Page 1
Retail Series Calibration Manual January 5, 2018 PN 170120 Rev A...
Page 2
All information contained within this publication is, to the best of our knowledge, complete and accurate at the time of publication. Rice Lake Weighing Systems reserves the right to make changes to the technology, features, specifications and design of the equipment without notice.
Retail Series Calibration Manual Counter-Top Scales RS-130/RS-160 Follow the steps below to complete the calibration process. 1. Turn on the scale while pressing until 6.3 displays. CHANGE 2. Press and hold until SETUP displays. ZERO 3. Press to enter calibration mode. The WEIGHT display flashes CAL-1 and shows UNIT 1.
Retail Series Calibration Manual Astra ow to calibrate the Ast Follow the steps bel ra scale. 1. Turn the scale off. 2. Turn the scale on while pressing . Release after the second beep. C00 TEST displays. 3. Press followed by .
Retail Series Calibration Manual SR-2000 Follow the steps below to calibrate the SR-2000 scale. 1. Press 2. Enter 495344 and press 3. Select from the list. 4. With no weights on the platter, press to initialize the calibration values. The initial A/D value should be 20000 ± 1.
Retail Series Calibration Manual The A/D value should be 80000 ± 2. 8. Remove the weight. 9. Save the calibration data by pressing the located underneath the platter. Memory Switch Figure 1-9. Uni-5 and Uni-7 Memory Switch location 10. Press and press 11.
Retail Series Calibration Manual Hanging Scales BC-4000L2H Follow the steps below to calibrate the BC-4000L2H scale. 1. Turn the scale off. 2. Turn the scale on while holding down the key until it beeps twice. C00 TEST MODE displays. 3. Enter a using the keypad and press the .
Page 14
Retail Series Calibration Manual 4. Select from the list. CALIBRATION 5. To select Single Range or Multi Range, enter and press as needed. 495344 SINGLE MULTI 6. With no weight on the platter, press to initialize the A/D values. The initial value should be 20000 ± 1.
Page 16
Retail Series Calibration Manual 11. Press the in the upper right corner. Figure 2-7. 12. Verify the weight readings in operation mode. Visit our website www.RiceLake.com...
Retail Series Calibration Manual IP-EMZ/OMNI-4000 Follow the steps below to calibrate the IP-EMZ scale and the OMNI-4000 auto wrapper. 1. Press Figure 3-2. 2. Enter 495344 and press 3. Select from the list. 4. With no weight on the platter, press to initialize the A/D values.
Retail Series Calibration Manual 10. Press the in the upper right corner. Figure 3-6. 11. Verify the weight readings in operation mode. Uni-7 RP Follow the steps below to calibrate the Uni-7 RP scale. 1. Press 2. Enter and press to access hidden menus.
Retail Series Calibration Manual 9. Save the calibration data by pressing the located behind a white plastic plug on the side of the remote Memory Switch scale base. Figure 3-10. Uni-9 RP Memory Switch location 10. Press the in the upper right corner.
Page 24
Retail Series Calibration Manual 2. Enter and press the key. 495334 TEST MODE Figure 3-15. 3. Select from the list. CALIBRATION Figure 3-16. 4. With no weight on the platter, press the key to initialize the A/D values. ZERO Figure 3-17.
Page 26
Retail Series Calibration Manual 5. Press , to select the range needed. SINGLE MULTI 6. With no weight on the platter, press to initialize the A/D values. The initial value should be 20000 ± 1. 7. Place scale capacity weight on the platter and press The A/D value should be 80000 ±...
Page 28
Specifications subject to change without notice. Rice Lake Weighing Systems is an ISO 9001 registered company. 230 W. Coleman St. • Rice Lake, WI 54868 • USA U.S. 800-472-6703 • Canada/Mexico 800-321-6703 • International 715-234-9171 • Europe +31 (0)26 472 1319...
Need help?
Do you have a question about the Retail Series and is the answer not in the manual?
Questions and answers