Faults
The user may encounter faults while operating the laser system. When a fault occurs a fault code
appears on the display followed by a brief text message. There are two types of faults: 1) errors,
or 2) warnings.
Errors faults are critical and will cause the laser to stop operating. These faults codes begin with
the prefix 'E' meaning error. Some of these errors require that you contact the Cynosure Service
Department.
Warnings are faults that indicate an improper laser state or operator error. These fault codes
begin with the prefix 'W' meaning warning. For example, an attempt to fire the laser without a
fiber present causes a warning fault to occur. The operator can usually correct these faults.
All faults codes and messages are listed in the following table, along with the probable cause and
corrective action if applicable.
Fault Code Table
Fault Code and Text Message
E01
Water Flow Fault; Cycle Power
E02
Open Shutter Fault
E03
Close Shutter Fault
E04
Fuse Open; Call Service
E05
Checksum Error; Call Service
E07
IGBT Fault - YAG
E08
Heater Fault; Call Service
E09
Coolant Sensor Fault; Call Service
E10
No High Voltage
E13
No Simmer YAG
E14
HVPS Fault
E15
Dump Fault; Call Service
E16
Dump Fault; Call Service
E17
Dump Fault; Call Service
E18
H2o Flowmeter
E19
HVPS Fault
Low Trans; Replace
E23
Fiber/Handpiece
YAG Calibration Error; Call
E24
Service
E25
YAG Cal Failed, Energy Low; Call
Cynergy Operator Manual
Troubleshooting
Probable Cause/Corrective Action
Broken water pump or blown fuse. Needs service.
Safety shutter stuck closed. Do not use laser until
serviced.
Safety shutter stuck open. Do not use laser until
serviced.
Blown capacitor bank fuse or shorted dump circuit.
Needs service.
Computer program memory error. Needs service.
YAG IGBT overload sensed. Cycle power to continue.
Open thermostat or heater. Blown heater fuse. Needs
service.
Open thermistor sensor or unplugged. Needs service.
Capacitor bank or HVPS wire shorted. No read back
voltage. Needs service.
Broken lamp or blown fuse. Needs service.
HVPS over-temperature.
Fuse fault at power up. Needs service.
Dump always on. Needs service. Dump shorted. Failed
dump TX. HVPS failed.
Dump not working. Needs service. Fiber connections
broken to cap bank. HVPS disconnected or failed.
Defective water flow sensor. Needs service.
HVPS over-voltage or open-circuit.
Damaged fiber, handpiece or focus lens, needs
replacement.
Incorrect Resonator Port or Cal Port sensor calibration.
Needs service.
YAG cal port energy too low to use. Needs service.
850-1065-000, Rev. 6
Page 59 of 78