OPERATION MANUAL
Professional Thermocouple
Thermometers

Advertisement

loading
Need help?

Need help?

Do you have a question about the 94915 and is the answer not in the manual?

Questions and answers

Summary of Contents for EBCHQ 94915

  • Page 1 OPERATION MANUAL Professional Thermocouple Thermometers...
  • Page 2 The (three input) offers fast response and laboratory accuracy. This data logging thermometer is suitable for most temperature measurement and data logging applications. These thermometers work with two K-type thermocouples and one IR temperature probe as temperature sensors, and it offered 0.1 °C / 0.1°F resolution. It internal memory can keep up to 18,000 records per channel.
  • Page 3 Specification General Specifications Operating Temperature 0 °C to 50 °C (4 °F to 122 °F) Noncondensing Storage Temperature 10 °C to 50 °C (14 °F to 122 °F) Temperature Range: K type thermocouples -200 °C to 1372 °C (-328 °F to 2501 °F) IR Temperature -30 °C to 550 °C (-22 °F to1022 °F) Temperature accuracy...
  • Page 4 Button 1. Press to turn the thermometer on or off. 2. Press to step through the maximum, minimum, and average readings. When readings, shows the maximum, minimum, and average of the logged readings. To exit MAN/MIX/AVG mode, press the button for 3 seconds to return to normal operation.
  • Page 5 8. Press to start recording and press again to stop recording .if enter a Setup option .scroll to the Setup option you want to change. Press to decrease the displayed setting. 9. Press to enter a Setup option. Press again to store the displayed setting in memory.
  • Page 6 Using the thermocouple (s) Plug the thermocouple(s) into the T1or T2 input connector(s). Press to turn on the thermometer. After 1 second the thermometer displays the first reading. If no thermocouple is plugged into the selected input or the thermocouple is "open," the display shows "- - - -". Press and hold this button 3 second to turn off the Power.
  • Page 7 Entering or Exiting Setup When the thermometer is in Setup mode, the display shows SETUP. Press button for 5 seconds start or exit Setup. Changing a Setup Option 1. Press to scroll to the setup option you want to change. 2.
  • Page 8 Sample Rate The tertiary display shows the sample rate of the temperature data recording. When the meter is power ,the default sample rate is 1 second.. The tertiary display show “ 0 0 0 1”. The order is minute---second. Press adjust the order.
  • Page 9 2. Set the time. The adjust order is year--- month ----day---- hour --- minute. Press to adjust the order. Press until the setting you want to use appears on the display. If the time is adjusted, Press to store the new setting in memory. Erase Memory Operation When one want to clear the memory, enter CLR setup operation, Press...
  • Page 10 Displaying Temperatures 1. Press to select the correct temperature scale. 2. Hold or attach the thermocouple(s) to the measurement location. The temperature reading appears in the primary display. Notes The display shows "- - - -" when a thermocouple is not connected.
  • Page 11 Record Data Press to turn on the thermometer. After 1 second the thermometer displays the first reading (18000). The reading is the number of the thermometer that can record datas. When memory full, thermometer displays the first reading “FULL”, please erase Memory Press , the meter will start recording, and press again will stop recording.