Gtos (Gate Turn-Off Thyristor) - Larkin M5500 User Manual

Power semiconductor tester
Table of Contents

Advertisement

GTOs (Gate Turn Off Thyristors)
Make sure the Voltage Control is at the zero position (fully counter clock-
wise) before turning the unit on. Connect the Anode, Gate and Cathode leads
to the device to be tested. DO NOT MIX UP THE ANODE AND
CATHODE LEADS OR HIGH VOLTAGE WILL BE APPLIED TO
THE GATE!
Place the function switch in the PRV position and press the "TEST" button.
Slowly increase the voltage while observing the Peak Voltage and Peak
Leakage Current on the panel meters.
In a properly functioning device, the rated voltage should be reached before
the current starts to rise rapidly, (the "break-over point").
Note: Most GTOs are "Anode Short" type, which means that the anode break
-over voltage is only between 17 and 19 volts.
Place the function switch in the PFV position and repeat procedure described
above.
Place the function switch in the GATE position and slowly raise the Voltage
Control while observing the "TRIGGER INDICATION" light.
If the light does not turn on, the Gate on the SCR is faulty, and the device
should be replaced.
Note: The Gate on a GTO turns on at a lower voltage than a SCR. Typical
values are between 0.6 and 0.8 volt.
- 10 -