8.
Specifications
8.1 Probe
Type
Acoustic working
frequency
Technology
Dimensions
Weight
Scanning width
Probe materials
Acoustic output
Applicable system
Classification
Cleaning
Disinfection
Sterilization
Operating conditions:
Ambient temperature
Contact surface temperature : max. 42°C
(Temperature of examinee)
Relative humidity
Storage conditions:
Temperature
Relative humidity
: L44 Probe
: 6.5MHz
: Linear Array Probe
: See Fig. 9.
: Approx. 0.43kg
(Including cable and connector)
: 38mm
: Biocompatible allergy free components
: According to IEC60601-2-37
(See Main Unit manual.)
: Depending on production and upgrade status
for detailed information, contact a service
support.
: MDD classification IIa.
: Applicable detergents are listed in the
suppliers list
: Applicable disinfectants are listed in the
suppliers list
: ETO gas sterilization
Plasma sterilization
: +10 - +35°C
: 30 - 85%
: -10 - +55°C
: 10 - 95%
(subject to no condensation)
( 21 )