Technical data
Permissible
15% – 90%, non-condensing
humidity (opera-
Note: Do not expose the device to the maximum humidity of 90% for more than 96
tion)
hours.
Maximum oper-
10,000 ft (20.67 Hg)
ating altitude
3000 m (700 hPa)
above sea level
Number of fixed
6
programs
Number of fixed
5
training plans
Number of pro-
Up to 200
gram memories
Number of chan-
10 output pulses in multiplex mode
nels
Maximum peak
150 V
level of the
The maximum peak level of the output voltage decreases due to the internal resistance
output voltage
of the end stages with decreasing load impedance.
Pulse width set-
50 μs – 400 μs in steps of 25 μs
ting
Pulse rise setting 0.0 s – 1.0 s in steps of 0.1
Frequency set-
2.0 Hz – 10.0 Hz in steps of 0.5; 10.0 Hz – 150.0 Hz in steps of 1.0
ting
Pulse pause set-
0.0 s – 1.0 s in steps of 1.0; 1.0 s - 10.0 in steps of 0.1
ting
Pulse time set-
1.0 s – 10.0 s in steps of 0.1
ting
Time specifica-
0.5 min – 5.0 min in steps of 0.5; 5.0 min – 30.0 min in steps of 1.0; 30.0 min – 90.0 min
tion setting
in steps of 2.0
Effective current
≤ 0.48 mA r.m.s./cm² (at 500 ohm)
density per cm²
of electrode area
Effective output
Ueff = 15 V at 500 ohm (30 mA)
value
Assuming a load impedance of 500 ohm, this produces a current intensity level of
30 mA r.m.s. The output voltage does not contain any DC component.
Output waveform
– bipolar