6.7
ANALYSIS ERRORS
6.7.1
Analysis Error
Monitor Timing: When measuring/analyzing
Monitor Method:
Coagulation
Error code
ERROR0001
(Temp Error)
ERROR0002
(Slight Coagulation)
ERROR0004
(Analysis Time
Over)
ERROR0008
(Coag. Curve Error)
ERROR0016
(Turbidity Level
Over)
ERROR0032
(No coagulation)
ERROR0064
(Measurement
Error)
ERROR0100
(Range Over)
ERROR0200
(Fbg Range Over)
Chromogenic/Immunoassy
Error code
ERROR0401
(Trans Light Low)
ERROR0402
(Trans Light High)
ERROR0404
(No Linearity)
ERROR0408
(Reaction Curve Error)
ERROR0416
(Range Over)
ERROR0528
(No Polynomial
Adjustment)
CA-500 Series S/M
Status
Temperature error occurred when
sample was being analyzed.
Detected coagulation curve shows
extremely small change.
Coagulation did not finish within
preset detection time or within Auto
Extension Mode detection time.
1. Curve shows a drop in some area.
2. Curve shows too steep in some
areas.
3. As buffer was cool, clotting time
was extended by the influence of
air bubbles. (Fbg)
Turbidity was too high to analyze.
Clotting could not be detected.
Due to any problem before photo
detection, such as insufficient sample
plasma, instrument error, insufficient
reagent, Mechanical Stop Switch
operation, analysis could not be done.
The detection time at 50% point is
exceeding the preset time, i.e.
coagulation is drastic.
Status
Transmitted Light was too low.
Transmitted Light was too high.
Reaction Curve had no linearity.
Reaction Curve changed to the
opposite direction.
Antigen became range over (with
Immunoassy only).
Polynomial adjustment could not
be performed (when using
VlinIntegral mode only).
Refer to 6.2 Temperature Errors.
Reanalyze the sample to make
comprehensive judgment, taking sample,
reagent, etc. into account.
Reanalyze the sample to make
comprehensive judgment, taking sample,
reagent, etc. into account. Or conduct
another test with prolonged detection time.
Reanalyze the sample by pre-dilution.
Reanalyze the sample after checking
samples and reagents (storage status blood
collecting method, etc.) for any errors, and
make comprehensive judgment.
Reanalyze the sample after preparing
required amount of plasma.
Reanalyze the sample after checking
samples and reagents (storage status,
blood collecting method, etc.) for any errors,
and make comprehensive judgment.
Check the reagent, lamp, etc. and reanalyze.
Check the reagent, lamp, etc. and reanalyze.
Reanalyze the sample.
Check the setting, reagent and reanalyze.
Reanalyze the sample by re-dilution, etc.
Check the setting, reagent and reanalyze.
6-27
Action
Action
Revised December 2001
8