Siemens SITRANS F Operating Instructions Manual page 32

Electromagnetic flowmeters
Hide thumbs Also See for SITRANS F:
Table of Contents

Advertisement

Installing/Mounting
4.8 Potential equalization
● Isolate the sensor from cathodic protected pipes using insulated bolts.
● Use bypass cable between the mating flanges.
Note
Remote mounted sensor versions
If the above is not acceptable, remote mounted sensors can alternatively be connected as
follows:
● Connect coil current cable shield at sensor end via a 1.5 µF condensator
● Make sure that electrode cable shield is not connected at both ends
32
SITRANS MAG 1100/1100 F
Operating Instructions, 08/2018, A5E03433301-AD

Advertisement

Table of Contents
loading

Table of Contents