1.13. Preparations before Use
Radiofrequency lesion generator
1. Check the rated voltage is correct for the equipment before connecting the power.
Caution
The equipment may be damaged if it is not connected to the correct voltage.
2. Warning: To avoid risk of electric shock, this equipment must only be connected to a power
supply with protective earth.
3. Avoid using the equipment at an unsanitary or flammable place.
4. Power on, Power off Procedure
4-1. Before Surgery
1) Connect the power cable to the RF generator
2) Press the power switch
3) Check the main menu
4-2. After Surgery
1) The output is stopped
2) Press the power switch
3) Disconnect the power cable from the RF Generator
Peristaltic pump
1. Connect the power cable to the pump's rear part.
2. Warning : To avoid risk of electric shock, this equipment must only be connected to a power
supply with protective earth.
Electrode tubing set connection
Preparation materials:
Coolant container (3L capacity)
Cooled IV bag (1 - 3L)
1. Make sure that the IV bag is sufficiently cooled before the treatment.
2. Use the saline solution as coolant before the treatment.
Note: 2L coolant is appropriate for a 12 minute-long treatment. The pump's flow rate is
appropriately 100ml/min.
3. The coolant temperature is indicated on the generator when the coolant is connected and
circulating through the pump. The cooling temperature is normally less than 20℃. When the
cooling temperature is over 25℃, ensure that the coolant's temperature is maintained by
placing the IV bag into the coolant's storage container.