Tan bead nucleic acid extractor (non-sterile) (54 pages)
Summary of Contents for TANBead Maelstrom 9610
Page 1
(Compatible with power 100V-120V only) Maelstrom 9610 TAN Bead Nucleic Acid Extractor User Manual (Non-Sterile) 039.M1602.X01Rev V1.0_2021.06.11...
Table of Contents Introduction ..................6 Instrument Overview ..............11 Installation & Get Started ............15 On-screen Display & How to Use ........16 Technical Support ................38 Cleaning and Maintenance ............39 Disposal ....................39 Patents ....................40 About Manufacturer ..............40 Appendix ..................41...
Page 3
About Manual The label on the instrument, the User Manual, and other packaging material may contain following symbols: Catalog number Non-Sterile Serial number Temp Limit (Operating) Specification of fuse Biologic Risks Manufacturer The date of manufacture An administrative marking that indicates conformity with health, safety, and environmental protection standards for products sold within the European Economic Area (EEA) In Vitro Diagnostic Medical Device...
Page 4
About Instrument WARNING Use a power cord that meets your country’s standard. In case of any questions about the power cord, contact your local distributor or reseller. Do not use the instrument with voltage other than the voltage specified on the device.
Page 5
About Instrument CAUTION Never attempt to remodel the instrument without the permission from the manufacturer. Doing so may lead to a fire or electric shock. Do not bump or knock the instrument. Doing so may cause malfunctions. Repairs to the instrument must be performed by agencies authorized by Taiwan Advanced Nanotech Inc.
Purified nucleic acids can be obtained after cell lysis, nucleic acids absorption, wash and elution. Intended Use Maelstrom 9610 shall be used in combination with TANBead Nucleic Acid Extraction Kit to extract DNA/RNA from specimen. Maelstrom 9610 can handle magnetic beads for extraction and purification of nucleic acid.
Page 7
9610 in a location that meets the following criteria: Choose a location that has good air circulation. Place Maelstrom 9610 on the table capable of bearing at least the weight of Maelstrom 9610 (≒ 95 kg). Do not use Maelstrom 9610 in a location where there is large temperature variability or high humidity.
Page 8
1. Introduction Safety Instructions and Guidelines This device can be used with potentially biohazardous materials. Use appropriate personal protective equipment (gloves, safety goggles, lab coat, etc.) for handling and disposing of biohazardous materials. Under normal condition, sound pressure level from Maelstrom 9610 does not exceed 80 dB and cause a hazard.
Page 9
1. Introduction Safety and EMC Information Safety Requirements The medical device has passed the tests and conformed to the standards of IEC 61010-1:2010, “Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements”. ...
Page 10
1. Introduction Equipment list Please note that following accessories accompanying with instrument may vary from region to region. Transmission cable x1 Power cord x1 Tips combo pack x1 Magnetic separator x1 Auto Tube base x 6 User manual x1 =Remark= Power cord varies from region to region.
2. Instrument Overview SPECIFICATION ITEM Model Maelstrom 9610 Weight Approx. 95 kg Dimensions 870(L) x 575(W) x 700(H) mm Power rating 100-120Vac, 50/60Hz, 8A (Class I) Fuse 250V, 10A Max. Throughput 96 samples per run Process. volume 50μl -1600μl Spin speed...
Page 12
2. Instrument Overview ① ② ③ USB type A port is for the USB type A use of barcode scanner. USB type B port is for the connection with PC. Power Button & USB type B USB I/O Use “►” and “◄” ...
Page 13
① USB type A for use of upgrading HMI program ② USB type A for use of upgrading PLC program ③ Power inlet ④ Fuse Please note that only TANBead technicians and authorized technicians or service engineers shall engage in HMI and PLC program upgrade.
Page 14
2. Instrument Overview Control Panel To turn off the alarm To pause the program. Press again to resume the program. To turn on/off the LED light. To initiate the program Emergency to stop machine...
3. Installation & Get Started Installing Maelstrom 9610 is easy. Please follow the guide below to correctly set up the instrument. Please note that this instrument weights around 95kg, it is highly recommended to have 3 or more persons to handle when taking it out from the box and must pay attention to the safety.
4. On Screen Display and How to Use Create an User ID An user ID is required to access the instrument. For a brand new machine, there is a need to create an user in advance. Click the icon to create your own user ID. Tab the icon to create a new user ID User...
Page 17
4. On Screen Display and How to Use Save User ID After creating the user ID, please must click to save and then program will automatically lead you access the system. Instrument is trying to save new user ID After save, you will be guided to the main page.
Page 18
4. On Screen Display and How to Use Possible Error Message Some messages may occur when you try to log in or create an user User ID not found Make sure you input correct, existing user ID account User ID repeated Please change the name of creating an ID Account is full...
Page 19
4. On Screen Display and How to Use Main Page There are six functions that can be used on main page. To test on mounting and ejecting spin tips To execute the extraction program To export extraction records REPORT To use UV & HEPA to sterilize the instrument STERILIZE To edit, add, and delete extraction program PROGRAM...
Page 20
4. On Screen Display and How to Use Tip (Test on Mount and Eject Spin Tips) Tip function is designed to test the performance of mounting and ejecting spin tips. Tip Point (Setting): Parameter setting value upon mounting tips ...
Page 21
4. On Screen Display and How to Use How to test mounting and ejecting spin tips Step 1: Click “All Home” to set X/Y/Z-axis to zero All Home Step 2: Place tip combo pack (for use of mounting spin tips) or blank 96DW plate (for use of ejecting spin tips) to #8 plate position Tip function is designed to test the performance of mounting and ejecting...
Page 22
Run function is designed to execute nucleic acid extraction program stored in the instrument. Maelstrom 9610 is capable of storing up to 25 programs. Please follow the guideline of reagent kits and correctly select the program. Step 1: Click the program first, and then click [OK] to continue...
Page 23
Sample ID can bear 18 characters input. Once completed, tab [Next] to continue. No.: Total number of Sample ID Pos.: Hole position on the plate ID: Name of Sample ID =Remark= Maelstrom 9610 can be compatible only with USB-type infrared barcode scanner.
Page 24
4. On Screen Display and How to Use Run (Execution of Extraction Program) Step 5: Follow the guide on screen to place the reagent plate onto correct plate accordingly. Please note that the 1 plate to be placed is always the tip combo pack. =Remark= Eject tip plate is allowed to changed according to user’s needs.
Page 25
4. On Screen Display and How to Use During the Run Understanding every symbol in the operation process is important. It helps you clearly know the status of nucleic acid extraction. Pause: Cease an extraction program temporarily No. of Well Name of Buffer From plate 1 to plate 6, name of buffer varies from kits to kits Currently running well...
Page 26
4. On Screen Display and How to Use During the Run Understanding every symbol in the operation process is important. It helps you clearly know the status of nucleic acid extraction. Mixing Beads collection In the course of beads collection, whether you click Pause button or Stop button, the instrument will not act immediately until the collection process comes to the end.
Page 27
Report (Export extraction records) Report function enables you to view experiment records and the identity Report that performed the experiment. Maelstrom 9610 is designed to store the last 15 experiment records. By clicking View button, you can navigate the detail of sample ID.
Page 28
4. On Screen Display and How to Use Sterilize (UV Sterilization & HEPA Filter) It is highly advised to use UV to sterilize the instrument after each Sterilize experiment. Set UV lighting time first, and then tab icon to start ...
Page 29
To add a new program, you need to use Copy to duplicate a program first, and then use Edit to write the protocol. Delete a program no longer used to save the memory space. Maelstrom 9610 can store up to 25 programs and each program can support totally 35 steps.
Page 30
4. On Screen Display and How to Use Edit a Program Tab existing value to change the setting Program Name of program, up to 10 characters Plate No. of plate Vol.(μl) Total processing volume Preheat(°C) Heating a plate to a particular temp before mixing step Preheat temp range from 25°C to 100°C Action...
Page 31
4. On Screen Display and How to Use Edit a Program Step Extraction step Maelstrom 9610 supports up to 35 steps in a program Plate No of plate Temp Protocol temp required to perform nucleic acid extraction Protocol temp and Preheat temp are different. When...
Page 32
4. On Screen Display and How to Use Add a Program To add a new program, you need to use Copy to duplicate a program first, and then use Edit to set desired protocol. Please find a number to add a program. Then, tab a program source, and click Copy icon to duplicate a program to assigned number.
Page 33
4. On Screen Display and How to Use Delete a Program To delete a program, tab the program no longer need and then click Delete icon to execute. Click Yes to confirm the deletion...
Page 34
Setting barcode scanner management. Admin user is allow to assess Language, Time, and Barcode functions. Language Maelstrom 9610 supports multi-language display, including English, Traditional Chinese, Simple Chinese, Korean and Spanish. Time When a new instrument arrives in, we strongly suggest user setting right time for the instrument.
Page 35
4. On Screen Display and How to Use Setting (Language) Maelstrom 9610 supports multi-language operating interface including English, Traditional Chinese, Simple Chinese, Korean and Spanish. Tab Language icon to access Default is English, simply tab icon to switch display language...
Page 36
4. On Screen Display and How to Use Setting (Time) Set correct time for the system according to the local time Click icon to set Date and Time for the system; once set up is done, tab [OK] to apply the latest setting.
Page 37
4. On Screen Display and How to Use Setting (Barcode) Maelstrom 9610 can be compatible with USB-type infrared barcode scanner. Click the icon to ON/OFF of barcode scanner If barcode scanner is set OFF, even USB-type barcode scanner is connected to Maelstrom 9610, scanned data will not be...
If there is a need to contact us, you can call our rep number at +886- 3-3167568 or email us at service@tanbead.com for further assistance. To efficiently solve the problems and answer your questions, please provide us your instrument serial number when you talk to our technician personnel.
To clean the magnetic rods, wipe with a soft cloth dampened with pure water. Do not use alcohol solvent. If the magnetic rods cannot be cleaned, please contact TANBead (Tel: +886-3-3167568) for technical assistance. 7. Disposal In case the product is to be disposed of, the relevant legal regulations are to be observed.
8. Patent Patent List US09616398B2 EP2937136 Canada CA2862946 Japan JP6151735B2 Korea KR101696517B1 China CN104971638B Taiwan TWI526245B WIPO WO2016127292 9. About Manufacturer Here comes the detail of instrument manufacturer and its EC Rep. Manufacturer:Taiwan Advanced Nanotech Inc. Manufacturer Add:No. 2, Aly. 12, Ln. 81, Longshou St., Taoyuan District., Taoyuan City 330, Taiwan ...
10. Appendix Fuse Replacement on Maelstrom 9610 Specification of Fuse: UL recognized component Rated 10A, 250VAC x2pcs UFE, Size 5.2Φ x 20mm Precautions of Fuse Replacement Please note that severe electrical shock which may cause physical injury or death could be resulted from working on an instrument as the high voltage power supply is in operation.
Page 42
PC (Type A to PC, Type B to instrument). (USB Type B) (USB Type A) Execute the utility program “Maelstrom 9610 Master” You can find the utility program from the CD that comes with the instrument. Load the CD to the PC and use Windows Explorer to find the folder named “Maelstrom Master”.
Page 43
10. Appendix How to use Maelstrom Master to update extraction protocols in the instrument? Program Editing After connection, the utility program will automatically read all protocols stored in the instrument. Select the protocol first, and then choose an editing function to apply. Please not that the protocol you select will be marked in Blue.
Page 44
10. Appendix How to use Maelstrom Master to update extraction protocols in the instrument? Remember to SAVE settings After adjustment of protocol contents, please MUST click “Save” to get most updated contents effective. Once you click “Save”, the utility program will automatically update the protocol contents to the instrument.
Page 56
2. 儀器概述 ① 繁 體 中 ② ③ 文 USB type A 接口:用於USB USB type A 式條碼掃描槍使用 USB type B 接口:用於連接 電腦端使用 電源按鈕及USB接口 USB type B 請利用 “►” 及 “◄” 按鈕 轉盤按鈕 可手動控制轉盤轉向 轉盤上總計有8個載座, 其中1/2/3/6號載座備 有獨立溫度控制模組。 轉盤 8號載座是專門放置旋...
Page 57
2. 儀器概述 後視圖 ① 繁 體 中 ② 文 ③ ④ ① USB type A 接口:更新人機程式使用 ② USB type A 接口:更新PLC程式使用 ③ 電源線接口 ④ 保險絲 請注意,只有台灣圓點奈米工程師或是經認證授權的工程師可以 進行人機程式及PLC程式的更新作業。...
Page 100
2. 设备概述 ① ② ③ 简 体 中 文 USB type A 接口:用于USB USB type A 式条形码扫描枪使用 USB type B 接口:用于连接 计算机端使用 电源按钮及USB接口 USB type B 请利用 “►” 及 “◄” 按钮 转盘按钮 可手动控制转盘转向 转盘上总计有8个载座, 其中1/2/3/6号载座备 转盘 有独立温度控制模块。 8号载座是专门放置旋...
Page 101
2. 设备概述 后视图 ① ② 简 ③ 体 中 ④ 文 ① USB type A 接口:更新人机程序使用 ② USB type A 接口:更新PLC程序使用 ③ 电源线接口 ④ 保险丝 请注意,只有台湾圆点奈米工程师或是经认证授权的工程师可以 进行人机程序及PLC程序的更新作业。...