INSTRUMENTS & TOOLS
Be aware of the force that you apply to screws, as they or their housings can break if too much force is used.
For information about the parts and components used when servicing a BVI 3000 system, see
on page 87. The following tools are required in order to service the system:
• Screwdriver with interchangeable bits
• #1 Phillips bit
• #2 Phillips bit
• T15 star bit
/
‑inch wrench or socket wrench
3
16
• Bar clamps (2)
• Tweezers
• Pliers
• File
• Ionized air gun
• EPROM extractor
• Isopropyl alcohol, 70% or higher
• Loctite 422 bonder (0135‑0042)
• Double‑sided tape (0410‑0011)
* See
Part Numbers
on page 87 for orderable part number
4
• All‑purpose cleaner
• Kimtech wipes (or equivalent)
• Cotton swabs (Q‑tips)
• Disposable gloves
• Master key (0810‑0110)
• PC station with access to a printer
• Certified adult bladder phantom (0620‑0068)
• BVI 3000 string phantom kit (0620‑0156)
• BVI 3000 test console (0570‑0090)
• BVI 3000 test probe (0570‑0091)
• BVI 3000 battery charger (0400‑0132)
• BVI 3000 battery (0400‑0039)
• Thermal paper roll for printer (0800‑0042)
Part Numbers
*
*

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BLADDERSCAN BVI 3000 and is the answer not in the manual?

Questions and answers

ali
February 18, 2025

How to reset the calibration timer

1 comments:
Mr. Anderson
February 18, 2025

To reset the calibration timer on the Verathon BladderScan BVI 3000, follow these steps:

1. On the Setup Now screen, press the SEL+ or SEL- buttons until Calibration Date Menu is selected.
2. Press the GO-> button to enter the Calibration Date Menu.
3. Enter the desired next calibration date for the instrument.
- Verathon recommends setting this to 14 months from the currently configured date.
4. Press the Set button to confirm the new calibration date.
5. Press the EXIT button to return to the Main Configuration screen.

This process updates the calibration timer for the device.