INSTRUCTION MANUAL
LCD ADJUSTMENT UNIT
※The operator should read this Instruction Manual carefully and handle the
device correctly.
1-8-18 Norimatsu-Higashi,Yahatanishi-ku,Kitakyushu 807-0837 Japan 
Phone No.(8193)691-3731  Fax No.(8193)691-3735 
http://www.matsushima-m-tech.com 
E-mail sales@matsushima-m-tech.com 
FOR
TYPE: GRAPHIC COM4
TGRAPHIC_COM4-003E
Rev0:1. Mar, 2018

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GRAPHIC COM4 and is the answer not in the manual?

Questions and answers

Summary of Contents for Matsushima GRAPHIC COM4

  • Page 1 TGRAPHIC_COM4-003E Rev0:1. Mar, 2018 INSTRUCTION MANUAL LCD ADJUSTMENT UNIT TYPE: GRAPHIC COM4 ※The operator should read this Instruction Manual carefully and handle the device correctly. 1-8-18 Norimatsu-Higashi,Yahatanishi-ku,Kitakyushu 807-0837 Japan  Phone No.(8193)691-3731  Fax No.(8193)691-3735  http://www.matsushima-m-tech.com  E-mail sales@matsushima-m-tech.com ...
  • Page 2: Table Of Contents

    Table of contents Safety precautions ・・・・・・・・・2 1. Overview ・・・・・・・・・3 2. Mounting/dismounting the LCD adjustment unit ・・・・・・・・・3 3. Key functions ・・・・・・・・・3 4. Display description ・・・・・・・・・4 4.1 Measurement display ・・・・・・・・・4 4.2 Menu display ・・・・・・・・・4 4.3 Waveform display ・・・・・・・・・5 4.4 PC Remote active display ・・・・・・・・・5 4.5 Error display ・・・・・・・・・6...
  • Page 3: Safety Precautions

    Cause for any peripheral equipment or device. d. Accident beyond control and force majeure (fire, earthquake, flood, riots, etc.). Lack of instructions to MATSUSHIMA for information or safety requirements that can be predicted only by customers’ side. This warranty conditions do not limit customers’ legal right.
  • Page 4: Overview

    Dismounting: Rotate LCD adjustment unit left to the direction "OPEN" and remove. 3. Key functions Detailed key functions of the LCD adjustment unit are described below. LOCK OPEN Fig. 3.1 GRAPHIC COM4 external view Table 3.1 Key functions Key  When menu selected  When enter/select ...
  • Page 5: Display Description ・・・・・・・・・4

    4. Display 4.1 Measurement display ① Measured value Display current measured value and respective unit. Measured value can be displayed in distance (m/ft), 12.345m  ① level (m/ft), level percent (%) or current (mA) mode. ② Distance  ② Display mode ③ APS ...
  • Page 6: Waveform Display

    4.3 Waveform display ① ⑤   ① Measured distance   D= 35.000m Current measured distance   ② Y axis ② ④ Reflection level (unit: dB) ③   ③ X axis Distance (unit: m or ft) 図4-3-1. エコーカーブ画面 Fig. 4.3.1 Echo curve view  ...
  • Page 7: Error Display

    4.5 Error display ① Error code (E****) When there is fault in the level meter or error occurs in 12.345m  communication between level meter and adjustment unit, then error code will be presented. Distance  Refer to table below for error code details. ①...
  • Page 8: Adjustment And Functions

    Adjustment and functions 5.1 Start up ≪Start up view≫ After power up the level meter initialize view is displayed. GRAPHIC COM4  REV.*.**  ≪Measured value≫  Display mode Description Distance Distance from level meter Level Distance Distance from tank base 12.345m  Level % Level percentage Current Current output value...
  • Page 9: Sensor Identification Setting

    5.2 Sensor identification setting In this menu you can adjust following parameters: Tag, Descriptor, Message and Date. 5.2.1 Tag In the menu item "Tag" enter tag name with less than 16 characters. (Default: SENSOR) 【Measurement view】  Character input area will Tag change Push 『Ent』...
  • Page 10 5.1.2 Descriptor In the menu item "Descriptor", enter descriptor with less than 16 characters. (Default: PULSE-RADAR) 【Measurement view】  Push 『Ent』 to Character input area Descriptor change enter main menu. will be activated. 12.345m  PULSE-RADAR̲  Push 『+』or 『®』  OK  button to select input Distance ...
  • Page 11: Message

    5.2.3 Message In the menu item "Message", enter message with less than 21 characters. (Default: LEVEL METER) 【Measurement view】 Character input area Message change Push 『Ent』  to enter will be activated. LEVEL METER̲  main menu. Push 『+』or 『®』 12.345m   OK  button to select input @ A B C D E F G H I J K L M N O characters.
  • Page 12: Date

    5.2.4 Date Set installation date of level meter. (Default: 2018/1/1) 【Measurement view】  Push 『Ent』  to enter main menu. 12.345m  Identification setting 1 Push 『Ent』to set        = PULSE-RADAR  date change.  3.Message  Distance         = LEVEL METER   4.Date         = 2018/09/01   ▼  Ent ...
  • Page 13: Basic Setting

    5.3 Basic setting In this menu basic setting for level meter measurement will be adjusted. 5.3.1 Application Measurement unit, measuring object and level change rate will be set. 5.3.1.1 Measurement unit (Meas.unit) Selects measuring unit of level meter from meter (m) and feet (ft). (Default: m) ※This setting will affect all measuring units.
  • Page 14 5.3.1.2 Measuring object Measuring object can be selected form “Liquid” and “Solid”, to change parameter sets for easier measurement. (Default: Liquid) ※The measuring object will change other parameters simultaneously. 【Measurement view】 Push 『Ent』  to enter main menu. 12.345m  Push『Ent』to select Application Distance ...
  • Page 15 5.3.1.3 Level change rate Measuring object level change rate can be selected form “Normal” and “Fast” to change parameter sets for easier measurement. (Default: Normal <1m/min) ※The level change rate will change other parameters simultaneously. 【Measurement view】  Push 『Ent』  to enter main menu.
  • Page 16: Measurement Range Span

    5.4 Measurement range span (Meas. range span) Measurement range span sets full/empty distance and related percentage. Here, measurement distance is distance from level meter. 【Measurement view】  Push 『Ent』  to enter Push『→』to move Meas. range span  main menu. 12.345m   1.Full setting  cursor to desired    Distance = 01.000m ...
  • Page 17: Damping

    5.5 Damping Sets amount of damping in number of seconds to respond fast step inputs. 【Measurement view】  Push 『Ent』  to enter main menu. 12.345m  Damping view will be Damping  displayed. Distance  Push 『 Ent』 to enter edit          =  0sec  entry.
  • Page 18 5.6 Current output setting (Current output set.) 5.6.1. 0-100% current output selection Sets 0-100% current output values. (Default: 4-20mA)                         『4-20mA』= 100% level (Upper level) is 20mA 0% level (Lower level) is 4mA 『20-4mA』=...
  • Page 19: Current Output Setting ・・・・・・・・・17

    5.6.2 Alarm current output Sets output current value during fault condition. Alarm current selection: Output current value when measurement is not valid because of no reflection echo loss or other reasons. Selection =「Hold」 「Sel.val.」 「Max」 「Min」 ・ Hold: Output remains at previous measured result during alarm. ・...
  • Page 20 Cont’d from previous page Current output set.  3.1 Push 『Ent』  to enter Sel. 3.Max current sel.  alarm current value. = 20mA  4.Min current sel.  Sel. alarm current val.         = 4mA  will be indicated by block 5.Sel.alarm current val.         =  22mA  cursor. ▼  Ent  × 1 click Current output set.  3.1 Use『+』or『®』and change 3.Max current sel. ...
  • Page 21 5.7 Echo learning Sets mask to unwanted reflections (false echoes or noise echoes) reflected from obstructions in the tank. Important:In most cases sets distance from level meter to unwanted reflections as echo learning distance. If there is true echo between the level meter and the masking distance, ! ...
  • Page 22: Echo Learning

    Cont’d from previous page Echo learning 3.2.1 ・"Selectable execution Push『Ent』set to  Distance  = 2.00m  ‘Update’ 【Clear/Update/Addition】  Execution = Update  OK  Clear   :Deletes all echo learning data. Update  :Deletes current echo learning data Ent  × 1 click and writes new data. Addition :Adds echo learning data Echo learning 3.2.1 To execute echo...
  • Page 23: Tw Adjustment

    5.8 TW adjustment TW (Time window) adjustment is used to keep TW peak temporary when peak moved outside of TW. (This setting used only when application->level change rate =Fast>1m/min). If TW adjusted by this setting is not correct then must be set manually to correct value. 【Measurement view】 ...
  • Page 24: Graph Display

    6. Graph display Displays either of echo curve only or echo detection curve + echo curve + time window from current active measurement. Waveforms are updated automatically. ・Echo curve: Reflection waveform (EC) ・Echo detection curve + EC + TW: Detection waveform (EDC) + Reflection waveform (EC) + Time window (TW) 6.1 Echo curve (EC) Displays echo curve of current active measurement.
  • Page 25 6.2 Echo detection curve + EC + TW Displays echo curve, echo detection curve and time window of current active measurement simultaneously. ※This view will not return to measurement view automatically. 【Measurement view】  Push 『Ent』  to enter main menu. 12.345m  Use『®』button to change Distance ...
  • Page 26: Self Test

    7. Self test Runs self diagnosis routine and if there are any fault detected then error code will be displayed. 【Measurement view】  When diagnosis Push 『Ent』  to enter Self test  completed diagnosis main menu. 12.345m  result will be displayed. OK  If there are no any fault Diagnosis result ...
  • Page 27: Reset

    8. Reset There are two reset options. Use "Measuring reset" to restart measurement without affecting parameters. Use "Parameter reset" to reset parameters to the default settings. Important:"Parameter reset" ・ Parameter reset returns various parameters to instrument default. Please take note of current settings before execute parameter reset.
  • Page 28: Current Output Test

    9. Current output test Outputs entered simulation value in order to test the functioning of the current output. (Level input range: -10.00% to +110.00%, current input range: 3.55mA(3.60mA:Microwave Level Meter under Ver2.0) to 22.00mA) Important : When you wish to end the simulation and return the instrument to actual ! ...
  • Page 29: Hart Communication Settings

    10. HART Communication setting Sets HART polling address and request preamble number for HART communication. 10.1 Polling address Sets polling address for HART communication. (Default: 0, input range: 0 to 63) Important: If polling address in not multidrop mode were changed unintentionally use『®』and ! ...
  • Page 30: Preamble Number

    10.2 Preamble number Set preamble number for HART communication. (Default:5, Input range: 5 to 20) Important: If preamble number in not multidrop mode were changed unintentionally use 『®』 and !  『+』buttons set to default value 5. 【Measurement view】  Push 『Ent』  to enter main menu.
  • Page 31: Multidrop Mode

    10.3 Multidrop mode ※This mode can be set in Microwave Level Meter over Ver2.0 Set ‘Disable’ or ‘Enable’ of Multidrop mode. (Default:Disable) Important: When Multidrop mode is enable , !  9. Current output test and 5.6 Current output setting (Current output set.)can not set values. 【Measurement view】 ...
  • Page 32: Dynamic Variables

    10.4 Dynamic valuables setting ※This mode can be set over in Microwave Level Meter over Ver2.0 Set ‘Distance’ or ‘Level Distance’ or ‘Level%’ in each Dynamic valuables. (Default:PV=Distance、SV=Level Distance、TV=Level%、 QV can not be seleted)  【Measurement view】 Push 『Ent』  to enter main menu. 12.345m  Select Distance/Level HART communication    8 Distance/Level%.
  • Page 33: Sensor Information

     6. Reset   7. Current output test   8. HART communication   9. Sensor information ▼  Ent  × 1 click Sensor information view Sensor information Manufacturer  will be displayed.      = Matsushima  Push 『+』 button to select  Device type  next sensor information      = MWLM-PR26  items.  Serial number      = 1  ▼ × several times + ...
  • Page 34: Display

    12. Display Sets measurement view display mode and display language. 12.1 Language setting Sets display language for Graphic COM4 to English or to Japanese (Kana). (Default: English) 【Measurement view】  Push 『Ent』  to enter main menu. 12.345m  ディスプレイ Language selection value 1.ゲンゴ ...
  • Page 35 12.2 Measurement view display mode Select measurement view display mode from distance, level Distance, level percent and current output. (Default: Distance) 【Measurement view】 Push 『Ent』  to enter main menu. 12.345m  Display selection will be Display   1.Language  indicated by block cursor.  ...
  • Page 36 12.3 Backlight setting Select Auto or ON or OFF in Backlight function.(Default: Auto) Important:In case of connecting Graphic com4 to Microwave Level Meter under Ver2.0, !  Backlight function can not be selected except for [Auto]. 【Measurement view】  Push 『Ent』  to enter main menu.
  • Page 37: Contrast

    12.4 Contrast Sets the contrast of the screen. (Default: 3) 【Measurement view】  Push 『Ent』  to enter main menu. Push 『Ent』 , then value to 12.345m  Display   be changed will be 4.Contrast  indicated by block cursor.           = 3 Distance  Each time you press 『+』, the number increases and the contrast becomes...
  • Page 38: Parameter List

    13. Parameter list Parameter Display abbreviation Range Default Menu Menu      Identification setting 1  Identification setting 1.1  Tag  Less than 16 ASCII characters  SENSOR  Descriptor 1.2  Description  Less than 16 ASCII characters  PULSE-RADAR  Message 1.3  Message  Less than 32 ASCII characters  LEVEL METER  Installation date 1.4  Installation day  2010/1/1 to 2155/12/31  2018/1/1  Basic setting 2  Basic setting  − ...
  • Page 39 Parameter Display abbreviation Range Default HART communication 8  HART communication  −  −  setting Polling address Polling address  0〜15  0  Preamble number Preamble number  5〜20  5  Multidrop mode Multidrop mode  Enable/Disable  Disable  ※Level Meter over Ver2.0 Dynamic Variables Dynamic Variables  PV/SV/TV=  PV=Distance  ※Level Meter over Ver2.0 Distance/Level Distace/Level%  SV=Level Distace ※QV can not be selected  TV=Level%  Sensor information 9 ...
  • Page 40: Troubleshooting

    LCD adj. unit failed Ensure LCD adj. unit attached properly ※When the device failure or measurement failure occur, measurement distance will not change even waveform being changed. ※If above mentioned problems persist, please contact your local Matsushima sales office concerning device failure.