Troubleshooting
E 40
EEPROM data incorrect
Table 4: Error messages [334|335|382|383|384|385|833|834|835|874|875|876|877|878|952|956]
3.4 Error messages [769|799|780|869|899]
If a malfunction or defect occurs during operation of the scale, it is shown on the display
as an error message. Error messages are displayed in the following form:
"E mm.xx"
E:
Error
mm: Number of the module reporting the error
(10 = weight calculation, 11 = electronic display)
xx:
Error number (Table 5)
Displayed
Load cell defective, supply
E XX. 11
leads damaged or electronics
defective
Preload too great, load cells
E XX.12
damaged or electronics
defective
Limit for automatic zero tracking
E XX.13
reached
03.04.2013/JEG
Cause
17 (24)
seca serva: Recalibrate scale
Replace electronics
Remedy
Diagnosis with seca serva:
Check load cell connections
Poss. replace load cell or
electronics
Check preload on scale
Diagnosis with seca serva:
Poss. replace load cell or
electronics
Place load of >0.5d/s on scale
seca
30-34-00-864
Need help?
Do you have a question about the 334 LC and is the answer not in the manual?
Questions and answers