FAULTS AND WARNINGS

The Mini-GL constantly monitors its sub-systems. The following are conditions that may interfere with 'normal' system
operation and will display a warning or fault message. Note the "C" and "W" indicates a CAL or Warmup is required.
-1.
Mini-GL Fault Type
1 - BUBBLE DETECT
CIRCUIT FAULT
2 - ROM CHECKSUM
3 - SHUTTER FAULT
5 - HV TOLERANCE
FAULT
6 - CALIBRATION FAULT
7 - DI SYSTEM FAULT
8 - DCD PRESSURE
FAULT
9 - WARM-UP TIMEOUT
10 - DELIVERY SYSTEM
FAULT
12 - ENERGY OUT OF
RANGE FAULT
13 - TRIGGER SWITCH
FAULT
15 - TRANSMISSION
FAULT
WARNING
(16) REPLACE CANISTER
(17) PURGE REQUIRED
EXIT TO CLEAN WINDOW?
1010-06-3020 Revision B2
~ CANDELA CORPORATION PROPRIETARY ~
FAULT
HP Bubble Circuit Test didn't detect a change in the signal (with DCD-enabled HP).
1.1
(Internal [Candela] or External Cylinder DCD system must be installed)
Canister Bubble Circuit Test didn't detect a change in the signal (with DCD-enabled HP).
1.2
(Internal [Candela] DCD system must be installed)
2
Calculation of checksum at power-up does not match checksum value in memory.
3
Shutter isn't in correct state when checked. Does not respond to actuation to correct state.
5.1
HV sample does not settle to within +-8 volt at End of Charge (EOC) within 1.5 seconds.
5.2
HV sample is not within +-3% of HV reference when HVPS is at EOC.
5.3
HV sample is not within +-3% of HV reference after DCD pre-spray already occurred.
6.1
CAL Low slope point not in range of 500V to 1800V.
6.2
Expected Head Energy (xHD) is calculated > Max HD Energy of 75 J.
6.3
Laser failed to CAL to desired Fluence within 25 pulses.
6.4
CAL required a HV > 2000V for desired Fluence.
7.1
DI temp < 60degC when not in warm-up.
7.2
DI temp > 70degC while in Ready state.
DI pressure switch does not change when power turned on. OR
7.3
DI pump is not ON or DI pressure switch not actuated.
DI temp < Thermistor Open Temperature (5) OR
7.4
DI temp > Thermistor Shorted Temp (98) while in Ready state
DCD pressure < 105 psi while in READY & INT_DCD & HP w/DCD & spray setting is non-zero.
8.1
8.2
DCD Pressure > 135 psi & INT_DCD
9.1
DI temperature < 62degC after 60 minute warm-up
9.2
DCD pressure < 105psi after 60 minute warm-up (Only when DCD enabled)
10.1
HP type changed or unrecognized while in Ready (15 consecutive reads at 50 ms interval)
10.4
Fiber not detected while in Ready (15 consecutive reads at 50 ms interval)
10.5
Slider button pressed while in Ready State. Cannot change spot-size while in Ready.
12.1
On last treatment pulse, the head energy (HD) was 14% lower than expected head energy (xHD). C
12.2
On last treatment pulse, the head energy (HD) was 14% higher than expected head energy (xHD). C
12.3
The head energy (HD) of the last pulse > Max Treatment HD Energy of 75 J.
13
The redundant trigger switches were in two different states for > 1 second while in ready.
15.1
Transmission (Tx) is < 70%. (HD or CP energy > 5.0 J)
15.2
Transmission (Tx) is > 100%. (HD or CP energy > 5.0 J)
DESCRIPTION
Displays once when DCD Canister Pulse count = 0
Displays if canister bubble is detected in Ready and DCD is enabled.
Bubble percentage is outside of tolerable range (~15% bubbles).
Transmission is < 75% - Delivery system degrading - Clean Window
DESCRIPTION
C
C
C
C
C
C
C
CW
CW
CW
C
W
W
W
W
C
C
C
C
C
C
Page 14 of 14