Summary of Contents for Vatech EzRay Premium VEX-S100W
Page 1
User manual Model: VEX-S100W Version: 1.35 • English Full version Postal code: 18449 13, Samsung 1-ro 2-gil, Hwaseong-si, Gyeonggi-do, Korea www.vatech.co.kr...
Page 5
For further information not covered in this manual, please contact us at: Phone: +82-1588-9510 E-mail: gcs@vatech.co.kr This document is originally written in English. The EzRay Premium is referred to as Equipment in this manual. Manual Name: EzRay Premium (Model: VEX-S100W) User Manual Document number: VDH-UM-056 Version: 1.35...
Page 6
Notice Left blank intentionally EzRay Premium User Manual...
Notice Table of Contents Notice Table of Contents General and Regulatory Information Manufacturer's Liability ............... 1 Owner and Operator's Obligations ............1 Conventions in this Manual ..............2 Marks and Symbols ................3 Standards and Regulations ..............4 Safety Instructions General Safety Guidelines ..............
Page 8
Notice Cleaning and Disinfection Cleaning .................... 34 Disinfection ..................34 Maintenance Maintenance ..................35 Service Disposing of the Unit Technical Specifications 11.1 Mechanical Specifications ..............41 11.2 Technical Specifications ..............43 11.3 Electrical Specifications ..............46 11.4 Environmental Specifications ............46 X-Ray Dose Data 12.1 X-Ray Dose Table ................
Genuine VATECH approved equipment and components have been used always. All maintenance and repairs have been performed by a VATECH authorized agent. The equipment has been used normally in accordance with the user's manual.
1. General and Regulatory Information Conventions in this Manual The following symbols are used throughout this manual. Make sure that you fully understand each symbol and follow the instructions which accompany it. To prevent personal injury and/or damage to the equipment, please observe all warnings and safety information included in this document.
1. General and Regulatory Information Marks and Symbols Symbols Description Location Alternate current Attention: consult accompanying documents Label Dangerous voltage Power board Protective earth (Ground) Power board Off (power: disconnect from the main switch) Main switch On (power: connect to the main switch) Main switch IEC60601-1 The degree of Protection from Electric Shock...
1. General and Regulatory Information Symbols Description Location This symbol indicates that electrical and electronic equipment must not be disposed Label of as unsorted municipal waste and must be collected separately. ESD susceptibility symbols indicate that an Board item is susceptible to damage from package electrostatic discharges.
Page 13
1. General and Regulatory Information Classifications (IEC60601-1 6.1): Protection against the ingress of water: Ordinary Equipment: IPX0. Protection against electric shock: Class I equipment, Type B Applied Parts. EzRay Premium User Manual...
Page 14
1. General and Regulatory Information Left blank intentionally EzRay Premium User Manual...
2. Safety Instructions Safety Instructions General Safety Guidelines This equipment is designed and manufactured to ensure maximum safety of operation. Operate and maintain it in strict compliance with the safety precautions and operating instructions contained in this manual. This equipment must only be operated by legally qualified persons.
Page 16
2. Safety Instructions Do not place flammable materials near this equipment. Never touch the patient while also touching the SIP/SOP connectors. Do not connect this equipment to a Multiple Portable Socket-Outlet (MPSO) or extension cord. The EzRay Premium X-Ray equipment complies with IEC standard 60601-1-2. ...
3. The System Overview The System Overview EzRay Premium is an intraoral source dental X-ray system intended for intraoral imaging. It consists of an X-ray generator, X-ray controller, beam limiting device, an operation panel, and a mechanical arm. The X-ray controller allows for accurate exposure control, and an adjustable mechanical arm allows for easy positioning.
3. The System Overview General View of the Equipment Item Description Cone Varied sizes adapted to films and digital sensors. (Collimator) Tube Head Generates high–frequency X-rays 3, 4 Articulating Arm Extends the arm Horizontal Arm 3 length options: 450 mm, 600 mm, 900 mm On / Off switch Turns the equipment ON/OFF Controls and manages the system;...
Page 19
3. The System Overview Available Option Items Arm’s length available: 450 mm, 600 mm, or 900 mm. Conehead (Collimator): 6 types; Length Shape Diameter (from the focal spot to edge of the cone) 200 mm default 60 ± 2(mm) Round 300 mm optional...
3. The System Overview Control Panel Description Exposure Button X-Ray Emission Indicator LCD (kV, mA, Time, Error Code, DAP) Power Status Indicator Parameter Adjustment Button Save Button Tooth Selector Adult / Child Selector Receptor Selector EzRay Premium User Manual...
Page 21
3. The System Overview 1. Exposure button When you take an exposure, you must press and hold down the exposure button for the duration of the exposure. 2. X-Ray emission indicator Ready: turns green X-Ray On: turns orange Refer to section 5.4 Tables of Exposure Time. COOLING TIME: The equipment requires a cooling time equal to 15 times the exposure time before taking another image.
Page 22
3. The System Overview 5. Parameter Adjustment Key The preprogrammed settings (tube voltage/current and exposure time) can be changed by the user. For manual adjustment, refer to “5.5 X-ray Generation Limits”. Icon Meaning Parameter Adjustment key: kV → mA → sec in a cycle. Up / Down for the kV, mA, and sec.
Page 23
3. The System Overview 8. Adult / Child selector Icon Meaning Adult Child 9. Receptor selector Icon Meaning Film Digital EzRay Premium User Manual...
Page 24
3. The System Overview Left blank intentionally EzRay Premium User Manual...
4. Preparations for Exposure Preparations for Exposure Turning the Equipment On Turn ON the equipment. The following displays light up: Power status indicator Anatomical selector Adult or Child selector Receptor selector Displays the exposure parameters: kV, mA, and exposure time EzRay Premium User Manual...
4. Preparations for Exposure Selecting the Cone Type Select the cone type to be used in the exposure. Use of the optional long cone is recommended to keep the patient’s absorbed dose as low as possible. Assembling the Cone: Long 30 cm (12″) Cone The long cone is attached into position by pushing it into the short cone and rotating it so that the guidelines on the short cone and long cone line up.
5. Using the System Using the System Exposure Parameters Select the receptor type: Film or Digital sensor. The indicator light of the selected projection will turn on. Select the type of patient: Adult or Child. The indicator light of the selected projection will turn on.
5. Using the System Positioning Positioning the Patient To obtain high-quality intraoral radiography with maximum details, take extra care in all steps of the radiography process: positioning the patient, the x-ray generator, and the imaging system; exposing the film or the intraoral sensor, and processing the film. The long cone (optional) can be attached into the short cone.
Page 29
5. Using the System Positioning Instructions Paralleling technique: the film or sensor is placed in a film holder which is used to align the film parallel to the long axis of the teeth. 1. Film or Sensor 2. Long axis of the tooth Bisected angle technique: the patient holds the film or sensor in place with his / her finger.
Page 30
5. Using the System Maxillary Incisor The x-ray beam is directed downward at 45°. Mandibular Incisor The x-ray beam is directed upward at 25°. Teeth Angle of inclination Incisor Maxilla +45˚ Incisor Mandible -25˚ EzRay Premium User Manual...
Page 31
5. Using the System Maxillary Molar and Premolar The x-ray beam is directed downward at 30°. Mandibular Molar and Premolar The x-ray beam is directed upward at 5°. Teeth Angle of inclination Molar and Premolar Maxilla +30˚ Molar and Premolar Mandible -5˚...
Page 32
Conventional dental films Digital sensors, such as VATECH’s EzSensor Phosphor plate Placing the receptor properly is important. Check your dental radiography text for information about the proper placement of the imaging receptor. Improperly positioning the film or sensor results in errors on the radiograph, such as distorted teeth and roots, elongation, magnification, and/or overlapping contacts.
5. Using the System Exposure The operator must instruct the patient to refrain from moving during the entire exposure. Ask the patient not to move. Move as far away from the X-ray tube as the length of the cable from the control panel permits.
5. Using the System Tables of Exposure Times (Default) The following exposure timetables were established with a unit equipped with a cone that corresponds to a focus-to-skin distance of 20 cm (8 inches) and 30 cm (12 inches) respectively. Standard: Option: Cone 20 cm Cone 30 cm...
5. Using the System X-Ray Generation Limit Tube Voltage, Tube Current, and Exposure Time value cannot be set to the maximum value simultaneously. If the parameter you set exceeds the safety limit, the “Over Value” message is displayed on the top left of the LCD. Unless this message is cleared, you cannot perform irradiation.
Page 36
5. Using the System Left blank intentionally EzRay Premium User Manual...
6. Emergency Measures Emergency Measures In instances of abnormal operation, error messages will be displayed on the LCD panel. If a problem persists, please request assistance from the customer support information services. Error Messages Error Code Cause Actions to be taken Turn off the power and check the connection between the panel board The connection between...
Page 38
6. Emergency Measures Error Code Cause Actions to be taken The kV feedback is higher than the set value by more " than + 20 kV during X-ray irradiation. The mA feedback is lower than the set value by more "...
Page 39
6. Emergency Measures Troubleshooting Problem Cause Solution Equipment is Connect the equipment disconnected Nothing lights up The circuit breaker is Reset the Circuit breaker tripped Control unit is Connect the control unit Control unit does not light disconnected Control unit is defective Call a qualified technician Check if the power plug is Equipment does not turn...
Page 40
6. Emergency Measures Left blank intentionally EzRay Premium User Manual...
7. Cleaning and Disinfection Cleaning and Disinfection The liquid can cause damage to the equipment. When cleaning or disinfecting, liquids may enter the equipment or the release button via the ventilation slots. Do not spray the equipment with Cleaners or disinfectant agents.
The use of unsuitable Cleaners or disinfectant agents and methods can damage the equipment and accessories. Only use the Cleaners or disinfectant agents specified or approved by VATECH The following Cleaners or disinfectant agents have been evaluated for safe use on the surfaces.
8. Maintenance Maintenance Maintenance Do not use force to unplug cables. The film holder can be autoclaved up to 145˚C or cleaned with alcohol-based solutions. Do not keep the equipment or its parts in a humid place or near a liquid substance.
Page 44
8. Maintenance Tasks Period Ensure there is no wear or damage to the exposure switch Monthly cable. Always turn off the equipment before performing any maintenance. EzRay Premium User Manual...
9. Service Service VATECH requires periodic constancy tests to ensure image quality and the safety of the patient and operator. Only VATECH authorized technicians can perform inspection and service of this equipment. For the technical assistance, contact VATECH service center or your local VATECH representative.
Page 46
9. Service Left blank intentionally EzRay Premium User Manual...
10. Disposing of the Unit 10. Disposing of the Unit To reduce environmental contamination, this equipment is designed to be as safe as possible to use and dispose of. Many components of this equipment are environment- friendly and can be recycled. All parts and components that contain hazardous materials must be disposed of in accordance with disposal regulations.
Page 48
10. Disposing of the Unit Left blank intentionally EzRay Premium User Manual...
11. Technical Specifications 11. Technical Specifications 11.1 Mechanical Specifications Dimensions: Wall-mounted Type EzRay Premium User Manual...
Page 50
11. Technical Specifications Horizontal arm (3 options) Note 1 Note 4 450 mm (17.7”) 713 mm (28”) 2,003 mm (78.9”) 600 mm (23.6”) 863 mm (34”) 2,153 mm (84.8”) 900 mm (35.4”) 1,163 mm (45.8”) 2,453 mm (96.6”) Item Description Total 31.5 kg (69.4 Ib) Tube Head...
11. Technical Specifications 11.2 Technical Specifications X-Ray Generator Item Description Model DG-10A05T3 Generator size Mono-tank: 140 x150 x 70 2.72 kg (Inverter: 0.92 / Monotask: Weight 1.8) Rated output power Max. 0.5 kW Tube Voltage (kV) 50 - 70 Tube Current (mA) 4.0 –...
Page 52
11. Technical Specifications EzRay Premium User Manual...
Page 53
11. Technical Specifications Tolerances of the focal spot on the reference axis X = ± 0.4 mm (sideways), Y = ± 0.4 mm (in-depth), Z = ± 0.4 mm (in height) Reference axis to which the target angle and the focal spot characteristics of the X-Ray tube refer EzRay Premium User Manual...
11. Technical Specifications 11.3 Electrical Specifications Item Description Power supply voltage AC 100 - 240 V Frequency 50 / 60 Hz (Single phase) Power rating 500W Maximum power 1.2kVA The input line voltage depends on the local electrical distribution system. ...
12. X-Ray Dose Data 12. X-Ray Dose Data The X-ray dose data is extracted from the X-Ray Dose Test Report for the EzRay Premium. The X-ray doses of EzRay Premium in the test report were measured in accordance with the IEC collateral standards. EzRay Premium was designed in accordance with Part 1.
12. X-Ray Dose Data 12.2 Leakage Dose Test Report Test Equipment Information Instrument Manufacturer Model S / N Dose Meter Victorian 101114 / 101377 Test Condition Distance from focal point [m] Applied Tube Voltage Peak [kVp] Applied Tube Current [mA] Applied Exposure time [s] Basic Cone [mGy/hr] Direction [°]...
Page 57
12. X-Ray Dose Data The tested area on the plane of radiation is defined within a radius at 1 m from the focal spot. [Measurement Points (Transverse / Longitudinal)] Beam Direction [Measurement Values (Transverse)] EzRay Premium User Manual...
Page 58
12. X-Ray Dose Data [Measurement Values (Longitudinal)] EzRay Premium User Manual...
13. Electromagnetic Compatibility (EMC) Information 13. Electromagnetic Compatibility (EMC) Information Guidance and manufacturer’s declaration - electromagnetic emissions. The EzRay Premium is intended for use in the electromagnetic environment specified below. The customer or the user of the EzRay Premium should assure that it is used in such an environment.
Page 60
13. Electromagnetic Compatibility (EMC) Information Guidance and manufacturer’s declaration - electromagnetic immunity The EzRay Premium is intended for use in the electromagnetic environment specified below. The customer or the user of the EzRay Premium should assure that it is used in such an environment. Immunity IEC 60601 Test Compliance...
Page 61
13. Electromagnetic Compatibility (EMC) Information Guidance and manufacturer’s declaration - electromagnetic immunity The EzRay Premium is intended for use in the electromagnetic environment specified below. The customer or the user of the EzRay Premium should assure that it is used in such an environment. IEC 60601 Compliance Electromagnetic...
Page 62
13. Electromagnetic Compatibility (EMC) Information Guidance and manufacturer’s declaration - electromagnetic immunity The EzRay Premium is intended for use in the electromagnetic environment specified below. The customer or the user of the EzRay Premium should assure that it is used in such an electromagnetic environment. Immunity IEC 60601 Compliance...
Page 63
No part of this manual may be reproduced, transmitted, or transcribed without the expressed written permission of the manufacturer. We reserve the right to make any alterations which may be required due to technical improvement. For the most current information, contact your VATECH representative. Manufactured by VATECH Co., Ltd.