    Thank you very much for purchasing ULTIAMTE XL-K/D. Please read this Operation Manual carefully regarding the instructions for use, handling method, and maintenance and inspection prior to use and store it in a place where users can review it anytime. Contents IMPORTANT INSTRUCTIONS AND WARNING ·············································...
  • Page 3 A. GROUNDING INSTRUCTIONS 8. USE PROPER EXTENSION CORD. Make sure your extension cord is in good condition. When using an extension cord, be sure to use one heavy enough to carry the current your 1. In the event of a malfunction or breakdown, grounding provides a path of least resistance for product will draw.
  • Page 4: Important Instructions And Warning

    Auto Cruise Switch Motor Handpiece Indicator Turbine Adapter (option) Power Switch Air Turbine (option) Motor Connector Knee Controller (Ultimate XL-K) Turbine Adapter Connector Pedal Fuse Box Speed Control Knob Handpiece Stand Turbine / Motor Selector Switch Foot Pedal Connector (Ultimate XL-D)
  • Page 5: Set Up Of Control Unit

    5. Preset the maximum rotation speed with the Speed Control Knob. 6. Depress the Knee Controller for ULTIMATE XL-K and the Pedal for ULTIMATE XL-D, Fig.3 and the motor will run. The rotation speed can be variably controlled within the preset 3.
  • Page 6: Speed Limit Mechanism

    4. Speed limit mechanism 7. Memory Function For the use of a small-diameter round bur or When the Power Switch is turned on, the rotation direction made when the switch was last fisher bur at 40,000min (rpm) or more, pressing turned off are restored.
  • Page 7: Replacement Of Fuse

    If the cause is uncertain, return the product to an authorized properly, Knee controller or Pedal may be faulty. To cancel this check, press NSK’s service shop for inspection. Forward/Reverse Selector Switch again. (5) "in": Initializing function 10. Maintenance Mode...
  • Page 8: Handling Of Motor And Handpiece

    (3) Insertion of Chuck 12. Handling of Motor and Handpiece Turn until finger-tight Turn until finger-tight Turn until finger-tight Thinly apply oil before insertion. Open the (1) Insertion or Removal of Bur Chuck Control Ring, insert the dummy bur or The chuck is opened by turning the Chuck Control LOCK the bur in use into the chuck, and turn the...
  • Page 9: Handpiece Stand

    14. Specification Power Switch is faulty. Ask for repair. Power on the Control Unit without Control Unit pressing the Foot Pedal. When it does Ultimate XL-K NE231 not work definitely, take off the Foot Model Control Unit display Ultimate XL-D...
  • Page 10 Ask for repair of the turbine adapter The air turbine does used. section in the manual. kit is bent or pinched. and/or the Ultimate XL-K/D unit. not rotate. Turn on the power again, and If it operates normally, there is no Error code E5...
  • Page 11: Disposing Product

    16. Disposing Product Please consult with dealer from whom you purchased it about waste disposal.