TROUBLESHOOTING TIPS

The following table provides troubleshooting tips for using the AED Trainer 3.
If you need additional assistance, please contact your local Philips
representative or go to www.philips.com/AEDsupport.
ISSUE
AED Trainer 3 does
not work and gives
beeping signals when
turned on.
AED Trainer 3 gives
"Analyzing" prompt
before pads are
applied to manikin.
AED Trainer 3 does
not respond when
pads are applied to
manikin.
POSSIBLE CAUSE
• No Operating
System Card is
installed.
• Incorrect card is
installed.
• Operating System
Card has corrupt or
missing files.
• The HeartStart
Training Pads II are
connected; the AED
Trainer 3 senses that
pads are in place as
soon as the pads
connector plug is
inserted.
• Laerdal LINK
Technology Training
Pads are incorrectly
placed on the
Laerdal LINK
Technology manikin.
SUGGESTED ACTION
• Install the Operating
System Card.
• Install the original
Operating System
Card.
• Obtain replacement
Operating System
Card.
• To use HeartStart
Training Pads II, apply
the pads to the
manikin before
connecting them to
the AED Trainer 3.
Alternatively, use
HeartStart Training
Pads III or the
Laerdal LINK
system, so that the
AED Trainer 3
automatically detects
when pads are
correctly placed.
• Correct the pads
position.