► Conversions to the device Unauthorized modifications to the device are not permitted and can lead to unforeseeable dangers. If, despite this, modifications are required, consult with Endress+Hauser. ► Repair To ensure continued operational safety and reliability, Carry out repairs on the device only if they are expressly permitted.
CM14 Safety instructions Observe the specifications in the separate supplementary documentation that is an integral ► part of these Instructions. Designated use The transmitter evaluates measured values of an analytical sensor and shows them on its multicolored display. Using the unit's outputs and limit relays, processes can be monitored and controlled.
Page 6
Safety instructions CM14 CAUTION Cause (/effects) Where applicable, consequences of failure to comply Avoidance measure ► Indicates a potentially hazardous situation which, if not avoided, may result in minor to ► moderate injury. NOTICE Cause (/effects) Where applicable, consequences of failure to comply Avoidance measure ►...
CM14 Identification Identification Device designation 2.1.1 Nameplate Compare the nameplate with the following illustration: A0015221 å 1 Nameplate of the transmitter (example) Device designation Order code, serial number and ID number of the device Power supply Power consumption Output variables...
Identification CM14 Certificates and approvals CE mark, Declaration of Conformity The process display unit is designed to meet state-of-the-art safety requirements, has been tested and left the factory in a condition in which it is safe to operate. The device meets the relevant standards and directives as per EN 61 010-1 "Safety requirements for electrical equipment for...
CM14 Installation Installation Incoming acceptance, transport, storage The permitted ambient and storage conditions must be observed. The precise specifications can be found in Section "Technical data" (® ä 33). 3.1.1 Incoming acceptance On receipt of the goods, check the following points: •...
Installation CM14 Installation procedure The required panel cutout is 92 mm x 45 mm (3.62 in x 1.77 in). A0015216 å 2 Installation in the panel 1. Screw the threaded rods (item 2) into the positions provided on the mounting frame (item 1).
CM14 Wiring Wiring WARNING Danger from electrical voltage The entire connection of the electrical system must take place while the device is de- ► energized. Danger if protective ground is interrupted The protective ground connection must be established before any other connection is made.
Wiring CM14 Connecting the transmitter A0015215 å 3 Connection diagram of the transmitter Terminal Description Terminal for Memosens cable, brown, sensor power supply U+ Terminal for Memosens cable, white, sensor power supply U- Terminal for Memosens cable, green, Com A...
CM14 Wiring Post-connection check Device condition and specifications Notes Are the device or cables damaged? Visual inspection Electrical connection Notes Does the supply voltage match the specifications on the nameplate? 24 to 230 V AC/DC (–20 % / +10 %) 50/60 Hz...
Operation CM14 Operation The easy operating concept of the device makes it possible for users to commission the device for many applications without a printed set of Operating Instructions. Display and device status indicator/LED A0015891 å 4 Display of the device...
CM14 Operation A0010420 • Open the configuration menu • Confirm an entry • Select a parameter or submenu offered in the menu A0010421 Within the configuration menu: • Scroll step-by-step through the parameters/menu items/characters offered • Change the value of the selected parameter (increase or decrease)
Operation CM14 Accept entry. If this icon is selected, the information entered is accepted at the position and the user exits the editing mode. Reject entry. If this icon is selected, the information entered is rejected and the user exits the editing mode. The text configured beforehand remains unchanged.
CM14 Commissioning Commissioning Installation check and switching on the device Make sure that all post-connection checks have been carried out before you commission your device: • "Installation check" checklist, (® ä 10). • "Post-connection check" checklist, (® ä 13). Once the operating voltage is applied, the green LED lights up and the display indicates that the device is operational.
Commissioning CM14 Parameter Configuration options Description Out main 0/4 mA Numerical value 0.000 to 99 999 Physical value corresponding to the lower range 1.0 pH limit of the analog output. If the set value is undershot, the current output goes to the saturation current 0/3.8 mA.
Page 19
CM14 Commissioning Parameter Configuration options Description Operating mode pH (pH sensor) Configures the operating mode. mV (ORP sensor) Unit pH, mV Unit of the physical value. Format none, one, two Configure the number of decimal places for the display. Damping main value...
Commissioning CM14 Parameter Configuration options Description Set point 2 Numerical value Only for function In band or Out band. Hyst. Numerical value Configures the hysteresis. Not for function Error. Delay time 0 to 60 s Configure the delay until the relay switches.
Page 21
CM14 Commissioning Parameter Configuration options Description Manufact. ID Read only Display the manufacturer's ID Manufact. name Read only Display the manufacturer name...
Calibration and adjustment CM14 Calibration and adjustment Definitions 7.1.1 Calibration (as per DIN 1319): Determining the relationship between the measured or expected value of the output variable and the corresponding true or correct value of the measured variable (input variable) for a measuring device under specified conditions.
Page 23
Calibrating with buffers with pH 7.0 and 4.0 is recommended for most applications. You use calibration buffers to perform two-point calibration. The quality buffers supplied by Endress+Hauser are certified and measured in an accredited laboratory. The accreditation (DAR registration number "DKD-K-52701") confirms that the actual values and the maximum deviations are correct and traceable.
Calibration and adjustment CM14 The calibration is not completed successfully or is canceled and is not valid. Possible reasons: • The sensor is old or contaminated. As a result, the permitted limit values for the slope and/ or zero point are exceeded.
CM14 Maintenance 10. Remove the sensor from the measurement medium, flush with distilled water, dry it and place it back into the measurement medium. To calibrate the ORP sensors, you can also remove them from the medium and calibrate them in the laboratory.
CM14 Troubleshooting Troubleshooting The following section provides you with an overview of possible causes of errors to provide you with an initial troubleshooting aid. 10.1 Troubleshooting instructions WARNING Danger from electrical voltage Do not operate the device for troubleshooting purposes while it is open! ►...
Page 28
Troubleshooting CM14 Sample display: F 61 sensor elec. A0015896 M 915 USP warning A0015897 S 844 Process value A0015898 C 107 Calib. active A0015899 Diagnostics Message text Description code Sensor data invalid. Remedy: Sensor data • Update date of the transmitter •...
Page 29
CM14 Troubleshooting Diagnostics Message text Description code No sensor communication. Possible reasons: • No sensor connection • Faulty sensor connection • Short-circuit in the sensor cable • Short-circuit in the neighboring channel F100 Sensor comm. • Sensor firmware update canceled with an error Remedy: •...
Page 30
Troubleshooting CM14 Diagnostics Message text Description code Faulty parameter checksum Possible cause: F846 Param error Firmware update Remedy: Reset parameters to factory defaults F847 Couldn't save param Parameters could not be saved F848 Calib AO1 Faulty calibration values for analog output 1...
Page 31
CM14 Troubleshooting Diagnostics Message text Description code Measured value outside the specified range. Measured value outside the specified range. Possible reasons: • Sensor in air • Air cushion in the assembly S844 Process value • Incorrect sensor inflow • Sensor defective Remedy: •...
CM14 Technical data Please enclose a note describing the fault and the application when sending the unit in for repair. 10.5 Disposal The device contains electronic components and must, therefore, be disposed of as electronic waste in the event of disposal. Please observe in particular the local waste disposal regulations of your country.
Page 34
Technical data CM14 11.2.4 Alarm output The alarm output is designed as an "open collector." In normal operation the alarm output is closed. In the event of a fault (F-fault, device without current) the "open collector" opens. Current max. 200 mA Voltage max.
Page 35
CM14 Technical data 11.5 Wiring 11.5.1 Electrical connection A0015303 Connection Description Terminal for Memosens cable, brown, sensor power supply U+ Terminal for Memosens cable, white, sensor power supply U- Terminal for Memosens cable, green, Com A Terminal for Memosens cable, yellow, Com B...
Page 36
Technical data CM14 Connection Description Terminal for analog output 1, - Terminal for analog output 2, + Terminal for analog output 2, - R11, R12, R13 Terminal for relay 1 R21, R22, R23 Terminal for relay 2 11.5.2 Supply voltage...
Page 37
CM14 Technical data Orientation The orientation is determined by the readability of the display. Max. viewing angle range +/- 45° from the central display axis in every direction. A0010351 å 6 Panel cutout 11.8 Environment 11.8.1 Ambient temperature range –10 to +60 °C (14 to 140 °F) 11.8.2...
Page 38
Technical data CM14 11.9 Mechanical construction 11.9.1 Dimensions A0015925 å 7 Dimensions of the transmitter 11.9.2 Weight 0.3 kg (0.66 lbs) 11.9.3 Material Housing, tube: Polycarbonate Front foil: Polyester, UV-resistant 11.9.4 Terminals Max. 2.5 mm² (22-14 AWG; torque 0.4 Nm (3.5 lb in)) line, relay...
Page 39
CM14 Technical data 11.10 Display and operating elements 11.10.1 Operating elements A0018699 å 8 Display and operating elements LC display for measured values and configuration data Status LED power supply connected Status LED alarm function Status LED limit function relay 1...
Page 40
Technical data CM14 Other standards and guidelines • IEC 60529: Degree of protection by housing (IP code) • IEC 61010-1: 2001 Cor 2003 Safety requirements for electrical equipment for measurement, control and laboratory use...
Need help?
Do you have a question about the CM14 and is the answer not in the manual?
Questions and answers