Cleaning and Sterilization/Disinfection Instructions
19. Draw a minimum of 50 mL of cleaning solution into the syringe and flush back through the channel.
20. Keep the Flexible Videoscope immersed for the manufacturer's recommended contact time (e.g. Enzol) immersion
for a minimum of 1 minute).
21. Remove the 60 cc syringe from the port and re-plug the port with the Luer plug (P/N 29100-BK).
22. Draw air into the 60 cc syringe and attach it to the rinse tube assembly (P/N 11301CDT). Holding the distal end of
the Flexible Videoscope out of the cleaning solution, flush the 60 cc of air to expel all the cleaning solution from
the channel.
23. Remove the Flexible Videoscope from the cleaning solution.
24. Discard all disposable cleaning accessories and use new accessories for water rinsing.

Water Rinsing

Water Quality Recommendation: Utility or critical water
1. Disinfect a new sink or basin to be used for rinsing following your healthcare facility's recommended method. After
disinfecting, be sure to thoroughly rinse the sink or basin with water to remove all of the disinfectant chemicals.
2. Fill the sink with a minimum of 10 L (2.6 gallons) of room temperature water. Ensure there is sufficient volume of
rinse water to COMPLETELY immerse the Flexible Videoscope.
3. Completely immerse the Flexible Videoscope in the water.
4. Thoroughly wipe the exterior of the entire Flexible Videoscope with a fresh, soft, lint-free cloth. A minimum of one
complete wipe should be done.
5. Attach the rinse tube assembly (P/N 11301CDT) to the cleaning adaptor (P/N 11301CD1) and attach the cleaning
adaptor (P/N 11301CD1) to the suction port. Then attach a 60 cc syringe to the rinse tube assembly (P/N
11301CDT).
6. Plug the instrument port with the Luer plug (P/N 29100-BK).
7. Draw a minimum of 50 mL of water into the syringe and flush back through the channel.
8. Remove the Luer plug (P/N 29100-BK) and attach a 60 cc syringe to the instrument port.
9. Draw a minimum of 50 mL of water into the syringe and flush back through the channel.
10. Remove the syringe from the instrument port and re-plug the port with the Luer plug (P/N 29100-BK).