Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Table of Contents 1 Safety Instructions 6.3.7 Settings of the Digital Outputs 1.1 Safety Instructions for the Device 6.3.8 Settings of the Function Buttons 1.2 Safety Instructions for Measuring Probes 6.3.9 Display of the Error Memory 1.3 Safety Symbols...
Page 4
Industrial Gaussmeter IGM11 Operating Instructions MAGSYS systeme 9.12.2 Main Commands 10 Table of Error Messages 9.12.3 Peak Value Function 11 Unit Conversion Table 9.12.4 Probe Functions 12 Technical Data 9.12.5 Memory Functions 13 Declaration of Conformity 9.12.6 Device Functions 14 Warranty and Copyright 9.12.7 Interface Functions...
Page 5
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Table of Illustrations Display Signal Input Minus Switched Basic Assembly of a Hall Probe Signal Output Plus Switched Flux Line Characteristics of an NdFeB Induction Disk 14 Signal Output Minus Switched...
Do not replace any parts and do not modify this product without prior and written consent by MAGSYS. Do not service this device. Return the device to MAGSYS magnet systeme GmbH or to your local supplier for repair and maintenance to ensure that all safety features remain.
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 1 Safety Instructions 1.3 Safety Symbols Safety symbols can be found on various spots of the device. Before using this connection or function, read the corresponding reference in the manual.
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 2 Brief Introduction 2 Brief Introduction Measurements with the gaussmeter utilize the Hall effect as principle of measurement. A Hall sensor is a symmetric current-carrying semiconductor. A magnetic field running perpendicular through this element generates an asymmetry on the chip and thereby creates an output voltage which, at first approximation, is proportional to the product of magnetic field strength and forced current.
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 2 Brief Introduction 2.3 Measuring Unit The gaussmeter shows measuring values in physical units of the SI-system and of the Gauss-CGS system (preferably used in North America). You can adjust the unit in the set-up menu.
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 2 Brief Introduction 2.5 Display The following shows a typical example of the display. Lifesign Measured value with unit Status of the digital Mea suring range in- / outputs and mode Figure 1 Display 2.6 Status Display...
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 3 Function of the Gaussmeter 3 Function of the Gaussmeter 3.1 The Hall Effect 3.1.1 Linear Properties of the Hall Probe The measurement is based on the deflection of charge carriers in a magnetic field inside a conductor.
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 3 Function of the Gaussmeter This implies: ⋅ ⋅ Hall ⋅ with Charge carriers Elementary charge of the electron (1.6022 × 10 Width of the path on which the electrons move...
Page 13
Hall effect. This becomes noticeable as a flux density modulated resistance. For measuring fast magnetic impulses, the device must have sufficiently high dynamics in order to equalize this effect. The gaussmeter IGM11 is optimized for this operating mode. 3.1.2.4 Temperature Dependence of the Sensitivity Due to the large band gap of the used Hall sensors the temperature dependence of the probe sensitivity is low.
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 3 Function of the Gaussmeter 3.2 Measurement Details The used Hall sensors contain a very small active semiconductor area ranging at approx.100 µm. The local resolution of this mesasuring method is thus rather high.
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 3 Function of the Gaussmeter Figure 4 Field Strength Pattern of an NdFeB Induction Disk The diagram in figure 4 shows the measurement with a Hall probe which is moved in parallel to the surface of the magnet within a distance of 1mm to the measuring surface.
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 3 Function of the Gaussmeter As B as well as B is measured in the unit Tesla, the magnetic field measured on the outside is often mixed up with the remanence.
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 3 Function of the Gaussmeter 3.2.4 External Static Magnetic Fields Particularly in sensitive measuring ranges, an external static magnetic field, as e.g. the Earth’s field, can already become clearly noticeable. These external magnetic fields lead to a corruption of the measuring result.
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 4 Control Elements and Connections 4 Control Elements and Connections 4.1 Front Panel Overview Figure 5 Front Panel Output of the measured value Display Status info ENTER-button Control of the set-up menu...
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 4 Control Elements and Connections 4.2 Ports Overview Probe Port EIA-232 / EIA-485 Port (Probe) 24V IO Ports Power Network Supply Port Figure 6 Ports Figure 7 Ports Top PROBE...
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 4 Control Elements and Connections Figure 8 Ports Bottom For add-ons Interface Power Supply Do not use Port for PLC Port for power RJ45 port for the LAN signals supply for the...
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 4 Control Elements and Connections 4.5 USB Interface The transfer of the measuring values and the entire gaussmeter control are possible via the installed standard USB interface on the front panel.
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 4 Control Elements and Connections 4.6 LAN Interface The transfer of the measured values and the entire control of the gaussmeter are possible via the standard LAN interface. The interface supports both the 10-Mbit/s-Ethernet (10BaseT) and the 100- Mbit/s-Ethernet (100BaseTX).
Page 23
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 4 Control Elements and Connections Please notice: Access authorization is not supported. This simplifies the • communication with the control programs. Only one connection is possible at a time. If one connection has •...
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 5 Operation 5 Operation 5.1 Buttons The required functions are selected and triggered via two buttons. More complex parameter settings can be made via menu functions in the set-up menu. Button functions in detail: With this button you confirm the entry, select the menu item or a numerical digit.
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 5 Operation 5.2 Display The following shows a typical display sample. Lifesign Status Signal Error Message Measuring Value with Unit Status of the Digital Measuring Range Inputs- / Outputs and Mode...
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 5 Operation 5.3 Status Display Apart from the measuring value the display shows various status information. The bottom left area of the display shows the currently selected measuring range with the currently selected unit.
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 5 Operation 5.5 Measuring Range The gaussmeter has 4 measuring ranges. The range limit value is shown at the bottom left area of the display. The range limit values depend on the ranges themselves and on the selected measuring unit.
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 5 Operation 5.6 Measuring Unit Select the desired measuring unit in the set-up menu. The measurement is carried out either in the unit Tesla, Gauss, kA/m or in Oersted. The display shows the selected measuring unit. The respective range limit value changes according to the selected measuring unit.
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 5 Operation The following shows the accuracies for DC-field measurements. Field Strength Accuracy (1σ) ≤ 1.5 T ±0.5 % > 1.5 T ±1.0 % 5.7.2 AC Field Measurement In the operating mode AC field measurement the effective value (RMS) is calculated from the determined AC field contributions.
Page 30
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 5 Operation The error of the AC field measurement is composed of the error of the DC field measurement and the error of frequency and shape factors. Error DC Field Measurement (1σ)
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 5 Operation 5.8 Peak Value Measurement The device has a fast DC field peak value detection. 5.8.1 Fast Peak Value Detection The fast detection of the maximum values of DC fields is required for short magnetic impulses, as they are generated in e.g.
Page 32
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 5 Operation The error of the fast peak value measurement consists of the error of the DC field measurement and of a frequency factor that depends on the measuring range. Error DC Field Measurement (1σ)
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 5 Operation 5.9 Probe Data For a correct measurement the gaussmeter must always know the necessary probe data. A connected probe contains a parameter memory which stores the probe parameters, the serial number and the name.
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 6 Set-up Menu 6 Set-up Menu The following set-up options can be adapted individually in order to be able to use the device for each application in an optimal way. By pressing both buttons for approximately 2 seconds, the set- up menu appears on the display.
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 6 Set-up Menu 6.3 Settings 6.3.1 Overview All settings are displayed. You can scroll through the display with both buttons. This is a display example. Display of measuring value and unit...
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 6 Set-up Menu 6.3.2 Measurement The following options can be set individually, in order to use the device optimally for each application. Measurement Settings DC Measurement Measurement of the DC field...
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 6 Set-up Menu 6.3.3 Parameters of the Serial Interface EIA-232 Settings for the serial interface EIA-232 (always 8 data bits are used) 1200 bps Rate 1200 bit/s 38400 bps 38400 bit/s...
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 6 Set-up Menu 6.3.5 Parameters of the LAN Interface Settings of the LAN interface nnn.nnn.nnn.nnn Settting of the IP address nnn.nnn.nnn.nnn Setting of the IP net mask 6.3.6 Settings of the Digital Inputs...
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 6 Set-up Menu 6.3.7 Settings of the Digital Outputs Settings of the digital output 1 With the following function RESET IF Logic function the output will be reset. With the following function SET IF the output will be set.
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 6 Set-up Menu 6.3.8 Settings of the Function Buttons Settings for button 1 (top / enter) Settings for button 2 (bottom / next) This button switches between RANGE Range the measuring ranges.
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 7 Power Supply Connection 7 Power Supply Connection Via connector J2 the device is supplied with DC power. The operating voltage must be between 12 V and 24 V. The power consumption is 200 mA maximum at a voltage of 12 V.
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 8 Signal Interfaces 8 Signal Interfaces 8.1 Introduction The device has three digital inputs and four digital outputs. The inputs and outputs are designed for a maximum voltage of 24 V. The inputs and outputs are insulated and potential free.
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 8 Signal Interfaces 8.2 Wiring of the Inputs Option1: The external control switches the external positive potential. Insulated DC input Used as current sinking inputs Special features Do not connect!
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 8 Signal Interfaces 8.3 Wiring of the Outputs Option 1: The gaussmeter switches the positive potential provided by the customer. The load is switched against the external minus pole. Insulated DC output Used as current source outputs.
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces 9 Serial Interfaces 9.1 Introduction Via the provided serial interfaces as well as via the LAN interface, all functions of the gaussmeter can be controlled by an external device (e.g. personal computer).
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces 9.2 Connecting the Gaussmeter to an External Controller The EIA-232 connector and the EIA-485 connector use the 9-pole connector of the type DE-9 on the top right side of the device (J4).
Page 47
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces EIA-485 Connector Connector Input/Output Function Data line A(-) Bus connection negative Data line B(+) Bus connection positive 5 V (max. 50 mA) Voltage supply for external terminal...
You can either establish a point-to-point connection with a control unit or you can access several devices via a bus structure. The IGM11 always operates as slave unit at this bus. This means it will transmit data only after a request.
The basic network should be provided for those devices with active bus terminal. With the IGM11 the resistances of the basic network can also be mounted in the connector casing. The 5 V connection provided for the basic network at Pin 8 of the connector J4 can thus be loaded with a maximum of 50 mA.
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces If the power supply of the interface is 5 V, we recommend 390 resistors (with 15 V approx. 1 k to power supply and 390 to common ground).
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces The potential equalization line and the signal cable should be layed in a • distance as small as possible to each other. Avoid cable extensions. • Conduct lines as tight as possible to grounded areas.
You can connect one master and a maximum of 31 slaves to the bus. The single slaves are selected by the master via an address character. An IGM11 can never take the transmission initiative automatically. The communication is carried out in half- duplex operation.
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces 9.4.2 Data Encoding The data telegram is encoded as follows: STX (Start of Text): 1 byte, ASCII character 0x02h (decimal 2) • LNG (Telegram length): 1 byte, containing the telegram length as binary •...
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces 9.4.5 The Address Byte (ADR) The address byte transmits the address number and further control information. The single bits in the address byte are specified as follows: Figure 27 EIA-485 Address Structure The device number will be transmitted in bit 0 to bit 4.
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces The IGM11 checks the following: starting pause before the master telegram, • parity of the single characters, • BCC check sum, • character frame of the single characters •...
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces In the set-up menu you can select the point-to-point connection through the parameter value P2P. 9.6 Samples of Data Transfer EIA-485 BUS Mode with bus address 01 Master: <STX><0x08><0x01>*idnt?<0x09>...
9.7.1 Operating Mode SCPI In this operating mode SCPI commands are evaluated and replied depending on the command type. The IGM11 will only reply on request. The command line must be terminated with the character <LF>. The reply is terminated with the characters <CR><LF>.
9.7.2 Operating Mode SHORT In this operating mode simple one-character commands are evaluated and replied depending on the command type. The IGM11 will only reply on request. The command line must start with <STX> and terminate with <ETX>. The reply will also be enclosed between <STX>...
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces 9.7.3 Operating Mode FLOW In this operating mode, the current measuring value is transmitted every 100ms. The measuring value is enclosed within <STX> and <ETX>. The syntax is: <STX>snnnn.<ETX>...
Abort Other control characters, e.g. tabulator (0x09h), can be sent from an external controller in order to get a clear format. They will be ignored by the IGM11. 9.10.2 Introduction to the SCPI Language The programming language SCPI (Standard Commands for Programmable Devices) defines the way a measuring device (here the gaussmeter) can communicate with a controller.
Page 63
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces 9.10.2.3 Command Separator “;“ Several commands within the same command line string are separated by a semicolon. By means of a semicolon the indicated path is not changed. The two following statements have the same meaning: CMD:LOW 1;:CMD:HIGH 5...
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces 9.10.3 SCPI Data Types The SCPI data language defines different data formats that are used in a program message and in a reply message. An SCPI device can normally accept commands and parameters in different formats.
The status system records different device states in several register groups. The individual messages are grouped in the several registers. One bit of these registers is related to one message respectively. The IGM11 has three different event registers: measurement event register •...
Page 66
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces :STATus:PRESet deletes the release register for the error data. When querying an event register, the device returns a decimal value that corresponds to the sum of the binary place values of all bits that are set in this register. For setting the bits in a release register, a decimal value is transmitted that corresponds to the sum of the binary place values of all bits set in this register.
Page 67
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces 9.10.4.5 Definition of the Register Bits Definition of the bits in the measurement event register Decimal Description Definition Value Overflow Range overflow during measurement Data available Measurement terminated. Data available.
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces 9.11 Summary of SCPI Commands The following conventions are used in the SCPI command syntax: Optional key words or parameters are issued in square brackets [ ] •...
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces 9.11.3 Peak Value Functions Reset the current peak values :PEAK:NULL Query the higher values of minimum and :PEAK:FAST:READ? maximum of the fast peak value detection Query the minimum values of the fast peak...
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces 9.11.6 Memory Functions :PAR:SAVE Store the set parameters :PAR:SAVW? Store the set parameters and queries the status :PAR:DEFA Default parameter settings 9.11.7 Serial Interfaces :SERI:COMA:BAUD[?] {1200|2400|4800|9600|14K Bit rate of EIA-232...
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces 9.12 SCPI Command Reference 9.12.1 Control Commands 9.12.1.1 *RST Description: Reset Command The gaussmeter is reset to default values. The internal parameters are set to default values. A complete reset of the device is carried out.
Page 72
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces 9.12.1.4 *ESR[?] Description: Standard Event Status Enable Query Reset or query the standard event release register. The gaussmeter returns the bit pattern of the register as a decimal value.
Page 73
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces 9.12.1.7 *OPC Description: Operation Complete Command) Set the bit "Operation complete" (bit 0) in the standard event register after the command has been completed. Mode: Command Parameter:...
Page 74
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces 9.12.1.10 :STAT:PRESet Description: Reset the measurement event register and the standard event register. All relevant bits are set to 0. Mode: Command Parameter: none Response: none Value: not relevant...
Page 75
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces 9.12.1.13 :STAT:MEAS:ENABle[?] Description: Set or query the measurement event release register. The gaussmeter returns the bit pattern as a decimal value. Mode: Command and query Parameter: {<Value>} within the range of 0..255...
Page 76
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces 9.12.2 Main Commands 9.12.2.1 :MODE[?] Description: Set or query the DC or AC field operating mode. Mode: Command and query Parameter: {DC|AC|PEAK} DC operation mode AC operation mode...
Page 77
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces 9.12.2.3 :RANGe:SET Description: Set the measuring range of the gaussmeter. Mode: Command Parameter: {0|1|2|3} Most sensitive range Least sensitive range Response: none Value: *RST Example: send :RANG:SET 2<LF>...
Page 78
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces 9.12.2.5 :READ? Description: Query the current measuring value. Identical with “MEAS?” Mode: Query Parameter: none Response: Measuring Value Value: not relevant *RST Example: send :READ?<LF> receive 2.546313e-01<CR><LF> 9.12.2.6 :MEAS? Description: Query the current measuring value.
Page 79
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces 9.12.2.7 :UNIT[?] Description: Set or query the physical unit of the measuring value. Mode: Command or query Parameter: TESL|APM|GAUS|OE|G|T|A|O TESL Unit is Tesla Unit is A/m GAUS...
Page 80
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces 9.12.3 Peak Value Function 9.12.3.1 :PEAK:NULL Description: Reset the current peak values. Mode: Command Parameter: none Response: none Value: not relevant *RST Example: send :PEAK:MODE NULL<LF> 9.12.3.2 :PEAK:FAST:READ?
Page 81
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces 9.12.3.4 :PEAK:FAST:READ:MAX? Description: Query the maximum values of the fast peak value detection. Mode: Query Parameter: none Response: Measuring value Value: not relevant *RST Example: send :PEAK:FAST:READ?<LF>...
Page 82
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces 9.12.3.7 :PEAK:SLOW:READ:MAX? Description: Query the maximum value of the slow peak value detection. Mode: Query Parameter: none Response: Measuring value Value: not relevant *RST Example: send :PEAK:FAST:READMAX?<LF> receive 9.11111e+00<CR><LF>...
Page 83
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces 9.12.4.3 :PROB:TYPE? Description: Query the probe type. Mode: Query Parameter: none Response: Probe type Type Type Transversal standard Axial standard Calibration probe Transversal high field Axial high field...
Page 84
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces 9.12.5 Memory Functions 9.12.5.1 :PAR:SAVE Description: Save parameters, units and ranges in non-volatile memory. Mode: Command Parameter: none Response: none Value: not relevant *RST Example: send :PAR:SAVE<LF> 9.12.5.2 :PAR:SAVW? Description: Save parameters, units and ranges in non-volatile memory.
Page 85
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces 9.12.6 Device Functions 9.12.6.1 :SN:UNIT? Description: Query the serial number of the device. Mode: Query Parameter: none Response: Serial number Value: not relevant *RST Example: send :SN:UNIT?<LF>...
Page 86
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces 9.12.6.4 :SN:CALI? Description: Query the calibration info. Mode: Query Parameter: none Response: Calibration info Value: not relevant *RST Example: send :SN:CALI?<LF> receive CAL 01SEP12 / 01SEP14<CR><LF> 9.12.6.5 :SYST:ERR? Description: Query the error queue.
Page 87
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces 9.12.7 Interface Functions 9.12.7.1 :SERI:COMA:BAUD[?] Description: Set or query the bit rate of the EIA-232 interface. Mode: Command or query Parameter: {1200|2400|4800|9600|14K4|19K2|38K4} Response: like parameter Value: not relevant...
Page 88
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces 9.12.7.4 :SERI:COMA:MODE[?] Description: Set or query the operating mode of the EIA-232 interface. Mode: Command or query Parameter: {SCPI|SHOR|FLOW} Response: like parameter Value: not relevant *RST Example: send :SERI:COMA:MODE SCPI<LF>...
Page 89
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces 9.12.7.7 :SERI:COMB:ADR[?] Description: Set or query the bus address of the EIA-485 interface. Mode: Command or query Parameter: {<Value>} within the range of 0..31 Response: like parameter...
Page 90
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces 9.12.7.9 :IP:MAS1[?] .. :IP:MAS4[?] Description: Set or query the IP sub net mask of the LAN interface. MAS1 is the most significant part of the IP subnet mask.
Page 91
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces 9.12.7.12 :DGIO:SET:IN :FUNC[?] Description: Set or query the function of a digital input. x is for inputs 1 to 3. Mode: Command or query Parameter: NO|HOLD|NLPK|NULL Parameter...
Page 92
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces 9.12.7.14 :DGIO:SET:OUT :FUNC[?] Description: Set or query the function of each output. x is for outputs 1 to 4. Mode: Command or query Parameter: NO|IFHI|IFLO|IFIN|IFOU|IFOV|IFRN Parameter Function No function assigned to this output...
Page 93
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 9 Serial Interfaces 9.12.7.16 :DGIO:SET:OUT :LOW[?] Description: Set or query the lower limit value for the limit value check of each output. x represents outputs 1 to 4. Only relevant for the selected compare operation.
Page 94
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 9 Serial Interfaces 9.12.7.18 :KEYS:TOP:FUNC[?] Description: Set or query the function of the upper button (ENTER) on the front panel. Mode: Command or query Parameter: NO|RANG|PKNL|NULL Parameter Function No function assigned to this key...
Page 95
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 10 Table of Error Messages 10 Table of Error Messages Errors inside the device are stored in the error queue. It can be read out via an SCPI command or via the set-up menu.
Page 96
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 10 Table of Error Messages EIA232 buffer overflow Buffer overflow on the serial interface EIA-232. If necessary, reduce the bit rate. Check the command. EIA485 parity error Character parity error on the serial interface EIA-485.
Page 97
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 11 Unit Conversion Table 11 Unit Conversion Table This table shows the relationship between the displayed measuring values. Displ Dimension Unit Conversion Mag. flux Tesla ⋅ density B Flux density Gauss −...
Page 98
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 12 Technical Data 12 Technical Data General Power Supply Supply voltage feed 11.4 V to 26.4 V DC; Current consumption (nominal): 120 mA with 12 V; 80 mA with 24 V (with display illumination) 70 mA with 12 V;...
Page 99
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 12 Technical Data Permanent output EIA-485 Norm EIA-485 Connector DE-9 plug Bit rate 1200 to 38400 bit/s; parametrizable Parity Even Function Data output via bus or point-to-point connection Bus terminal Possible via external resistors in the connector ®...
Page 100
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 12 Technical Data Logic 1 (min.) 20 V DC with maximum current Logic 0 (max.) 0.1 V DC with 10 k load Function Individually selectable via parametrization Output shunt Possible Connector Via common12-pole IO connector;...
Page 101
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 12 Technical Data Units Tesla Gauss Oersted Ampere/meter Meas. ranges 4.5 T 45 kG 45 kOe 3800 kA/m [10T] [100 kG] [100 kOe] [8500 kA/m] (1 mT) (10 G) (10 Oe)
Page 102
Date / Datum M. Kopka Dipl.Ing. CEO For further information, please contact your local MAGSYS magnet systeme sales office, agent or distributor, or MAGSYS magnet systeme GmbH, Rohwedderstr. 7, 44369 Dortmund, Germany. www.magsys.de Für weitere Informationen kontaktieren Sie bitte Ihr örtliches MAGSYS magnet systeme Vertriebsbüro, Handelsvertrete, Händler oder direkt MAGSYS magnet systeme GmbH, Rohwedderstr.
Page 103
The above-mentioned measures are the only and exclusive given hint. We will be striving to fix any mistake as fast as measures on the side of the buyer. MAGSYS magnet systeme possible. GmbH is not liable for direct, indirect, particular damages...
Page 104
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 15 Index 15 Index :SERI COMA BAUD ............87 MODE ............88 *CLS ..............71 PARY ............87 *ESE..............71 STOP............87 *ESR..............72 COMB *IDN ..............72 ADR............89 *OPC............72, 73 BAUD ............88 *RTS..............71...
Page 105
Industrial Gaussmeter IGM11 Operating Instructions magne t MAGSYS systeme Chapter 15 Index Data Transfer EIA-232......... 60 DC Field Measurements ........28 DC Field/AC Field..........28 Operation............24 Declaration of Conformity ......... 102 Output Data............65 Device Functions ........... 69, 85 Digital Inputs ............38 Digital Interfaces ..........
Page 106
Gaussmeter IGM11 Operating Instructions MAGSYS systeme Chapter 15 Index Table of Contents ..........3 Table of Illustrations ..........5 Units ..............97 Technical Data ............98 USB Interface............21 Telegram Exchange ..........54 User Data............62 Telegram Length ..........53 Telegram Syntax..........52 Wiring Details............. 50 Wiring of the Inputs ..........
Need help?
Do you have a question about the IGM11 and is the answer not in the manual?
Questions and answers