RKI Instruments 65-2397RK Operator's Manual

RKI Instruments 65-2397RK Operator's Manual

M2a transmitter

Advertisement

Quick Links

65-2397RK
CO
Transmitter
2
Operator's Manual
Part Number: 71-0185RK
Revision: C
Released: 9/1/20
RKI Instruments, Inc.
www.rkiinstruments.com

Advertisement

Table of Contents
loading

Summary of Contents for RKI Instruments 65-2397RK

  • Page 1 65-2397RK Transmitter Operator’s Manual Part Number: 71-0185RK Revision: C Released: 9/1/20 RKI Instruments, Inc. www.rkiinstruments.com...
  • Page 2 Frequency of calibration depends upon the type of use you have and the sensor types. Typical calibration frequencies for most applications are between 6 and 12 months, but can be required more often or less often based on your usage. 2 • 65-2397RK CO Transmitter...
  • Page 3 Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge.
  • Page 4: Table Of Contents

    Parts List ..............21 4 • 65-2397RK CO...
  • Page 5: Overview

    ± 5% of reading or ± 2% of full scale (whichever is greater) WARNING: When using the 65-2397RK, you must follow the instructions and warnings in this manual to assure proper and safe operation of the 65-2397RK and to minimize the risk of personal injury.
  • Page 6: Description

    Factory Adjust Pot Span Pot Zero Pot Amplifier Factory Adjust Pot Jumper Pins, Factory Use Only Detector Terminal + Test Point Strip - Test Point J-Box IR CO2 Detector Figure 1: CO Transmitter Component Location 6 • 65-2397RK CO Transmitter...
  • Page 7: Co 2 Detector

    A cap gasket seals the interface between the housing and cap. Plug-In CO Sensor The plug-in sensor is secured in the detector assembly by the housing cap. The sensor produces a millivolt output that corresponds to the detection range of the transmitter. 65-2397RK CO Transmitter • 7...
  • Page 8: Amplifier

    Three spacers installed on the back of the junction box control the distance of the junction box from a mounting surface and ensure that there is enough room to install a calibration cup 8 • 65-2397RK CO Transmitter...
  • Page 9: Installation

    The exhaled CO may cause an alarm. 3.10 3/4 NPT Female 3.65 6.80 M AX Rubber Spacers, J-Box IR CO2 Detector 1 1/2-20 Thread for Calibration Cup Figure 3: Mounting the CO Transmitter 65-2397RK CO Transmitter • 9...
  • Page 10: Wiring The Co 2 Transmitter

    Use appropriate conduit fittings and construction technique for the environmental rating of the controller. RKI controllers are typically rated NEMA 4X, are not explosion-proof, and must be installed in a non-hazardous location. 10 • 65-2397RK CO Transmitter...
  • Page 11 12. If shielded cable is used, connect the cable’s drain wire to an available chassis (earth) ground at the controller. RKI controllers typically have a ground stud that can be used to ground the cable’s drain wire. 65-2397RK CO Transmitter • 11...
  • Page 12: Startup

    4. Screw the calibration cup onto the bottom of the IR CO detector. 5. Screw the regulator into the 100% nitrogen calibration cylinder. 6. Use the calibration kit sample tubing to connect the regulator to the calibration cup. 12 • 65-2397RK CO Transmitter...
  • Page 13: Maintenance

    Investigate significant changes in the display reading. Biannually Calibrate the CO transmitter as described in “Calibration” on page 19 of this manual. 65-2397RK CO Transmitter • 13...
  • Page 14: Troubleshooting

    6 months. section. Some applications 5. If the calibration/response may require a more difficulties continue, contact RKI frequent calibration for further instruction. schedule. 14 • 65-2397RK CO Transmitter...
  • Page 15: Replacing Components Of The Co 2 Transmitter

    6. Make sure the cap gasket is in place and screw the detector housing cap back onto the detector housing body. 7. Turn on power to the controller. 8. Turn on the controller and place it into normal operation. 65-2397RK CO Transmitter • 15...
  • Page 16 LEL “R” White LEL “W” Green LEL “G” Black LEL “B” 9. Reinstall the detector terminal strip into its socket. 10. Reinstall the junction box cover. 11. Turn on or plug in power to the controller. 16 • 65-2397RK CO Transmitter...
  • Page 17 Table 4 and Table 5 below. Table 4:Reconnecting the Amplifier to the Controller Controller Amplifier Controller Transmitter Terminal Terminal Strip Strip (typical) PWR/SIG “-” - (DC -) PWR/SIG “S” S (4 - 20 mA In) PWR/SIG “+” + 24V 65-2397RK CO Transmitter • 17...
  • Page 18: Calibration Frequency

    6 months is adequate for most infrared CO transmitter applications. Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a calibration frequency of every 6 months. If an application is not very demanding, for example detection in a clean, temperature controlled environment where calibration adjustments are minimal at calibration, then a calibration frequency of every 9 to 12 months is adequate.
  • Page 19: Calibration

    4. Allow the gas to flow for one minute and verify a reading of 100 mV (±2 mV). If necessary, use the zero pot on the amplifier to adjust the reading to 100 mV (±2 mV). 5. Turn the regulator’s on/off knob clockwise to close the regulator. 65-2397RK CO Transmitter • 19...
  • Page 20: Setting The Response Reading

    The 0-20% volume, 0-50% volume, and 0-100% volume detectors will not display a reading in fresh air because their ranges are too large. 6. Store the components of the calibration kit in a safe and convenient place. 20 • 65-2397RK CO Transmitter...
  • Page 21: Parts List

    0-50% Volume, (includes detector, junction box, and amplifier) 65-2397RK-10 transmitter, 0-100% Volume, (includes detector, junction box, and amplifier) 71-0185RK Transmitter Operator’s Manual (this document) 65-2397RK CO 81-0070RK-01 Calibration cylinder, 2000 ppm CO , 34 liter steel 81-0071RK-01 Calibration cylinder, 5000 ppm CO...
  • Page 22 34 liter steel calibration cylinders (cylinders with external threads) 81-1051RK Regulator, with gauge and knob, 0.5 liter/minute continuous flow, for 34 liter alu- minum/58 liter/103 liter calibration cylinders (cylinders with internal threads) 81-1117RK Calibration cup 82-0006RK Pot adjust screwdriver, for calibration 22 • 65-2397RK CO Transmitter...