Failure to comply with any of the instructions contained in this manual can result in damage to the unit or improper installation, rendering the pass-through inoperable. Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
CleanMount® designs use lag bolts on either side of the chamber, thus requiring extra clearance in the wall cut-out Figure 1: (from left) Pass-through with Standard Mounting bracket and Pass-through with CleanMount® Mounting Bracket Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
Page 5
Finish the wall cut-out with an appropriate substrate and finishing material. As a general rule, the wall cut-out should have the same grade of finishing as the wall it is installed in (e.g. cleanroom-grade materials). Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
Section 2.6 Door Alignment. Figure 4: Pass-through with standard mounting brackets, fastened to the wall Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
Install the bracket covers by snapping each of them into the the body of the pass-through bracket clips (See Figure 8) Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
Page 8
Figure 8: Snap the bracket covers back into place to to the “clean”-side bracket, sandwiching the wall hide the lag bolts and other hardware between the brackets Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
10. Fasten the bracket to the wall surface only. Do not over-tighten the fasteners as this may bend the bracket and create gaps. 11. Test the pass-through doors again for proper functioning. Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
CleanMount® brackets (flush-mount), complete the installation instructions outlined in Section 2.4. Align the floor ramps with the bottom of the pass-through so that the two notches on the rear of the ramp hook onto the studs below the pass-through doors. Push the ramp in until it is flush with the surface of the pass-through and then set it down.
Protruding Exhaust Flange or Filter Housing If the pass-through arrived with either an exhaust flange or an ULPA filter housing preinstalled, these features will need to be detached from the pass-through prior to installation to allow the pass-through to fit through the wall cut-out.
Fold the wiper between strokes, and replace with a clean wiper often. Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
Terra Universal. All claims must be made in writing within 90 days of the date the product was shipped. Any claims not made within this time limit shall be deemed waived by the customer.
Need help?
Do you have a question about the Pass-Through and is the answer not in the manual?
Questions and answers