Safety Instructions This instruction manual provides you with important information about the OMRON ET720. To ensure the safe and proper use of this thermometer, READ and UNDERSTAND all of these instructions. If you do not understand these instructions or have any questions, contact 1-800-634-4350 before using this thermometer.
Important Safety Information Read the Important Safety Information in this instruction manual before using this device. Follow this instruction manual thoroughly for your safety. Indicates a potentially hazardous situation which, if not avoided, could result in death or Warning serious injury. •...
Page 5
Important Safety Information • If there is any temperature difference between the places where the unit is stored and where you are going to measure, leave the unit in the room where you are going to use it for more than 30 minutes to allow it to reach room temperature first, then measure.
Page 6
Important Safety Information • The “Repeatability” is 0.36 °F (0.20 °C). • The probe lens is the most delicate part of the thermometer. Use with care when cleaning the probe lens to avoid damage. a. When the probe lens becomes dirty, wipe it lightly with a soft dry cloth or a cotton swab. DO NOT wipe the probe lens with tissue paper or a paper towel.
2. Preparation Removing the Insulating Tape When using the thermometer for the first time, pull the insulating tape out of the battery compartment. Back of the unit The unit will turn on and after 1 minute the room temperature appears on the display. Note: •...
2. Preparation Switching between “°F” and “°C” This unit is set to “°F” as default. While the room temperature is displayed, press and hold the [START] button. While holding it down, press and hold the [ON/MEM] button until “°C” appears on the display with two beeps.
2. Preparation Setting the Buzzer The buzzer is activated by default, but can be deactivated. Note: The buzzer is only available in forehead measurement mode. Press the [ON/MEM] button. Press and hold the [ON/MEM] button for 3 seconds. The “ ”...
3. Using the Unit Taking a Reading Forehead Measurement Mode This mode displays the oral-equivalent value of the measured forehead temperature. Always make sure the probe lens is clean and undamaged, and the forehead is clean with no sweat, cosmetics, scars, etc.
Page 12
3. Using the Unit Hold the probe lens 0.4 to 1.2 inches (1 to 3 cm) away from the center of the forehead. Note: • Holding the unit too long in the hand may cause a higher ambient temperature reading of the probe.
Page 13
3. Using the Unit Move the unit away from the forehead and check the measurement result. The display lights up for 5 seconds. Note: • If the result is higher than 108.0 °F (42.2 °C), “Hi” appears on the display. If the result is lower than 93.2 °F (34.0 °C), “Lo”...
Page 14
3. Using the Unit Press and hold the [ON/MEM] button until “OFF” appears on the display. “OFF” appears on the display and the display will flash off, and then turn back on with the room temperature continually showing on the display. •...
Page 15
3. Using the Unit Surface Measurement Mode Surface measurement mode shows the actual and unadjusted surface temperature of an object, which is different from the body temperature. This mode can help you determine if the object temperature is suitable for a patient or baby (such as baby milk).
Page 16
3. Using the Unit Note: • Move the probe lens as close to the object as possible when using in surface measurement mode (0.4 inches (1 cm) apart is recommended). Do not touch the probe lens to the object directly. •...
3. Using the Unit Using the Memory Function This unit automatically stores the last 25 measurement results (excluding Hi/Lo results). Note: • If the memory is full, the unit will delete the oldest reading. • When the unit is switched between “°F” and “°C”, all the readings stored in the memory are deleted. Press the [ON/MEM] button.
4. Troubleshooting and Maintenance The Icons and Error Messages In case of any of the below problems occur during measurement, check that no other electrical device is within 12 inches (30 cm) of the device. If the problem persists, please refer to the table below. Error Display Possible Cause Solution...
Page 19
• Forehead measurement mode: The temperature Use the unit within its measurable range. If a reading is higher than 108 °F (+42.2 °C). malfunction still exists, please contact OMRON • Surface measurement mode: The temperature customer service. reading is higher than 176 °F (+80 °C).
Page 20
4. Troubleshooting and Maintenance Error Display Possible Cause Solution Change to a new battery. The battery is depleted. (Refer to sub-section 4.4) Remains on the display.
4. Troubleshooting and Maintenance Questions and Answers 1. The temperature indicated is rather high. • The unit has been stored in a cold room. When the unit has been stored in a cold room, the measured temperature may be rather high if you measure the temperature shortly after bringing the unit to a room that is at normal temperature.
Maintenance • Check the unit can be operated properly if it is dropped. If unsure, please contact OMRON customer service to have the unit checked. • When the probe lens becomes dirty, wipe it lightly with a soft dry cloth or a cotton swab.
4. Troubleshooting and Maintenance Replacing the Battery Battery: CR2032 lithium button battery Use the battery in accordance to the battery life. Note: To protect the environment, discard the used batteries in accordance with the local regulations. Release the screw and remove the battery cover. Remove the battery with a pointed object such as a pen or pencil.
5. Specifications Product Name No-Touch Forehead Thermometer Model # ET720 MC-720-Z Sensor Thermopile Temperature Display 4-digit, °F display in 0.1 degree increments 3-digit, °C display in 0.1 degree increments Accuracy Forehead measurement mode ±0.4 °F (±0.2 °C) within 95.0 to 107.6 °F (35.0 to 42.0 °C), ±0.5 °F (±0.3 °C) for other range...
Page 25
• The specification may be changed without prior notice. • This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co. Ltd., Japan. • IP classification is degrees of protection provided by enclosures in accordance with IEC 60529.
6. FCC Statement FCC CAUTION Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. This device complies with part 15 of FCC Rules, Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
7. Limited Warranty The OMRON ET720 No-Touch Forehead Thermometer excluding batteries, is warranted to be free from defects in materials and workmanship appearing within 1 year from the date of purchase, when used in accordance with the instructions provided with the thermometer. The above warranty extends only to the original retail purchaser, and only to products purchased from an Omron authorized seller who is subject to and follows Omron’s quality control standards, unless otherwise prohibited by...
Page 28
7. Limited Warranty This warranty provides you with specific legal rights, and you may have other rights which vary by jurisdiction. Because of special local requirements, some of the above limitations and exclusions may not apply to you. FOR CUSTOMER SERVICE Visit our web site at: OmronHealthcare.com Call toll free:...
Accompanying Documents in the Scope of IEC60601-1-2:2014 Important information regarding Electromagnetic Compatibility (EMC) ET720 conforms to IEC60601-1-2:2014 Electromagnetic Compatibility (EMC) standard. Further documentation in accordance with this EMC standard is available at OmronHealthcare.com/emc. Refer to the EMC information for ET720 on the website.
9. Some Useful Information How the No-Touch Forehead Thermometer Works This unit measures the temperature at the surface of the forehead by sensing infrared emission of the skin at the measured point, and then calculates the oral-equivalent value using a unique algorithm based on actual body temperature survey data. About Body Temperature Body temperature refers to the stable temperature of the inner body (also known as core body temperature).
Page 31
Forehead temperature (°C) Surface measurement mode Displays the actual temperature at the surface of the forehead. (Survey conducted by Omron) (room temperature at 23 °C) 36.5 37 36.5 37.5 38 37.5 38 38.5 39...