracket onto Trunnion using four (4) 1/4-20 hex head bolts. 9. Mount Handgrip Assembly on Tube Support Bracket (in front of the x-ray tube) using four (4) 1/4-20 x 1" hex head bolts (see Figure 9). Model QS-550 - Service Manual...
Page 22
13. Route Handgrip cable (from terminal block TB1) to D-connector on rear side of the Handgrip Assembly and connect. 14. Connect collimator cables as required (refer to collimator manual and to Figure 27 or 28 in Chapter 5, Diagrams). Model QS-550 - Service Manual...
Page 23
4. Route Handgrip cable (from terminal block TB1) to D-connector on rear side of the Handgrip Assembly and connect. 5. Connect collimator cables as required (refer to collimator manual and to Figure 27 or 28 in Chapter 5, Diagrams). Model QS-550 - Service Manual...
Page 24
Chapter 2 Assembly & Installation TUBE FORK 1/4-20 X 2" HEX HEAD BOLTS (4X) Figure 11. Mounting the Tube Fork Model QS-550 - Service Manual...
Page 25
(4) mounting blocks on the underside of the Transverse Arm Assembly. 3. Position the Transverse Arm Lower Cover as shown in Figure 14 and secure it to the Transverse Arm using four (4) 6-32 x 3/4" flat head screws. Model QS-550 - Service Manual...
Page 26
HEAD SOCKET SCREW (6X) Figure 13. Installing the Transverse Arm Inner Cover TRANSVERSE ARM OUTER COVER 6-32 X 3/4" FLAT HEAD SCREW (4X) TRANSVERSE ARM LOWER COVER Figure 14. Installing the Lower Transverse Arm Cover Model QS-550 - Service Manual...
Page 27
Vertical Column) and add trim weights as required (see Figure 15). ACCESS OPENING NOTE Be careful not to drop trim weights when adding through opening in Ver- tical Column. Figure 15. Trim Weight Access Opening (Rear Side of Vertical Column Model QS-550 - Service Manual...
Page 28
2. The space between electromagnetic lock and contacting part of the Floor Track should not exceed 20/1000" (ideally between 10/1000” and 20/1000”). Adjust electromagnetic lock brake height as required. 3. When electromagnetic lock height is correct, re-tighten two (2) 1/4-20 x 1” hex head bolts. Model QS-550 - Service Manual...
Page 29
Brake Assembly to terminal block TB1 as follows: •RED TB1-6 •BLACK TB1-2 •YELLOW TB1-1 •GREEN TB1-1 •BROWN TB1-5 •ORANGE TB1-4 7. Re-install left Base Access Panel on the Base Assembly. 8. Re-install terminal strip cover on back of tube stand. Model QS-550 - Service Manual...
Page 30
4. When switch height is correct, tighten the two (2) 10-32 x 1/4" socket head cap screws. 5. On SID switch cable (from terminal block TB1), connect red and black wires to NO and COM terminals, respectively, on Vertical SID Switch Assembly. Model QS-550 - Service Manual...
Page 31
Vertical Trolley by installing upper halves of Cable Clamps to lower halves using two (2) 8-32 x 1” socket head cap screws. 4. From the Terminal Strip Assembly, route the tubestand cables along with the +24 VDC/+15 VDC power supply cable, to the x-ray generator. Model QS-550 - Service Manual...
Page 32
2. Bring the Vertical Trolley down to the desired 40" SID mark on the Vertical each other. Column. 3. Mark the location for the Vertical SID Switch actuator. Model QS-550 - Service Manual...
Page 33
1. Set tube position for 40" (100 cm) tabletop SID. 2. Affix label on Vertical Column so that the 40" (100 cm) mark lines up with the arrow in the table SID window of the SID Indicator Bracket. Model QS-550 - Service Manual...
Page 34
3. Use collimator horizontal cross light to align the table. Rotate the x-ray tube to check the pitch adjustment as required. 4. Verify vertical movement and alignment with bucky stand or cas- sette holder using collimator light. Model QS-550 - Service Manual...
Chapter 3 Theory of Operation OVERVIEW The Model QS-550 Tubestand is a free-standing x-ray tube support device (i.e., does not require wall and/or ceiling mounting) that provides a wide range of tube positions through longitudinal, vertical, rotational, and transverse tubestand motions. The Handgrip Assembly is designed for operator convenience with ergonomically posi- tioned push button switches that enable tubestand movement when pressed.
Page 38
SID switch, which routes the +24 VDC to the Longitudinal Lock Coil L3. The longitudinal brake (magnetic lock) is activated, locking the x-ray tube in the current SID position. To re-activate the Auto-Stop function, both the Longitudinal Lock Release and Auto-Stop buttons must be pressed. Model QS-550 - Service Manual...
Page 39
"T" for table receptor or "W" for wall receptor. • When Quantum Medical Imaging, LLC. x-ray generator is in APR mode, it displays the current exposure selections (e.g., kVp, mA, time, etc.). When not in APR, displays either "MANUAL" or "AEC", depending on the mode selected.
Page 40
• A1U5, A1U6 and potentiometer A1R5 control the contrast voltage out- put to LCD Display EL31-005. Adjusting A1R5 changes the voltage level to the LCD Display. Varying this voltage level increase/decreases the contrast setting of the LCD display. Model QS-550 - Service Manual...
Check for proper counterbalancing and smoothness of vertical (up/down) tubestand motion • Verify that all ground conductors are properly and securely installed and free of corrosion or damage • Check all electrical cabling and wiring for wear and fraying Model QS-550 - Service Manual...
Page 44
2905 Veterans Memorial Highway Ronkonkoma, New York 11779 USA Phone: (631) 587-5800 When ordering replacement parts, supply the following information: • Model and serial number of equipment • Part number • Part description • Quantity required Model QS-550 - Service Manual...
Page 45
(ANGULATION MAGNET) Figure 21. Model QS-550 Tubestand Parts Location Diagram (LONGITUDINAL MAGNET) Figure 22. Model QS-550 Tubestand Parts Location Diagram...
Page 46
Chapter 4 Service Instructions Figure 23. Model QS-550 Tubestand Parts Location Diagram NOTE: Vertical SID switch and actuator are shown in Figure 23; the Horizontal SID switch, located in the Base Assembly, and longitudinal actuators have the same respective part numbers.
Page 47
Chapter 4 Service Instructions Figure 25. Model QS-550 Tubestand Parts Location Diagram Figure 26. Model QS-550 Tubestand Parts Location Diagram Model QS-550 - Service Manual...
Page 48
Chapter 4 Service Instructions THIS PAGE INTENTIONALLY LEFT BLANK Model QS-550 - Service Manual...