AEM
Monitor Setup Sheet
EC REP – according to MDD93/42/EEC
1275
MDSS GmbH
Schiffgraben 41
EM2+E only
30175 Hannover, Germany
Locate the on/off switch located on the side of the
AEM Monitor and turn it on. Verify that all visual
indicators illuminate and two beeps are heard.
Plug the green INHIBIT ADAPTER into the ESU
return electrode receptacle. Then plug the return
electrode into the adapter. Verify the amber RETURN
ELECTRODE light turns off.
Plug the cylindrical end of the AEM CORD ADAPTER
into the ESU, plug the other end into the AEM CORD
receptacle of the AEM Monitor.
NOTE: Do not sterilize the AEM CORD ADAPTER.
Plug either end of the AEM CORD into the AEM
CORD ADAPTER.
Plug the other end of the AEM CORD into the
instrument of choice.
Verify the amber AEM CORD light turns off and the
green READY light is on. The AEM Monitoring
System is READY to function.
ENCISION recommends that a backup AEM Cord and
AEM Cord Adapter be available at all times.
©Copyright 2010
Printed in USA
00701-008

Advertisement

loading
Need help?

Need help?

Do you have a question about the AEM and is the answer not in the manual?

Questions and answers

Summary of Contents for Encision AEM

  • Page 1 Verify the amber RETURN ELECTRODE light turns off. Plug the cylindrical end of the AEM CORD ADAPTER into the ESU, plug the other end into the AEM CORD receptacle of the AEM Monitor. NOTE: Do not sterilize the AEM CORD ADAPTER.
  • Page 2 ENCISION Technical Support at 800-998-0986. STEP 6 CAUTION: Do not use the AEM Monitor without the Inhibit The front panel display will indicate the Adapter plugged into the ESU return electrode receptacle. The intensity of the RF current being delivered Inhibit Adapter allows the AEM Monitor to signal the ESU to the accessory.