Output indication 
does not change. 
Fault LED is 
lighted. 
[NOTE] 
(1)When removing Sensor, loosen M5 bolts and nuts. 
Securely tighten them at installation after cleaning it. 
(2)If insulation extension is required, consult with Sales Dept. of Matsushima.(non-free) 
(3)Do not cut probe length too short in case any electric charge may not be detected.
・Power is not supplied. 
・The wiring between Sensor and Transducer is 
 either off or wrong. 
・Material sticking on Probe electrode. 
・Probe electrode is in contact with pipe inside 
 (GND). 
・Device is in trouble. 
・CPU abnormal. 
・Wiring is wrong. 
・The supplied power supply is out of allowable 
  range. 
・CPU abnormal. 
・Confirm power line is wired 
  to Terminal AC. 
  If not supplied, supply power. 
・Check wiring  of exclusive  cable. 
・Remove Sensor and check if any 
・Measure inside diameter and 
・Consult with Sales Dept. to 
・Power OFF, then ON. 
・ Check wiring of power/output 
  (See Page 6, 5.How to connect.)
  If wiring is wrong, rewire 
・Check power supply with a Tester 
・ Power OFF, then ON. 
- 13 -
(See Page 6, 5.How to connect.)
If  wiring  is  wrong,  rewire 
correctly. 
 sticking on Probe electrode. 
If yes, power off and wipe them 
off with moist cloth. (See NOTE)
 Probe electrode  length and  check
 if they are not in contact. 
If  yes,  change  electrode  length.
(See NOTE) 
check the soundness of Device. 
signal lines, exclusive cable.
correctly. 
etc.. 
If it is out of the allowable 
range, apply the one within the 
range. 