AtriCure ASU2-115 Service Manual

Ablation and sensing unit
Hide thumbs Also See for ASU2-115:

Advertisement

ABLATION AND SENSING UNIT
(ASU)
SERVICE MANUAL
Model ASU2-115
ASU3-230
AtriCure, Inc.
6033 Schumacher Park Drive
West Chester, Ohio 45069
USA
Telephone: 513-755-4100
Technical Support: 866-349-2342
P000050.F

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ASU2-115 and is the answer not in the manual?

Questions and answers

Summary of Contents for AtriCure ASU2-115

  • Page 1 ABLATION AND SENSING UNIT (ASU) SERVICE MANUAL Model ASU2-115 ASU3-230 AtriCure, Inc. 6033 Schumacher Park Drive West Chester, Ohio 45069 Telephone: 513-755-4100 Technical Support: 866-349-2342 P000050.F...
  • Page 2: Table Of Contents

    REPLACING FUSES .........................17 8.1. Tools Needed ........................ 17 8.2. Procedure ........................17 CLEANING AND DISINFECTING ..................18 9.1. ASU Cleaning and Disinfection Instructions..............18 9.2. Footswitch Cleaning and Disinfection Instructions............18 10. ATRICURE CONTACT INFORMATION ................18 P000050.F Page 1 of 18...
  • Page 3: Section 1. Introduction

    Please read all information carefully. Failure to properly follow the instructions may lead to serious consequences. For the user’s convenience, the AtriCure Ablation and Sensing Unit will be referred to in this Service Manual as the “ASU”. The AtriCure Bipolar Handpiece will be referred in this Service Manual as the “Handpiece”.
  • Page 4: Warnings And Precautions

    SERVICE MANUAL Warnings and Precautions The safe and effective use of RF energy is highly dependent upon factors under the control of the operator. There is no substitute for a properly trained operating room staff. It is important that the operating instructions supplied with the ASU be read, understood and followed before use.
  • Page 5: Responsibility Of The Manufacturer

    SERVICE MANUAL 2.3. Responsibility of the Manufacturer AtriCure is responsible for safety, reliability, and performance of the equipment only if: • Installation procedures in the User’s manual are followed. • Other than fuse replacement, only persons authorized by AtriCure carry out modifications or repairs.
  • Page 6: Power And Voltage Output Restrictions

    SERVICE MANUAL 3.7. Power and Voltage Output Restrictions • Power output is available over a 20Ω - 1,000Ω range. • The maximum power and voltage output, Device Code Maximum Output Power Maximum Output Voltage 57.0 Vrms 28.5 W at 114Ω 77.5 Vrms 15.0 W from 26.3Ω...
  • Page 7 SERVICE MANUAL ASU2 & ASU3 POWER OUTPUT Device Code A and F Device Code H Device Code D Device Code E Device Code G 1000 1100 1200 Load Impedance (Ohms) Figure 1 – Power vs. Load (clamp algorithm) P000050.F Page 6 of 18...
  • Page 8: Theory Of Operation

    460 kHz, with a maximum power as indicated in Section 3.7. The AtriCure ASU is designed to operate with the AtriCure Handpiece. The ASU and Handpiece are designed for use without a neutral electrode. The Footswitch is the input device used to activate RF energy delivery.
  • Page 9: Power On Self-Test

    SERVICE MANUAL Power on Self-Test The ASU goes through a Power On Self Test (POST) every time the power switch is turned on. The POST is software driven; it is a comprehensive test of the generators components. If an Error is generated during this test, the ASU goes into FAULT Mode. This Mode is entered upon detection of any unrecoverable error condition during any Mode of operation.
  • Page 10 SERVICE MANUAL Configuration Register Error • Code P08 is generated if a configuration register error is detected. MCU COP Error • Code P09 is generated if the COP watchdog timer expires. Switch Stuck Test • Code P10 is generated if the Footswitch is depressed during the POST. Real Time Clock Test •...
  • Page 11: Optional Performance Verification Procedure

    6.1. Tools Needed Tools needed to complete optional performance verification procedure: • ASU Service Test Cable. Available through AtriCure. Part Number A000075. • 50Ω , 100Ω, and 200Ω Load • Oscilloscope, capable of reading both Voltage and Current 6.2.
  • Page 12: Maximum Power Output

    SERVICE MANUAL 6.3. Power Output This test is to confirm the ASU is putting out the correct power. NOTE: It is assumed that the one performing these tests has a basic knowledge of how to operate the Oscilloscope in use. The accuracy of the measurement will vary with the test equipment used.
  • Page 13: Errors And Fault Mode

    SERVICE MANUAL 6.4. Errors and Fault Mode These tests are to confirm that errors are being detected properly. 6.4.1. Error Generation Test 1. With footswitch and service test cable plugged in and no load connected to cable, turn on the ASU. 2.
  • Page 14 SERVICE MANUAL 2. Do a visual inspection of the Footswitch and electrical power cords. • Cords should not be frayed or damaged and electrical power cords should have proper grounding. 3. Verify that the ASU is stable when placed on a flat level surface. 4.
  • Page 15: Troubleshooting

    Broken Handpiece cable Replace Handpiece Fault in Footswitch Replace Footswitch Fault in Handpiece Replace Handpiece Internal ASU failure Contact AtriCure Technical Support If the lack of ASU RF power output persists, contact AtriCure Technical Support. P000050.F Page 14 of 18...
  • Page 16: Error Codes

    If a fault condition should occur, the numeric displays on the front panel will display an error code. If an Error Code of E07 through E09, PO1 through P11, or F01 through F14 appears, try turning the ASU power off, then on. If the problem persists, contact AtriCure Technical Support.
  • Page 17 All ground wires should go to the same grounded metal with wires that are as short as possible. 4. If the above steps do not remedy the situation, contact AtriCure Technical Support to arrange evaluation of the ASU. 7.4.2.
  • Page 18: Replacing Fuses

    Input Voltage Selector Switch on the back panel of the ASU indicates the selected Input Voltage for this unit. The operator should not adjust The Input Voltage Selector Switch. This setting should only be adjusted by the manufacturer or by an authorized AtriCure technical service representative.
  • Page 19: Cleaning And Disinfecting

    The Footswitch cannot be sterilized. The Footswitch may be disinfected using a standard hospital alcohol solution applied with a cloth. AtriCure Contact Information If further assistance is needed, please contact AtriCure Technical Support. P000050.F Page 18 of 18...