Philips Avent Plus SCF373/11 User Manual page 27

Single/double electric breast pump
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Emissions
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The Philips Avent single/double electric breast pump Plus is intended for use in the electromagnetic
environment listed below. The user of the device should ensure that it is used in such an environment
to prevent improper operation.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Radiated RF EM fields
IEC 61000-4-3
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
Rated power frequency
magnetic fields IEC
61000-4-8
Electrical fast transient/
burst
IEC 61000-4-4
Surges
IEC 61000-4-5
Compliance
Class B
Complies
Class A Category
Complies
Immunity test level
±8 kV contact
±15 kV air
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
385 (27), 450 (28), 710 (9),
745 (9), 780 (9), 810 (28),
870 (28), 930 (28), 1720
(28), 1845 (28), 1970 (28),
2450 (28), 5240 (9), 5500
(9), 5785 (9) MHz (V/m)
30 A/m
50 Hz or 60 Hz
±2 kV
100 kHz repetition
frequency
±1 kV Line-to-line
±2 kV Line-to-ground
Electromagnetic environment –
guidance
The device is suitable for use in locations in
residential environments and in
establishments directly connected to a low
voltage power supply network which supplies
buildings used for domestic purposes.
Electromagnetic environment -
guidance
Floors can be a source of static electricity.
When the air is very dry (<20 % RH) the user is
advised to prefer to use the device in a room
with concrete or wooden floors (instead of
using it in a room with a synthetic floor).
The device is immune against any RF wireless
communication device such as smart phone,
Wi-Fi or Bluetooth device at a distance of
30 cm or more. Avoid a closer distance as this
may result in improper operation.
Mains power quality expected of a typical
commercial or residential environment.
Mains power quality expected of a typical
commercial or residential environment.
English
27

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Avent plus scf371/11Avent plus scf338/11