Inogen GS-100 Technical Manual

Inogen GS-100 Technical Manual

At home stationary oxygen concentrator

Advertisement

INOGEN AT HOME STATIONARY OXYGEN CONCENTRATOR
TECHNICAL MANUAL
96-05070-00-01
Revision D
GUIDANCE TO AUTHORIZED SERVICE AND REPAIR CENTERS
This Manual applies to the following Inogen, Inc. products:
• Inogen At Home Oxygen Concentrator, model #GS-100
• Inogen At Home AC Power Cord, model #RP-109
CONFIDENTIAL and
INOGEN AT HOME OXYGEN
Page 1 of 15
96-05070-00-01 rev D
DCR 20-328
PROPRIETARY
CONCENTRATOR TECHNICAL MANUAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GS-100 and is the answer not in the manual?

Questions and answers

Summary of Contents for Inogen GS-100

  • Page 1 TECHNICAL MANUAL 96-05070-00-01 Revision D GUIDANCE TO AUTHORIZED SERVICE AND REPAIR CENTERS This Manual applies to the following Inogen, Inc. products: • Inogen At Home Oxygen Concentrator, model #GS-100 • Inogen At Home AC Power Cord, model #RP-109 CONFIDENTIAL and...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS INFORMATION FOR PROVIDERS OF THE INOGEN AT HOME ........3 1.1................3 AUTION AND ARNING TATEMENTS SETTING UP A PATIENT ON INOGEN AT HOME ............4 2.1..................4 ECOMMENDATIONS FOR 2.2....................4 YSTEM OMPONENTS 2.3.
  • Page 3: Information For Providers Of The Inogen At Home

    Proper care in relaying this information will not only enhance the user’s experience with the Inogen At Home, but will also protect the patient, prolong the life of the device, and help you avoid unnecessary service calls and complaints from users.
  • Page 4: Setting Up A Patient On Inogen At Home

    4. Plug into closest AC outlet. 5. Turn on the Inogen At Home by pressing the on/off button. 6. Use the + (increase flow) or – (decrease flow) buttons to adjust the Inogen At Home to the flow setting prescribed by the physician or clinician.
  • Page 5: Servicing The Inogen At Home

    The Inogen At Home delivers up to 5 L/min of 90% (+6%/-3%) oxygen. At each flow setting, the Inogen One generates a specific amount of oxygen (1 L/min per setting), which is delivered continuously to the patient. The device will deliver the same volume of oxygen regardless of the patient’s breath rate.
  • Page 6 14. Remove and replace Output Filter per section 3.1.6 below. 15. Unit is now ready for reuse. 16. Remove PPE and wash hands with soap and water for a minimum of 20 seconds. CONFIDENTIAL and INOGEN AT HOME OXYGEN Page 6 of 15 96-05070-00-01 rev D DCR 20-328 PROPRIETARY...
  • Page 7 Do not disassemble the Inogen At Home or any of the accessories or attempt any maintenance other than tasks described in this Technical Manual unless you have completed a training course through Inogen. Disassembly of the Inogen At Home or any of the accessories without NOTE proper training certification will void the product warranties.
  • Page 8 This filter is intended to protect the user from small particles in the product gas flow. The Inogen At Home includes an output filter conveniently located behind the removable cannula nozzle fitting. Inogen suggests that this filter be replaced between patients.
  • Page 9: Maintenance By The Patient

    3.2.1. Cannula Replacement The nasal cannula should be replaced on a regular basis. A single lumen cannula of 4 to 100 feet in length must be used. Inogen has certified its performance data with the Salter Laboratories 16SOFT cannula. Do not use total cannula tubing length exceeding 100 feet with the Inogen CAUTION At Home unless proper saturation has been verified by a clinician.
  • Page 10: Expected Service Requirements

    Additional replacement filters may be obtained from Inogen (RP-400). There are other types of filters inside the Inogen At Home that provide additional filtration. Maintenance of these filters is not required under normal operating conditions.
  • Page 11: Classifications

    IP 21 Degree of Safety for Application in Presence Not suitable for such applications of Anesthetic Gases: 5. INOGEN AT HOME ERROR ALARMS 5.1. DEFINITIONS ⚠ 5.1.1. Alarm LED: The alarm LED is positioned above a warning symbol ( ) on the user interface panel and is used to indicate when an alarm is occurring.
  • Page 12: Low Priority Alarms

    The compressor is at its end of life or there is a leak in the concentrator. concentrator’s Return the concentrator for service. compressor has reached its maximum speed. CONFIDENTIAL and INOGEN AT HOME OXYGEN Page 12 of 15 96-05070-00-01 rev D DCR 20-328 PROPRIETARY CONCENTRATOR TECHNICAL MANUAL...
  • Page 13: Medium Priority Alarms

    Check for kinked tubing and of specified flow listen for proper exhaust noises. If setting for 1 problem persists, return the minute. concentrator for service. CONFIDENTIAL and INOGEN AT HOME OXYGEN Page 13 of 15 96-05070-00-01 rev D DCR 20-328 PROPRIETARY CONCENTRATOR TECHNICAL MANUAL...
  • Page 14: High Priority Alarms

    (see below) encountered an the device. See Error Code table below. error and is shutting down. CONFIDENTIAL and INOGEN AT HOME OXYGEN Page 14 of 15 96-05070-00-01 rev D DCR 20-328 PROPRIETARY CONCENTRATOR TECHNICAL MANUAL...
  • Page 15: Error Code Table

    5.7. CONTACTS for MORE INFORMATION Inogen, Inc. 301Coromar Drive Goleta, CA 93117 1-866-765-2800 http://www.inogenone.com http://www.inogen.net CONFIDENTIAL and INOGEN AT HOME OXYGEN Page 15 of 15 96-05070-00-01 rev D DCR 20-328 PROPRIETARY CONCENTRATOR TECHNICAL MANUAL...

Table of Contents