Qsu s ® ® ilbiss uction erVice anual MODEL 7314 SERIES CAUTION-Federal (U.S.A.) law restricts this device to sale by or on the order of a physician. Assembled in USA...
It has been dropped or damaged . It has been dropped into water . Instead return the product to an authorized DeVilbiss Healthcare service center for examination and repair . Check the collection container for cracks before each use . Do not use if cracked .
G E N E R A L I N F O R M AT I O N Do not place the suction unit in a position that would make it difficult to disconnect the AC power cord. 10 . When the device is used under extreme operating conditions and in the carrying case, the surface temperature of the device may exceed 73°C .
I M P O R TA N T PA R T S 7314 Series DeVilbiss Vacu-Aide QSU Suction Unit with Disposable Container and with Disposable Container and with Reusable Container and Internal Filter Cartridge External Bacteria Filter External Bacteria Filter Vacuum gauge Vacuum gauge Vacuum gauge...
S E T U P A N D O P E R AT I O N Fully charge battery for 17 hours . (7314P series only) 5B & 5C . Reusable Container and Disposable Container w/ external Bacteria Filter Connection: Connect either end of the 4-1/2”...
Page 7
S E T U P A N D O P E R AT I O N 7B . 7314D series (non-battery label) 11 . Adjust the suction level . NOTE– The 7314D series is not factory equipped with an 12 . Verify suction level . NOTE– Always verify suction level internal rechargeable battery.
“Off’ position. If the battery does not charge, please be sure the model you are using has a battery installed prior to contacting your authorized DeVilbiss Healthcare provider. NOTE– The internal rechargeable battery is sealed lead-acid. Contact your local authorities for instruction on proper disposal.
Do not reverse direction of filter. Additional filters (7305D-608 12/pack) may be purchased from your authorized DeVilbiss Healthcare provider . NOTE– Do not substitute any other material for this filter cartridge. Substitution may lead to contamination or poor performance;...
C L E A N I N G I N S T R U C T I O N S 5A. Remove filter cartridge and 4-1/2” (11.43 cm) tubing and WARNING set aside . To prevent possible risk of infection from contaminated cleaning/disinfection solutions, always prepare fresh solution for each cleaning cycle and discard solution after each use.
C L E A N I N G I N S T R U C T I O N S Remove bacteria filter, elbow, and 3-3/8” (8.57 cm) tubing Disposable Collection Container (with and set aside . External Bacteria Filter) NOTE–...
C L E A N I N G I N S T R U C T I O N S Remove bacteria filter, elbow, and 4-1/2” (11.43 cm) tubing Reusable Collection Container (with and set aside . External Bacteria Filter) Turn unit “off”...
C L E A N I N G I N S T R U C T I O N S For single patient use: Keep outer surface clean by wiping with clean, damp cloth . Soak in 1 part vinegar (>=5% acetic acid concentration) to 3 parts water (131ºF-149ºF/55ºC-65ºC) solution for 60 minutes .
CAUTION– Do not submerge in water as this will result in Patient Change damage to the vacuum pump. DeVilbiss Healthcare recommends that at least the following NOTE– Do not use any cleaners or disinfectants that contain ammonia, benzene and/or acetone to clean the unit.
T R O U B L E S H O O T I N G TROUBLESHOOTING GUIDE PROBLEM CAUSE SOLUTION No power at receptacle Use other receptacle Power cord is loose or not connected Connect power cord Defective AC to DC adapter Replace adapter 1 .
S E R V I C E A N D C O M P O N E N T R E P L A C E M E N T FUNCTIONAL TEST AND EVALUATION BATTERY REPLACEMENT Remove bottom cover (refer to cover removal instructions) . The following should be performed after repair or between patients: Disconnect battery wire connector from circuit board and...
S E R V I C E A N D C O M P O N E N T R E P L A C E M E N T VACUUM ADJUSTMENT KNOB W/ PISTON CYLINDER AND CUP SEAL GAUGE REPLACEMENT REPLACEMENT Remove bottom cover (refer to cover removal Remove compressor (refer to compressor replacement...
F I G U R E A N D V I E W S 6’ Patient Tubing Vacuum Adjustment Knob w/Gauge Handle AC to DC Adapter/ Quilt Charger Filter Cartridge w/ Tubing Power Cord Collection Container Motor Bumpers Top Cover Feet Bottom Cover...
Page 19
F I G U R E A N D V I E W S LED’s Battery Connector Compressor Connector Printed Circuit Board Printed Circuit Board Compressor Connector 7314D Series 7314P Series NOTE– Power switch and DC input jack are located on back NOTE–...
PA R T S L I S T & A C C E S S O R I E S PARTS LIST 7314P-614 Battery 7314D-616 Bottom Cover w/Feet & Bumpers 7314D-605 Compressor & Piston Assembly 7310P-627 Cover Screws (15/pkg) 7305D-615 Cup Seal (5/pkg) 7314D-627 Exhaust Chamber...
S P E C I F I C AT I O N S / C L A S S I F I C AT I O N S Size (including container) 8 .3 H x 8 .0 W x 8 .5 D (21 .1 cm x 20 .3 cm x 21 .6 cm) (not including AC to DC adapter) Weight (including container) 7314P Series - 6 .6 lb .
Healthcare’s option if the unit has not been tampered with or used improperly during that period . Make certain that any malfunction is not due to inadequate cleaning or failure to follow the instructions . If repair is necessary, contact your DeVilbiss Healthcare Provider or DeVilbiss Service Department for instructions: U .S .A .
G U I D A N C E A N D M A N U FA C T U R E R ’ S D E C L A R AT I O N WARNING Medical Electrical Equipment needs special precautions regarding Electromagnetic Compatibility (EMC) and needs to be installed and put into service according to the EMC information provided in the accompanying documents.
Page 24