KSL-H AND KSL-Z by KEELER CONTENTS INDICATIONS FOR USE ..................3 SAFETY ......................... 3 PHOTOTOXICITY ..........................3 WARNINGS AND CAUTIONS ......................4 CONTRAINDICATION ........................5 CLEANING AND DISINFECTION INSTRUCTIONS ............ 6 INSTALLATION OF DIGITAL UNIT ................6 KSL-H SERIES SLIT LAMP........................9 KSL-Z SERIES SLIT LAMP ........................9 CONTROLS......................
Page 4
Serial number Catalogue number Medical device Translation The Keeler Slit Lamp Digital IFU is designed and built in conformity with Directive 93/42/EEC, Regulation (EU) 2017/745 and ISO 13485 Medical Devices Quality Management Systems. Classifi cation: CE: Class I FDA: Class II The information contained within this manual must not be reproduced in whole or part without the manufacturer’s...
Brief description of the instrument This Keeler Slit Lamp can either be mounted onto a custom table top supplied by Keeler or can be mounted on a third parties table top (refraction unit) by suitably trained technicians. The Keeler Slit Lamp consists of 5 assemblies; illumination tower; observation system; XYZ translation base;...
Please note that the proper and safe functioning of our instruments is only guaranteed if both the instruments and their accessories are exclusively from Keeler Ltd. The use of other accessories may result in increased electromagnetic emissions or reduced electromagnetic immunity of the device and may lead to incorrect operation.
RF communications equipment. • Keeler digital KSL is not intended to be used with wireless technology. Do not plug a wireless dongle into USB port. It is advisable to disable wireless technology on the computer to prevent unauthorised access to the camera.
4. Safely dispose of used cleaning materials. 4. INSTALLATION OF DIGITAL UNIT For ‘Digital Ready’ variants of the Keeler Slit Lamp a Digital Camera Assembly (DCA) must be purchased separately, contact Keeler or your local distributor for details. Set up the Slit Lamp as directed in the Instructions For Use (IFU) EP59-70040 or EP59-70043, contact your distributor if you are unable to locate a copy of these instructions.
Page 9
KSL-H AND KSL-Z by KEELER 2. Fit the DCA to the rear of the magnifi cation block and tighten the securing knob to hold it in place. 3. Re-fi t the eyepiece assembly to the rear of the DCA and secure it by tightening the securing knob.
Page 10
KSL-H AND KSL-Z by KEELER 5. Remove the cover to the USB hub on the Slit Lamp base by undoing the small Phillips head screw and lifting off the cover. 6. Connect the shorter USB cable (3020-P-7107) to the socket on the underside of the DCA and to the appropriate socket on the USB hub located on the Slit Lamp base.
KSL-H AND KSL-Z by KEELER 4.1 KSL-H SERIES SLIT LAMP 1. Clip the auxiliary diffuser and blue fi lter in place on the Slit Lamp tower upright posts, above the level of the mirror. Align the slot to the fl attened post section.
KSL-H AND KSL-Z by KEELER 5. CONTROLS 5.1 BACKGROUND ILLUMINATION ADJUSTMENT KSL-H SERIES SLIT LAMP KSL-Z SERIES SLIT LAMP 5.2 EXPOSURE AND REVIEW BUTTONS Pressing these buttons provides adjustment of the camera exposure. Increase exposure time (milliseconds) Review freeze frame images forwards...
Slit Lamp has not been regularly serviced. The manufacturer’s warranty and terms and conditions are detailed on the Keeler UK website. The mirror, main illumination lamp and general ‘wear and tear’ are excluded from our standard warranty.
KSL-H AND KSL-Z by KEELER Emissions test Compliance Electromagnetic environment – guidance Voltage fluctuations / Complies flicker emissions IEC 61000-3-3 7.2 ELECTROMAGNETIC IMMUNITY Guidance and manufacturer’s declaration – electromagnetic immunity The Digital Camera Assembly is intended for use in the electromagnetic environment specified below.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Keeler Slit Lamp is used exceeds the applicable RF compliance level above, the Keeler Slit Lamp should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orientating or relocating the Keeler Slit Lamp.
KSL-H AND KSL-Z by KEELER Camera Assembly as recommended below, according to the maximum output power of the communications equipment. Rated maximum output Separation distance according to frequency of power of transmitter (W) transmitter (m) 150 kHz to 80MHz 80MHz to 800MHz 800MHz to 2.7GHz...
Page 17
KSL-H AND KSL-Z by KEELER Weight, packed (approx.) Digital Slit Lamp complete 20.0Kg, 75 x 54 x 45cm W x D x H Protection against ingress IPxO Class II ME equipment Insulation between mains parts and the functional earth provide at least two means of protection.
KSL-H AND KSL-Z by KEELER STORAGE CONDITIONS 1060 hPa °C °C 700 hPa TRANSPORT CONDITIONS 1060 hPa °C °C 500 hPa Vibration, sinusoidal 10 Hz to 500 Hz: 0.5g Shock 30 g, duration 6 ms Bump 10 g, duration 6 ms *This instrument does not meet the temperature requirements of ISO 15004-1 for storage and transportation.
KSL-H AND KSL-Z by KEELER 10. PACKAGING AND DISPOSAL INFORMATION Disposal of old electrical and electronic equipment This symbol on the product or on its packaging and instructions indicates that this product shall not be treated as household waste. To reduce the environmental impact of WEEE (Waste Electrical Electronic Equipment) and minimise the volume of WEEE entering landfills we encourage at product end of life that this equipment is recycled and reused.
Page 20
Contact Manufacturer Keeler Limited Clewer Hill Road Windsor Berkshire SL4 4AA UK Freephone 0800 521251 Tel +44 (0) 1753 857177 Fax +44 (0) 1753 827145 USA Sales Office China Office Keeler USA Halma China Group 3222 Phoenixville Pike 名称:沃迈(上海)机电有限 Building #50 公司...
Need help?
Do you have a question about the KSL-H and is the answer not in the manual?
Questions and answers