2. Application
2.2.4.7 Maintenance programs
Scroll to page 3 in the home screen and press the [Maintenance
programs] button.
You have the following program options:
– Cleaning program
– Service position
Cleaning program
The firing chamber is "cleaned" with the cleaning program. After a cleaning program, it is recommended to calibrate the
furnace. More detailed information on the cleaning program and on cleaning the firing chamber can be found in Chapter
3.2 "Cleaning"
Service position
To be able to clean the firing chamber or to remove particles from the firing chamber, the service position must be
executed. For that purpose, the furnace head is uncoupled from the mobile cooling tray area to enable access to the firing
chamber.
To clean the firing chamber, the furnace and the firing chamber must be cold.
The service position can be executed in only a few steps:
Step 1:
Open the service position
Scroll to page 3 in the home screen and press the [Maintenance
programs] button. Press the [Service position] button in the Care
Programs menu.
Step 2:
Moving to the service position
Press the [Go to service position] button. The furnace head then
moves to a higher position than in normal operation.