Summary of Contents for Smiths Medical CADD-Legacy PCA
Page 1
CADD-Legacy ® Infusion Pump Model 6300 Quick Reference Card for Clinicians Lock Level 0 (LL0) Display shows Threaded Mounting Hole programming information for use with Polemount and messages Bracket Adapter (back of pump) Power Jack used to plug in AC Adapter Battery Compartment (back of pump) Accessory Jack used to...
Page 2
Programming / Set-up sequence Batteries must be installed; pump must be stopped and in LL0. NOTE: Value Not Saved is displayed if a value is scrolled and ENTER/CLEAR is not pressed. Press NEXT to continue programming. Warning: If the pump is dropped or hit, inspect it for damage.
Page 3
Programming / Set-up sequence - Continued Press ENTER/CLEAR. Press NEXT. Enter Doses Per Hour Note: This screen will only appear if you have programmed a demand dose and dose lockout is less than 1 hour. Press or to select desired Doses Per Hour. Press ENTER/CLEAR.
Page 4
Pump Operations Change the Lock Level Stop the pump. Press LOCK. Press or until desired lock level appears. Press LOCK or ENTER/CLEAR. Press or until the lock level code appears. Press LOCK or ENTER/CLEAR. Start the Pump Press and hold STOP/START until (–...
Page 5
Pump Operations - Continued Stop the pump. Push down and hold the arrow button on the battery door while sliding the door off. Remove and discard old batteries. Important: Always stop pump before removing batteries. Install new batteries, matching polarities shown on the pump.
Page 6
Pump Operations - Continued Press NEXT until NEXT FOR BIOMED, ENTER FOR MAIN appears. Press ENTER/CLEAR to return to main screen. Or press NEXT to continue in Biomed settings loop. Changing the Cassette To Remove Used Cassette Stop the pump. Close all tubing clamps.
Programming screens Main Screen -Displays status of pump (STOPPED or RUN) Reservoir Volume -Volume of fluid in reservoir Units -Programming units (ml, mg, mcg) Concentration - -Concentration of medication in mg/ ml or mcg/ml Continuous Rate -Continuous rate of infusion (ml/hr, mg/hr, mcg/hr) Demand Dose -Amount delivered when patient...
Dosing To Start Demand Dose Pump must be running. Press DOSE or remote dose cord button. Note: If you wish to stop delivery of the demand dose, press and hold STOP/START to stop the pump. To Start Clinician Bolus Pump must be running. Press LOCK.
Page 9
Alarms & Troubleshooting Battery Depleted Two-tone alarm Batteries are depleted and cannot operate pump; install new batteries immediately. Battery Removed Pump Won’t Run Two-tone alarm With AC adapter attached, batteries have been removed while pump was running, or you have tried to start pump with depleted batteries.
Page 10
United States. CADD, CADD-Legacy, the Medication Cassette Reservoir and the Smiths Medical design mark are trademarks of Smiths Medical. The symbol ® indicates the trademark is registered in the U.S. Patent and Trademark Office and certain other...