Prosonic M FMU 40/41/42/43 with Foundation Fieldbus
!
"
!
5.5
Lock/unlock configuration
5.5.1
Software security locking
Enter a number ≠ 2457 in the "unlock parameter" (0A4) function in the "diagnostics" (0A)
function group.
The
symbol appears on the display. Inputs are no longer possible.
If you try to change a parameter, the device jumps to the "unlock parameter" (0A4) function.
Enter "2457"
Now change the parameters.
5.5.2
Hardware security locking
Press S, O and F simultaneously.
Inputs are no longer possible.
If you try to change a parameter, the following appears:
Press S, O and F simultaneously. The "unlock parameter" (0A4) function appears.
Enter "2457"
Now change the parameters.
Note!
A hardware locking can only be unlocked again via the display by pressing the O, S and F keys at
the same time again. It is not possible to unlock the hardware by communication.
5.6
Resetting the customer parameters
It is advisable to reset the customer parameters if you want to use a device with an unknown history.
Effects of resetting:
• All customer parameters are reset to their default values.
• Customer interference echo suppression is not deleted.
• Linearisation is switched to "linear", but the table values are kept. The table can be switched
back on in the "linearisation" (04) function group in the "linearisation" (041) function.
In order to carry out the reset, enter the number "33333" in the "reset" (0A3) function in the
"diagnostics" (0A) function group.
Caution!
A reset may lead to impairment of the measurement. As a rule, a basic calibration is required after
a reset.
Note!
The default values of each parameter are shown in bold in the menu overview in the appendix.
Operation
L00-fmrxf0a4-20-00-00-de-001
45