Coolant Recommendations; Connecting The Liquid Cooling System; C Onnecting The C Oolant H Oses - Oxford Instruments ANDOR Newton CCD Hardware Manual

Table of Contents

Advertisement

Newton
3.8 c
R
oolant
ecoMMendations
• Coolant temperature: Refer to the temperatures specified in Technical Specifications for the specific model. Note that
cooling performance may be affected by distance between camera head and cooler.
• Recommended coolant: water or water/glycol mix depending on the ambient environmental temperature during operation.
De-ionized water (without additives) may be used as the coolant. Some mains supply water is heavily mineralized (i.e.
"Hard") which could cause deposits in the water circuit inside the camera. This can reduce the flow-rate and cooling
efficiency. The specified cooling performance of the camera can be achieved with coolant flow rates of 2 litres per
minute, the maximum recommended pressure of coolant circulating through the camera head is 2 bar (30 PSI).
3.9 c
onnecting tHe
An overview for connecting a liquid cooling system is outlined below- please refer to the information supplied with your
cooling system for further information.
3.9.1 c
onnecting tHe
1. Remove the black rubber protective caps (x2) if fitted.
2. Press the hoses onto the connectors on the camera.
3. Confirm the hoses are connected securely.
4. Connect the other ends of the coolant hoses to the cooling system- refer to the cooling system manual.
Version 1.4 rev 13 Oct 2017
l
c
s
iqUid
ooling
ysteM
c
H
oolant
oses
Coolant hose connectors for
connection to the cooling system
20

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ANDOR Newton CCD and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Andor newton emccdAndor newton soAndor newton sy

Table of Contents

Save PDF