Page 1
INSTRUCTIONS SUCTION PUMP KV- -5 CAUTION Federal (USA) law restricts this device to sale 0086 by or on the order of a physician.
Page 2
IMPORTANT Please read this manual carefully before attempting to use the Olympus KV- - 5 suction pump as it contains important information on the proper care, handling and use of the equipment. In addition, read the manuals of any other units (endoscope, light source, etc.) that form part of the system.
CONTENTS STANDARD SETS & FEATURES ......... OPERATING PRECAUTIONS .
Page 4
The KV--5 is available in two configurations -- as a single--use suction system or a reusable suction system. Unpack the equipment and confirm the items in the standard set ordered are present and undamaged. Retain the packaging for future use. Contact Olympus if any parts are damaged or missing. 1.1 Standard set - Single- use suction system...
Page 5
K7503763 K7503760 K7503657 Suction jar mounting kit for Suction jar mounting kit for 4mm A/F Allen key x1 Olympus WM- - N60/D60, Olympus TC- - C1/NE Olympus WM- - 30/60 K7503556 K7503555 Supplied with standard sets: Supplied with standard sets:...
Page 6
K7505595 (Americas only) mounted suction jar Filter connecting tube 900mm pk 10 (non- - sterile) Suction jar mounting kit for K7503657 Olympus WM- - 30/60 K7503555 For use with trolley mounted suction jar Suction jar mounting kit for Olympus WM- - N60/D60...
Page 7
1.4 Features Label 5 Single- - use Microbial filter suction liner Power switch Filter Label 1 connecting tube Label 2 Suction jar and lid Vacuum gauge (for indication only) Label 3 Vacuum regulator Label 6 FIGURE 1-1 Rear panel IEC power receptacle Label 4 (with integral fuses) Potential...
D The electrical installation of the room where the unit is to be used must comply with local or national regulations. D The KV-5 is not intended to be used when positioned on inclined surfaces as the suction jar fluid cut--off mechanism may not function correctly.
D If aspirated material has been drawn past the microbial filter and into the pump, the unit must be removed from use and returned to Olympus for service. D To prevent liquid backflow within the patient connecting tube, always leave the unit running until the patient tube is disconnected from the suction jar.
Page 10
1 litre suction jar Suction jar holder (3) Remove the transit bung from the filter housing (retain for transportation) on the front of the unit and insert the microbial filter with the word ‘Olympus’ facing outermost (see Figure 3-2) FIGURE 3-2 Filter housing...
Page 11
FIGURE 3-4 Microbial filter Filter connecting tube 90˚ connector Filter connecting tube (6) Ensure the KV--5 power switch is OFF and the vacuum regulator is set to ‘MIN’, then connect the power cable into the IEC receptacle on the rear of the unit and a suitably grounded AC wall outlet or isolation transformer socket. (7) Set the power switch to the ‘l’...
Page 12
1 litre suction jar Suction jar holder (4) Remove the transit bung from the filter housing on the front of the unit and insert the microbial filter with the word “Olympus” facing outermost (see Figure 3-2) FIGURE 3-8 Filter housing Microbial filter (5) Connect the patient tube between the suction jar lid (Figure 3-9) and endoscope or other equipment as required.
Page 13
FIGURE 3-9 Patient connecting tube (6) Push the open end of the filter connecting tube onto the microbial filter and fully insert the 90˚ connector into the suction jar lid marked VACUUM (Figure 3-10). FIGURE 3-10 Microbial filter Filter connecting tube 90˚ connector Filter connecting tube (7) Ensure the KV--5 power switch is set to O and the vacuum regulator set to ‘MIN’, then connect the power cable into the IEC receptacle on the rear of the unit and a suitably grounded AC wall outlet or isolation transformer socket.
Ensure the suction jar is thoroughly rinsed with fresh water following disinfection. Clean the KV-5 casing, including the microbial filter housing, by wiping with a cloth dampened with detergent and water or alcohol impregnated swab. Thoroughly rinse off any detergent using a soft cloth dampened with clean water.
Failure to fit the transport screws when transporting the unit may result in damage to the pump. AINTENANCE AND REPAIR There are no operator--serviceable components inside the KV--5. Refer servicing to qualified service personnel. If repair of the KV-5 is required, contact Olympus.
Page 16
NOTE: Always replace a fuse with one of the correct type and rating. If a fuse continues to blow, contact Olympus. The thermal cut-out within the windings of the pump motor may have operated. Check the vents are clear. When the temperature have operated.
Resistance to The external surfaces of the Olympus KV- - 5 are resistant to: 2% aqueous neutral detergent, 70% ethyl alcohol, chemicals isopropyl alcohol, water.
Refer to Figure 7-1 The pump used in the KV-5 is a piston unit. The pump is connected to the suction jar by silicone tubing and has an in-line microbial filter. A suction tube connects the endoscope or other devices to the suction jar. When the KV-5 is switched on, a vacuum is created in the suction jar, which results in suction through the endoscope when the appropriate endoscope control is operated, or through other devices.