The total is seven sets of data. The next step is to put seven sets of data entered into the Instrument.
Pay attention to the input order in accordance with from small to large.
To enter the menu M48, then input the data set number "7", to be followed in the above seven sets of
data, we revised the setup of a multi-segment linearity correcting.
If you need to cancel linearity correcting, simply enter "0" in the menu M48.
Decommissioned linearity correcting, you just need to enter in the menu M48 data points (in this case,
is "7").
Noted: before the calibration of the instrument, you must first close the linearity correcting function. If
under the condition that the linearity correcting function has not closed, the calibration data sets arising
from the amendment must be dealing with the reverse curve of the data following the original
amendment, then input to the Instrument. Reverse amendment is very complicated, and should be
avoided.

4.37 How to save / restore frequently-used pipe parameters

You can use menu window 27 to store or restore from the internal Flash memory, as many as 9
different pipe parameter configurations.

5. Troubleshooting

5.1 Power-on Error Displays and Counter-Measures

The FDT-25W ultrasonic flow meter provides an automatic power-on diagnosis for the hardware
problems. When any message (with the power on) in the following table displays, counter-measures
should be taken.
Error message
ROM Testing Error
Segment Test Error
Stored Data Error
Timer Slow Error
Timer Fast Error
Date Time Error
0.99 )
1.
)
Causes
Problem
with
software
The
parameters
entered by the user
lose integration.
Problem
with
timer-keeper or the
crystal oscillator.
Number errors with Initialize
Counter-measures
the
(1)Power on again
(2)Contact with factory
When this message displays, the
user should press ENT key, and all
the configuration will be restored to
the default state.
the
(1)Power on again
(2)Contact with factory