Hearing aids HearLink 9030 not connecting to iphone
1 comments:
Mr. Anderson
March 16, 2025
Philips HearLink miniRITE T hearing aids may not be connecting to an iPhone due to several possible reasons:
1. Compatibility Issues – The iPhone model may not be compatible with the hearing aids. A list of compatible devices is available at hearingsolutions.philips.com/compatibility. 2. Wireless Interference – Other wireless devices or networks may be affecting the connection, as the hearing aids operate on the 2.4 GHz ISM band. 3. Device Settings – The hearing aids may not be properly paired with the iPhone, or Bluetooth may be disabled. 4. Battery Issues – The hearing aid battery may be low or depleted. 5. Unverified Accessories – Using accessories not certified by the manufacturer may affect performance.
For troubleshooting, users should check compatibility, ensure Bluetooth is enabled, re-pair the devices, and verify battery levels. If issues persist, consulting a hearing care professional is recommended.
Summary of Contents for Philips HearLink miniRITE T
Page 1
HearLink Receiver in-the-Ear Hearing Aids miniRITE T miniRITE T R Instructions for Use...
Page 2
Model overview This booklet is valid for the Philips HearLink™ family in the following hearing aid models: MNR T R MNR T FW 1.1 Philips HearLink 9030 MNR T R 05714464033171 (HER9032) Philips HearLink 7030 MNR T R 05714464033188 (HER7032) ...
Page 3
The following speakers are available for the above model: Speaker 60 Speaker 85 Speaker 100 Power Receiver Mold speaker 100 Power Receiver Mold speaker 105 MicroShell 60 MicroShell 85 ...
Page 4
Intended use Intended use The hearing aid is intended to amplify and transmit sound to the ear. Indications for use Bilateral or unilateral impaired hearing of sensorineu- ral, conductive or mixed type ranging from a slight (16 dB HL*) to profound (95 dB HL*) degree of hearing loss, with an individual frequency configuration.
Page 5
IMPORTANT NOTICE The hearing aid amplification is uniquely adjusted and optimized to your personal hearing capabilities during the hearing aid fitting performed by your hearing care professional.
Table of contents About Your hearing aid, speaker and earpiece Getting started Charging time Battery performance Turn hearing aids ON/OFF using the charger Turn hearing aids ON/OFF using the push button Turn hearing aids ON/OFF Low battery indication Identify left and right hearing aid How to replace the disposable battery —...
Page 7
Daily use Cleaning Replace standard earpieces ProWax miniFit filter Clean customized earpieces Replace ProWax filter Hearing aid storage Change volume Change program Options Flight mode Mute your hearing aids Using your hearing aid with iPhone, iPad and iPod Using your hearing aid with Android devices Pairing and compatibility Wireless accessories and other options Continues on next page...
Page 8
Table of contents Warnings General warnings More info Troubleshooting Water & dust resistant (IP68) Conditions of use Technical information Mobile phone Warranty Your individual hearing aid settings Sound and LED light indicators Technical Data...
Your hearing aid, speaker and earpiece Microphones Ear grip Sound in Holds the speaker in place (optional) LED light Wax filter Protects the speaker from earwax damage Push button Upper part Lower part Nail grip Speaker (MNR T only) Sound out Opens the battery drawer Rechargeable battery...
Page 10
The hearing aid uses one of the following speakers: Standard speakers Power Receiver Mold MicroShell About Getting started Daily use Options Warnings More info...
Page 11
The speakers use one of the following earpieces: Standard earpieces Customized earpieces OpenBass dome LiteTip Bass dome, double vent MicroMold VarioTherm® LiteTip Power dome VarioTherm® MicroMold Grip Tip VarioTherm® is a registered trade- Available in small and large, left mark of Dreve.
Charging time (MNR T R only) Ensure you fully charge your hearing aids before first use and charge them every night. This ensures you start your day with fully charged hearing aids. If your hearing aid’s battery is completely drained, the normal charging time is: 3 hours 1 hour...
Battery performance In both MNR T R and MNR T, battery performance varies depending on your individual use and hearing aid settings. Streaming sound from a TV, mobile phone or connectivity devices can influence this performance. Rechargeable battery — MNR T R only If your rechargeable hearing aids do not perform for a full day, you may need to have the rechargeable battery replaced.
Turn hearing aids ON/OFF using the charger (MNR T R only) Your hearing aids automatically turn ON when removed from the charger. The hearing aid LED light turns GREEN after approximately two seconds. Wait until the hearing aid LED light blinks GREEN twice, confirming that it is ready for use.
Turn hearing aids ON/OFF using the push button (MNR T R only) The hearing aids can be turned ON/OFF using the push button. To turn ON Press and hold the lower part of the push button for approximately two seconds until the hearing aid LED light turns GREEN.
Turn hearing aids ON/OFF (MNR T only) The battery drawer is used to turn the hearing aids ON and OFF. To save battery life, make sure your hearing aids are switched OFF when you are not wearing them. To perform a quick reset of hearing aid settings, open and close the battery drawer.
Low battery indication Just before the battery runs out completely, you will hear four descending tones. To extend battery performance, ensure you stop any audio streaming. MNR T R: When the battery is running low, you will hear three alternate beeps.
Identify left and right hearing aid (MNR T R only) It is important to distinguish between the left and the right hearing aid, as they may be programmed differently. You can find left/right color indicators on the hearing aid itself and on 60 and 85 speakers as shown.
Page 19
Identify left and right hearing aid (MNR T only) It is important to distinguish between the left and the right hearing aid, as they may be programmed differently. You can find left/right color indicators on the hearing aid itself and on 60 and 85 speakers as shown. Indicator markings (either L or R) can also be found on 100 speakers and some earpieces.
How to replace the disposable battery — size 312 (MNR T only) 1. Remove 2. Uncover 3. Insert Fully open the Remove the sticky Insert the new bat- battery drawer and label from the + tery into the battery remove the battery. side of the new drawer with the + battery.
Page 21
4. Close MultiTool Close the battery drawer. You can use the MultiTool to You may hear a jingle through change the battery. Use the mag- the earpiece. netic end to remove and insert batteries. The MultiTool is provided by your hearing care professional.
Put on hearing aid Step 1 Step 2 Step 3 Ear grip (optional) Place the hearing Hold the bend of Gently push the ear- aid behind your ear. the speaker wire be- piece into your ear tween your thumb canal until the speak- You should always and index finger.
Cleaning The MultiTool contains a brush and wire loop for cleaning and removing earwax. If you need a new MultiTool, contact your hearing care professional. Magnet Wire loop Brush IMPORTANT NOTICE The MultiTool has a built-in magnet. Keep the MultiTool at least 30 centimeters (1 foot) away from credit cards and other magnetically-sensitive devices.
Page 25
Do not drop your hearing aid when handling it. Ensure you hold it over a soft surface to avoid damage while cleaning it. Clean the microphone openings Use the MultiTool brush to carefully Microphone openings brush debris away from the openings and the surface around the openings.
Replace standard earpieces It is important that you do not clean the standard earpiece (dome and Grip Tip). If the earpiece is filled with earwax, replace it with a new one. Grip Tip needs to be replaced at least once a month. Step 1 Step 2 Step 3...
Page 27
IMPORTANT NOTICE If the earpiece is not on the speaker when removed from the ear, the earpiece may still be in the ear canal. For further instructions, consult your hearing care professional.
ProWax miniFit filter The speaker has a white wax filter attached to the end where the earpiece is attached. The wax filter keeps earwax and debris from ProWax filter damaging the speaker. Ensure you replace the filter when clogged, or if the hearing aid does not sound normal.
Page 29
Replace ProWax miniFit filter 1. Tool 2. Remove 3. Insert filter filter Final filter Empty Remove the tool Insert the empty pin Insert the new wax from the shell. The into the wax filter in filter using the other tool has two pins, the speaker and pull pin, remove the tool, one empty for re-...
Clean customized earpieces Ensure that you regularly clean the Clean the vent by inserting the brush earpiece. through the hole, twisting it slightly. The earpiece has a white wax filter* that keeps earwax and debris from damaging the speaker. Ensure you replace the filter when clogged, or if the hearing aid does not sound normal.
Replace ProWax filter 1. Tool 2. Remove 3. Insert filter filter Old filter Remove the tool Insert the empty pin Insert the new wax from the shell. The into the wax filter filter using the other tool has two pins, in the earpiece and pin, remove the tool, one empty for re-...
Hearing aid storage (MNR T R only) When you are not using your hearing aid, the charger is the best place to keep it. To ensure the longest life of the rechargeable battery in the hearing aid, do not expose it to excessive heat. For example, do not leave the hearing aid in the sun in front of a window or in a car, even if the hearing aid is in the charger.
Change volume The push button lets you adjust the volume. When you increase or decrease the volume, you hear a beep. For information regarding button press times, see the table General settings overview for your hearing aid, in the Your individual hearing aid settings section at the end of this booklet.
Change program Your hearing aid can save up to four different programs configured by your hearing care professional. Depending on the program you choose (1,2,3 or 4), you hear one to four tones. For information regarding tones, see the Sound and LED light indicators section.
Page 35
Press the push button to switch between pro- grams. The program cycle switches one program forward when the upper part of the push button is pressed, for example program 1 to 2 or program 4 to 1. If the lower part of the push button is pressed, the program cycle goes backward, for example 2 to 1 or program 1 to 4.
Flight mode When Flight mode is activated, Bluetooth connectivity is turned OFF. However, the hearing aid is still turned ON and functioning. Be aware that pressing the push button on one hearing aid, activates Flight mode on both hearing aids. For more information about sounds and lights, see the Sound and LED light indicators section.
Page 37
MNR T To activate and deactivate Press and hold either end of the push button for at least seven seconds. A jingle and an LED light pattern confirm your action. Opening and closing the battery drawer also deactivates Flight mode.
Mute your hearing aids In both MNR T and MNR T R you can mute your hearing aids by using one of the following optional devices/app: • HearLink App • AudioClip • Remote Control How to unmute your hearing aids You can unmute your hearing aids by using one of the optional devices/app or by applying a short press to the upper or lower part of the push button on the hearing aids.
Page 39
IMPORTANT NOTICE Do not use the mute function as an OFF button, as the hearing aid is still using battery power in this mode.
Using your hearing aid with iPhone, iPad and iPod Philips HearLink is a Made for iPhone® hearing aid and allows for direct communication and control with iPhone, iPad® or iPod touch®.* For assistance in using your hearing aid with any of these products, contact your hearing care professional.
Using your hearing aid with Android devices Philips HearLink supports Audio Streaming for Hearing Aids (ASHA) and can be used for direct communication and control with selected Android™ devices.* For assistance in using your hearing aid with Android devices, contact your hearing care professional.
Pairing and compatibility For instructions on how to pair your hearing aid with iPhone, iPad, iPod touch or Android devices, visit the following website: hearingsolutions.philips.com/support/how-to-faq/ For a list of compatible iPhone, iPad, iPod touch and Android devices, visit the following website: hearingsolutions.philips.com/compatibility...
Wireless accessories and other options There are a range of accessories available as an enhancement to your wireless hearing aid. These enable you to hear and communicate better in everyday situations. AudioClip Remote Control A device that can be used A device that lets you change as remote microphone and program, adjust volume, or...
Page 44
HearLink App Telecoil An application that lets you Telecoil can help you control your hearing aid from hear better when using your mobile phone or tablet. a phone with a built- For iPhone, iPad, iPod touch, in loop or when in buildings and Android devices.
General warnings For your personal safety and to ensure (MNR T R only) correct usage, you should familiarize Only charge the hearing aids with a yourself fully with the following general designated charger. Other chargers warnings before using your hearing aid. risk destroying the hearing aids and Consult your hearing care professional if batteries.
Page 47
Usage of hearing aids (MNR T only) Hearing aids should be used only as Batteries have occasionally been directed and adjusted by your hearing mistaken for pills. Therefore, check care professional. Misuse can result in your medicine carefully before sudden and permanent hearing loss. swallowing any pills.
Page 48
General warnings Explosives Fatality hazards and risk of swallowing (MNR T R only) lithium-ion batteries or placing them in the ear or nose The hearing aid is safe to use under (MNR T R only) normal usage conditions. The hearing aid has not been tested for compliance Never swallow lithium-ion batteries with international standards...
Page 49
Rechargeable battery The safety of recharging batteries (MNR T R only) using a USB connector is determined by the external signal source. When Do not attempt to open the hearing connected to external equipment aid, as it may damage the battery. plugged into a power socket, this Never attempt to replace the battery.
Page 50
General warnings Active implants The Autophone magnet or MultiTool The hearing aid has been thoroughly (which has a built-in magnet) should tested and characterized for human be kept more than 30 centimeters health according to international (1 foot) away from the implant, e.g. do standards for human exposure (Specific not carry it in your breast pocket.
Page 51
Detached earpiece in ear canal The chemicals in cosmetics, hairspray, If the earpiece is not on the speaker perfume, aftershave lotion, sunscreen when removed from the ear, the lotion, and insect repellent can earpiece may still be in the ear canal. damage the hearing aid.
Page 52
General warnings Power hearing aid Use on aircraft Special care should be exercised in Your hearing aids have Bluetooth selecting, fitting, and using hearing wireless technology. On board an aids where the maximum sound aircraft, the hearing aids must be pressure capability exceeds 132 dB SPL put into Flight mode to deactivate (IEC 60138-4/IEC 711) as there may...
Page 53
Interference The hearing aids have been thoroughly tested for interference according to the most stringent international standards. Electromagnetic interference may occur in the vicinity of equipment with the symbol to the left. Portable and mobile RF (radio frequency) communications equipment can affect the performance of your hearing aids.
Page 54
General warnings Warnings to hearing care professional (iii) History of sudden or rapidly A hearing care professional should progressive hearing loss within advise a prospective hearing aid user the previous 90 days. to consult immediately with a licensed (iv) Acute or chronic dizziness. physician (preferably an ear specialist) Unilateral hearing loss of sudden before dispensing a hearing aid if the...
Page 55
Special care should be exercised in is purchased. Following the medical selecting and fitting a hearing aid evaluation, the physician will give you whose maximum sound pressure a written statement that states that capability exceeds 132 dB SPL as there your hearing loss has been medically may be risk of impairing the remaining evaluated and that you may be...
Page 56
General warnings The audiologist or hearing care Federal law permits a fully informed professional will conduct a hearing adult to sign a waiver statement aid evaluation to assess your ability declining the medical evaluation to hear with and without a hearing for religious or personal beliefs that aid.
Page 57
Children with hearing loss In addition to seeing a physician for medical evaluation, a child with a hearing loss should be directed to an audiologist for evaluation and rehabilitation, since hearing loss may cause problems in language development and educational and social growth of a child.
Troubleshooting MNR T R/T Symptom Possible causes Hearing aid is out of power Dead battery No sound Clogged earpieces (dome, Grip Tip, or mold) Hearing aid microphone muted Clogged sound outlet Intermittent or reduced sound Moisture Hearing aid earpiece incorrectly inserted Squealing noise Earwax accumulated in ear canal If your hearing aid plays eight beeps, four times consecutively,...
Page 59
2) On your phone, turn Bluetooth OFF and ON again 3) Turn the hearing aid OFF and then turn it back ON 4) Pair your hearing aid again (For guidance, visit: hearingsolutions.philips.com/support/how-to-faq/) Note Continues on next page If none of the above solutions work, consult your hearing care professional.
Page 60
Troubleshooting MNR T R only — To troubleshoot the charger, see your charger’s instructions for use. Symptom Possible causes The charger is not turned ON The hearing aid or charger’s temperature is either too warm The hearing aid LED light remains turned or too cold OFF when the hearing Charging is incomplete.
Page 61
Solutions Verify that the charger’s power plug is correctly connected or the power bank has enough battery Move the charger and hearing aid to a location with a temperature between +5°C and +40°C (+41°F and +104°F) Reinsert the hearing aid into the charger. This completes the charging within approximately 15 minutes.
Water & dust resistant (IP68) Your hearing aid is dust tight and Should your hearing aid come into protected against ingress of water, contact with water and stop working, which means it is designed to be gently wipe off any water and let the worn hearing aid dry.
Conditions of use (MNR T R only) Operating Temperature: +5°C to +40°C (41°F to 104°F) conditions Humidity: 5% to 93% relative humidity, non-condensing Atmospheric pressure: 700 hPa to 1060 hPa Charging Temperature: +5°C to +40°C (41°F to 104°F) conditions Humidity: 5% to 93% relative humidity, non-condensing Atmospheric pressure: 700 hPa to 1060 hPa Transportation Temperature and humidity shall not exceed the below...
Page 64
Conditions of use (MNR T only) Operating Temperature: +1°C to +40°C (34°F to 104°F) conditions Humidity: 5% to 93% relative humidity, non-condensing Atmospheric pressure: 700 hPa to 1060 hPa Transportation Temperature and humidity shall not exceed the below and storage limits for extended periods during transportation and conditions storage:...
Technical information The hearing aid contains the (MNR T R) following two radio technologies: The radio transmitter is weak and always below 9 mW equal to The hearing aid contains a radio 9.6 dBm in total radiated power. transceiver using short range magnetic induction technology (MNR T only) operating at 3.84 MHz.
Page 67
Additional information can device meets FCC and Innovation, be found in the Specification Guide Science and Economic Development on hearingsolutions.philips.com Canada’s RF exposure limits and has been tested while in contact USA and Canada with the human body. Use of...
Page 68
Note: The device complies with Part 15 of the FCC rules and with Innovation, Science and Economic Develop- ment Canada’s license -exempt RSSs standards. Operation is subject to the follow- ing two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including in- terference that may cause undesired...
Page 69
Note: — Reorient or relocate the receiving This equipment has been tested and antenna. found to comply with the limits for — Increase the separation between a Class B digital device, pursuant the equipment and receiver. to part 15 of the FCC Rules. These —...
Mobile phone Some hearing aid users have report- Whereas all hearing aids have acous- ed a buzzing sound in their hearing tic coupling, only the larger hearing aid when they are using cell phones, aids have the physical space for indicating that the cell phone and telecoil (inductive) coupling.
Page 71
performance and the telecoil coupling The immunity of this hearing aid indicating normal use. is at least M2/T2. The equipment The above equipment performance performance measurements, measurements, categories and sys- categories and system classifications tem classifications are based upon are based upon the best information the best information available, but it available but cannot guarantee that cannot be guaranteed that all users...
Page 72
Directive SBO Hearing A/S 2014/53/EU. Kongebakken 9 DK-2765 Smørum This medical device complies with Denmark Medical Device Regulation (EU) hearingsolutions.philips.com/doc 2017/745. Waste from elec- tronic equipment SBO Hearing A/S must be handled according to local regulations. Denmark...
Page 73
Description of symbols and abbreviations used in this booklet Warnings Text marked with a warning symbol must be read before using the device. Manufacturer The device is produced by the manufacturer whose name and address are stated next to the symbol. Indicates the medical device manufacturer, as defined in EU Regulations 2017/745.
Page 74
REACH Declaration REACH requires Philips Hearing Solutions to provide chemical content infor- mation for Substances of Very High Concern (SVHC) if they are present above 0.1% of the article weight. Recent information can be found on the website hearingsolutions.philips.com/REACH.
Page 75
Description of additional symbols used on labels Keep dry Indicates a medical device that needs to be protected from moisture. Caution symbol Consult instructions for use for warnings and cautions. Catalog number Indicates the manufacturer’s catalog number so that the medical device can be identified.
Page 76
International warranty Your device is covered by an legislation governing the sale of international warranty issued by consumer goods in the country the manufacturer. This international where you have bought your device. warranty covers manufacturing Your hearing care professional and material defects in the device may also have issued a warranty itself, but not in accessories such that goes beyond the clauses...
Your individual hearing aid settings To be filled out by your hearing care professional. General settings overview for your hearing aid Left Right Change volume Short press Short press Change program Long press Long press Volume control indicators Beeps at min/max volume Beeps when changing volume Beeps at start-up volume Battery indicators...
Sound and LED light indicators Different sounds and light patterns indicate the hearing aid status. The different indicators are listed on the following pages. For light indicators on your charger, see the charger’s instructions for use. Your hearing care professional can set sound and LED light indicators to match your preferences.
Page 80
ON/OFF Sound LED light LED light comments Jingle 4 descending (MNR T R only) tones Volume Sound LED light Shown once Start-up volume 2 beeps Minimum/maximum 3 beeps volume Volume up/down 1 beep Continuous or repeated Mute three times Long GREEN blink Short GREEN blink Long ORANGE blink Short ORANGE blink...
Page 81
Accessories Sound LED light LED light comments TV Adapter 2 different tones AudioClip 2 different tones Continuous or repeated three Flight mode Sound LED light times Flight mode 4 descending activated tones + short (MNR T R only) jingle Flight mode 4 descending deactivated tones + short...
Page 82
Warnings Sound LED light LED light comments Continuously 3 alternate Low battery blinking beeps 4 descending Battery shut down tones 8 beeps Microphone service Repeated four repeated 4 check needed times times The hearing aid LED light does not turn ON when See the Troubleshooting the hearing aid...
Page 83
Warnings Sound LED light LED light comments The hearing aid LED light blinks Continuously ORANGE when blinking. See the the hearing aid Troubleshooting is placed in the section. charger (MNR T R only) The hearing aid LED light blinks GREEN Continuously when the hearing blinking.
Technical Data miniRITE T R Philips Philips HearLink HearLink 0 dB SPL ref. 20 mPa 9030 7030 & 5030 Measured Peak OSPL90 106 dB SPL 106 dB SPL according to American HF Average OSPL90 103 dB SPL 103 dB SPL...
Page 85
Philips HearLink 9030 Philips HearLink 7030 & 5030 OSPL90 – Output Sound Pressure Level Input: 90 dB SPL. Technical setting: Full-on Gain Input: 50 dB SPL. Technical setting:...
Page 86
Technical Data miniRITE T R Philips Philips HearLink HearLink 0 dB SPL ref. 20 mPa 9030 7030 & 5030 Measured Peak OSPL90 117 dB SPL 117 dB SPL according to American HF Average OSPL90 114 dB SPL 114 dB SPL...
Page 87
Philips HearLink 9030 Philips HearLink 7030 & 5030 OSPL90 – Output Sound Pressure Level Input: 90 dB SPL. Technical setting: Full-on Gain Input: 50 dB SPL. Technical setting:...
Page 88
Technical Data miniRITE T R Philips Philips HearLink HearLink 0 dB SPL ref. 20 mPa 9030 7030 & 5030 Measured Peak OSPL90 124 dB SPL 124 dB SPL according to American HF Average OSPL90 120 dB SPL 120 dB SPL...
Page 89
Philips HearLink 9030 Philips HearLink 7030 & 5030 OSPL90 – Output Sound Pressure Level Input: 90 dB SPL. Technical setting: Full-on Gain Input: 50 dB SPL. Technical setting:...
Page 90
Technical Data miniRITE T R Philips Philips HearLink HearLink 0 dB SPL ref. 20 mPa 9030 7030 & 5030 Measured Peak OSPL90 127 dB SPL 127 dB SPL according to American HF Average OSPL90 123 dB SPL 123 dB SPL...
Page 91
Philips HearLink 9030 Philips HearLink 7030 & 5030 OSPL90 – Output Sound Pressure Level Input: 90 dB SPL. Technical setting: Full-on Gain Input: 50 dB SPL. Technical setting:...
Page 92
Technical Data miniRITE T Philips Philips HearLink HearLink 0 dB SPL ref. 20 mPa 9030 7030 & 5030 Measured Peak OSPL90 105 dB SPL 105 dB SPL according to American HF Average OSPL90 103 dB SPL 103 dB SPL National...
Page 93
Philips HearLink 9030 Philips HearLink 7030 & 5030 OSPL90 – Output Sound Pressure Level Input: 90 dB SPL. Technical setting: Full-on Gain Input: 50 dB SPL. Technical setting:...
Page 94
Technical Data miniRITE T Philips Philips HearLink HearLink 0 dB SPL ref. 20 mPa 9030 7030 & 5030 Measured Peak OSPL90 117 dB SPL 117 dB SPL according to American HF Average OSPL90 114 dB SPL 114 dB SPL National Standard...
Page 95
Philips HearLink 9030 Philips HearLink 7030 & 5030 OSPL90 – Output Sound Pressure Level Input: 90 dB SPL. Technical setting: Full-on Gain Input: 50 dB SPL. Technical setting:...
Page 96
Technical Data miniRITE T Philips Philips HearLink HearLink 0 dB SPL ref. 20 mPa 9030 7030 & 5030 Measured Peak OSPL90 123 dB SPL 123 dB SPL according to American HF Average OSPL90 119 dB SPL 119 dB SPL National Standard...
Page 97
Philips HearLink 9030 Philips HearLink 7030 & 5030 OSPL90 – Output dB SPL Sound Pressure Level Input: 90 dB SPL. Technical setting: 1000 2000 5000 10000 Full-on Gain Input: 50 dB SPL. Technical setting:...
Page 98
Technical Data miniRITE T Philips Philips HearLink HearLink 0 dB SPL ref. 20 mPa 9030 7030 & 5030 Measured Peak OSPL90 127 dB SPL 127 dB SPL according to American HF Average OSPL90 123 dB SPL 123 dB SPL National Standard...
Page 99
Philips HearLink 9030 Philips HearLink 7030 & 5030 dB SPL OSPL90 – Output Sound Pressure Level Input: 90 dB SPL. Technical setting: 1000 2000 5000 10000 Full-on Gain Input: 50 dB SPL. Technical setting: 1000 2000 5000 10000...
Page 100
Philips and the Philips Shield Emblem are registered trademarks of Koninklijke Philips N.V. and are used under license. This product has been manufactured by or for and is sold under the responsibility of SBO Hearing A/S, and SBO Hearing A/S is the warrantor in relation to this product.
Need help?
Do you have a question about the HearLink miniRITE T and is the answer not in the manual?
Questions and answers
Hearing aids HearLink 9030 not connecting to iphone
Philips HearLink miniRITE T hearing aids may not be connecting to an iPhone due to several possible reasons:
1. Compatibility Issues – The iPhone model may not be compatible with the hearing aids. A list of compatible devices is available at hearingsolutions.philips.com/compatibility.
2. Wireless Interference – Other wireless devices or networks may be affecting the connection, as the hearing aids operate on the 2.4 GHz ISM band.
3. Device Settings – The hearing aids may not be properly paired with the iPhone, or Bluetooth may be disabled.
4. Battery Issues – The hearing aid battery may be low or depleted.
5. Unverified Accessories – Using accessories not certified by the manufacturer may affect performance.
For troubleshooting, users should check compatibility, ensure Bluetooth is enabled, re-pair the devices, and verify battery levels. If issues persist, consulting a hearing care professional is recommended.
This answer is automatically generated