Manuale di Installazione e uso EVOSTYLE N.G.
Mod. MQI006-0
BEST-X LINE
INSTALLATION AND USER MANUAL
Mod. MQI006-0 Doc. HBE220-1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BEST-X Series and is the answer not in the manual?

Questions and answers

Summary of Contents for New Life BEST-X Series

  • Page 1 Doc. HBI070-2 Manuale Manuale di Installazione e uso EVOSTYLE N.G. Mod. MQI006-0 BEST-X LINE INSTALLATION AND USER MANUAL Mod. MQI006-0 Doc. HBE220-1...
  • Page 2: Table Of Contents

    Doc. HBE220-1 MANUAL Installation and User Manual BEST-X-DC WITH Mod. MQI006-0 TIM-X INDEX GENERAL ASPECTS ..........................1 Introduction ............................1 Descriptions ............................1 Technical data ............................5 Other Data ............................6 INSTALLATION AND USE ........................6 Use Conditions ............................6 General symbols ............................7 Symbols for transportation and storage....................
  • Page 3 Doc. HBE220-1 MANUAL Installation and User Manual BEST-X-DC WITH Mod. MQI006-0 TIM-X The Monoblock Does Not Emit X-Rays ..................51 Procedures for the filament current measures: .................54 Measurement of the filament current ..................54 Plates positioning ..........................57 Servicing of the x-ray unit ......................62 Extraordinary maintenance ......................63 Thermal Curves .............................
  • Page 4: General Aspects

    (conventional films, phosphor plates and videoradiografie). According to Directive 2007/47/CE (Legislative Decree 37/2010) it is classified in CLASS IIb. NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 5 (mm 400-800-1100), has a maximum extension which varies between 1730 mm and 2430, depending on the extension cord used. NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 6 The maximum time that can be set is 1 s (the display will read 1000) while the minimum time is 20 ms. (the display will read 0020); NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 7 Drawings, schematics, components lists, instructions for repairs: New Life Radiology is committed to providing, upon request, drawings, circuit diagrams, component parts lists, instructions, or other information that can serve for qualified technical personnel, to perform the repair of any parts that can be repaired.
  • Page 8: Technical Data

    Voltage accuracy ± 5 % Current accuracy ± 3 % Time accuracy ± 5 % Dose accuracy ± 5 % NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 9: Other Data

    Use conditions: Equipment for continuous operation with intermittent charge Installation: Permanent e mobile NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 10: General Symbols

    Red point, placed on the cover of the unit, indicate focal spot Alternating current Radiation emission symbol Follow annex instructions Symbol in conformity with European Directive 2002/96/Ec (Weee). NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 11 Equipment Serial Number. To be used for all communication with manufacturer / service technician TUBE X-ray Tube Serial number NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 12: Symbols For Transportation And Storage

    Doc. HBE220-1 Manual Installation and User Manual BEST-X-DC WITH Mod. MQI006-0 TIM-X Symbols for transportation and storage 27 kg NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 13: Information For Installation

    3 x 1.5 sq mm, shall be provided between the control unit and the support. Please note that the radiological groups are supplied without a plug NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 14 The connection to the external lamps must be done by connecting the output LAMP indicated with the letter A in the figure above. The lamp must be of value 30W 230-240 Vac. NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 15: Installation

    The system in wall version consists of the components shown in the following legend: wall plate extension double pantograph arm palmtop monoblock NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 16 Installation and User Manual BEST-X-DC WITH Mod. MQI006-0 TIM-X Quotas with relative useful space – EXTENSION FROM mm 400 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 17 Installation and User Manual BEST-X-DC WITH Mod. MQI006-0 TIM-X Quotas with relative useful space – EXTENSION FROM mm 800 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 18 Installation and User Manual BEST-X-DC WITH Mod. MQI006-0 TIM-X Quotas with relative useful space – EXTENSION FROM mm 1100 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 19  alternately, if the wall is not perfectly flat put adequate thicknesses so as not to deform the wall plate. NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 20 Extract the SEGER from the extension , and enter it in the wall plate as shown  in fig. 3 fig. 3 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 21 Replace the SEGER, unhook the front cover and remove the lower strip cover strips  (fig. 4) fig. 4 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 22 Connect the cable from the double pantograph arm to output X-RAY TUBE and the ground wire IN in its slot. fig. 5 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 23 6 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 24 Unscrew the screw from the front tubular carrier of the double arm pantograph  by lifting the cylinder and remove the half-moon(fig. 7) fig. 7 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 25 8 Reinsert the half-moon in its slot. Lower the cylinder, securing the double arm pantograph with the screw. NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 26 Doc. HBE220-1 Manual Installation and User Manual BEST-X-DC WITH Mod. MQI006-0 TIM-X fig. 9 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 27: Settings

    After making the adjustments necessary to close the wall plate with the respective cover, making sure to replace the lower back lining strip and the front cover of the extension. NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 28: Springs Adjustment Of The Double Pantograph Arm

    This procedure is identical for the section B in which the adjusting key must be  inserted in the hole 2. fig.11 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 29: Instructions For Installation In Column Mode

    The system in column mode consists of the components shown in the following legend: Detail cart vertical support monoblock double pantograph arm fuse box palmtop NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 30 Doc. HBE220-1 Manual Installation and User Manual BEST-X-DC WITH Mod. MQI006-0 TIM-X Quotas with appropriate useful space Fig. 12 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 31 4  screws 6x20 as shown in fig. 14 fig. 14 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 32 Remove the screws holding the card fuse holder to the vertical support  shown in Fig. 16 fig. 16 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 33 Mod. MQI006-0 TIM-X Insert the double pantograph arm in the vertical support as shown in fig. 17  fig. 17 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 34 Connect the cable coming from the double pantograph arm to the switch  board fuse holder as shown in figure 18 fig. 18 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 35: Palmtop Operation

    The maximum time that can be set is 1 s (the display will read 1000) while the  minimum time is 20 ms. (the display will read 0020); NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 36 Type of tooth selection Manual time decrease Menu Pre-Start mA Selection kV Selection Type of film selection Type of patient selection NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 37 It allows the manual decrease of the exposure time; Time schedule It allows customizing the time schedule; Battery level It shows the battery charge level; NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 38 It allows turning on/turning off the hand-held or put it on standby; Reset button It allows resetting the hand-held; Micro USB slot for recharging; NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 39: Operation

    Environmental working conditions from 0 to 40 C°. After possibly plugging in the battery, push and release the on/off button, the display will light up and show the “New Life Radiology” logo, the software version and, ultimately, the exposure parameters screen.
  • Page 40 In order to access another function push the Menu button where as to start the exposition just select the X-ray icon (Pre-Start). NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 41 To exit, in any case, it is sufficient to press the Cancel button. To carry out an X-ray, you need to keep the X-Ray button pressed until the cooling process screen appears. NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 42 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 43 In the above screen you can press the KV button and have a visual of the voltage (yellow line) and/or the mA button just to have a visual of the electrical current (blue line). NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 44 From here you can go back to the configuration screen pushing Radiographic or you can access various functions: - Parameters: The screen parameter is accessible by New Life Radiology personnel only - Matching: to connect the hand-held to a receiver - Disp.Options: to modify some functions of the display and to access the self-calibration...
  • Page 45: Matching Screen

    To go back to the menu, press the Menu button. NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 46: Display Options Screen

    From now on you can also access the self-calibration of the radiographic device pressing the Calib. button. NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 47: Calibration Screen

    Such result will be shown in the text line placed in the bottom. To go back to the Menu, press Back. NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 48: Functioning E Time Scale

    320 [ms] 3.26 [mGy] 2.39 [mGy] 1.74 [mGy] 480 [ms] 360 [ms] 240 [ms] 2.61 [mGy] 1.95 [mGy] 1.30 [mGy] NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 49 40 [ms] 0.57 [mGy] 0.38 [mGy] 0.38 [mGy] 60 [ms] 40 [ms] 40 [ms] 0.57 [mGy] 0.38 [mGy] 0.38 [mGy] NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 50 180 [ms] 0.68 [mGy] 0.68 [mGy] 0.68 [mGy] 180 [ms] 180 [ms] 180 [ms] 0.68 [mGy] 0.68 [mGy] 0.68 [mGy] NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 51 120 [ms] 1.34 [mGy] 1.00 [mGy] 0.80 [mGy] 180 [ms] 100 [ms] 100 [ms] 1.20 [mGy] 0.67 [mGy] 0.67 [mGy] NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 52 When using the appliance we recommend the use of protective devices such as collars or aprons. NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 53 The programmed exposure Press the Exit button and time has not been successfully repeat the exposition Exposition interrupted accomplished procedure NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 54: The Monoblock Does Not Emit X-Rays

    If the monoblock does not emit X-rays, as first check the continuity of the THERMOSWITCH; if this test is negative, you must bypass it as shown in fig.19 fig. 19 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 55 A and B indicated in fig 20b. fig.20a (wall version) NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 56 Doc. HBE220-1 Manual Installation and User Manual BEST-X-DC WITH Mod. MQI006-0 TIM-X fig.20b (column version) NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 57: Procedures For The Filament Current Measures

    After removing the back cover, making reference to fig.21, connect oscilloscope / tester  terminals across the resistor to 1KΩ. fig. 21 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 58 7 V (to each V corresponds 1 mA). As if 4 mA are set on timer the average value that is expected to read is about 4 V. NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 59 It checks the performance of the X-ray tube is in compliance with the project  specifications; The tests are stored in digital format and made available for those who request them;  NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 60: Plates Positioning

    Installation and User Manual BEST-X-DC WITH Mod. MQI006-0 TIM-X Plates positioning Below, in fig.23 it shows the layout plates positioning on BEST-X-DC  fig.23 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 61 Doc. HBE220-1 Manual Installation and User Manual BEST-X-DC WITH Mod. MQI006-0 TIM-X (“A” wall version) (“A” column version) NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 62 4. S/N XX.XY.XYZ (It represents the device serial number) XX It indicates the year; XY It indicates the month; XYZ indicates the serial sequential number; NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 63 Doc. HBE220-1 Manual Installation and User Manual BEST-X-DC WITH Mod. MQI006-0 TIM-X (“B”) (“C”) NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 64 1. S/N XX. T XYZ (It indicates the timer’s serial number) XX It indicates the year of manufacture T XYZ It is the palmtop’s progressive serial number NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 65: Servicing Of The X-Ray Unit

    NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 66: Extraordinary Maintenance

    The extraordinary maintenance interventions are not carried out by the operator Extraordinary maintenance CARRIED OUT BY THE NEW LIFE RADIOLOGY SRL TECHNICAL ASSISTANCE IN CASE OF NEED AND IN ANY CASE AFTER RELEVANT REPAIRS . We provide a guideline on controls, on minimum checks that are to be performed in any technical intervention on the plant: monoblock, support and pantograph stability;...
  • Page 67 The accessories and connection cables must be disinfected only with a cloth soaked  in disinfectant solution. Do not use solvents or corrosive disinfectants. NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 68 In case of monoblock's damage for impact or crushing with a consequent oil  leaking, prevent dispersal into the environment, proceeding to decommissioning as indicated above. NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 69 Output voltage 5V DC Max output current 1.2A Type of output connector Micro USB Power Power voltage 90÷264 Vac NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 70 Beam limiting devices technical description distance FFD 200 mm max 60 mm  Limiter beam (cone collimator) coated in lead: mod. C.L.01 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 71 Focal spot position on the reference axis: See detailed figures Focal spot value for the reference axis: 0,3 mm X-ray tube: fig.24 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 72: Thermal Curves

    Installation and User Manual BEST-X-DC WITH Mod. MQI006-0 TIM-X Reference curves Thermal Curves Emission and filament characteristics DC Rating Charts fig.25 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 73: Safety Aspects

    2 meters and out of the path of the x-ray beam, in order to avoid the exposition to the stray radiation. NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it...
  • Page 74: Possible Drawbacks Found In The Radiographs Results

    Film accidental exposure to daylight or safety lamp darkroom no longer suitable Black line occurrence on X-rays Possible causes: NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 75: Recommendations

    During normal operation, the film must be positioned at 90 ° with respect to the cone collimator, as shown in fig.26 Upper jaw Lower jaw Fig.26 NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
  • Page 76 Although every effort is made to ensure accuracy information contained this manual, the New Life Radiology does not assume responsibility for mistakes, omissions inaccuracies. NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO) Italy phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...