Negative Pressure Test (Negative Pressure Holding Test) - Fresenius Medical Care 4008 S Service Manual

Hemodialysis device
Hide thumbs Also See for 4008 S:
Table of Contents

Advertisement

Chapter 7: Error messages
Error message
F04 Temperature
F05 Temperature
F06 Temperature
F07 Temperature
F08 Temperature
F09 Temperature
F10 Temperature
F95 Temperature

7.1.12 Negative pressure test (negative pressure holding test)

Error message
F01 neg. Pressure
F02 neg. Pressure
F03 neg. Pressure
7-10
Description
The temperature failed to stabilize within 15 minutes.
– Acknowledgement (T_DIAL1, X633L/B16)  X632/A24 is steadily changing
(change > 0.3 °C/15 sec).
Detuning in positive direction not higher than 3 °C (10 seconds).
– Acknowledgement (T_DIAL1, X633L/B16)  X632/A24 change in voltage
insufficient.
– Detuning (T_DETADJ, X632/A23) X633R/C21 insufficient
The monitor sensor indicates a constant value.
– NTC-3 defective
The test release is missing (max. test running time is 10 minutes).
– Run-time problem (software).
CPU1 failed to transmit a bibag
– Run-time problem (software).
®
bibag
NTC_BIB detuning not higher than 1 °C.
– Acknowledgement (NTC_BIB, X633R/C15)  ADW on P.C.B. LP 633-5,
change in voltage insufficient
– Detuning (BIBAG_TE, X634R/A13)  X633R/A20 insufficient
®
bibag
temperature display outside of measuring range (15 to 45 °C)
– Acknowledgement (NTC_BIB, X633R/C15)  ADW on P.C.B. LP 633-5
System error
Description
During the start phase a negative pressure of more than 450 mmHg has
developed (max. test running time 120 seconds).
– The hydraulic system is contaminated,
– the air separation pump started running
Setting the dialysate pressure to the test pressure (–300 mmHg to –450 mmHg)
was not possible (max. test running time 120 seconds).
Upon repetition of measurement, the range was extended from
–260 mmHg to –490 mmHg.
– Leakage in the hydraulic system
– The UF pump is defective
The working point (116 digits) of the differential amplifier cannot be set correctly
(max. test running time 120 seconds).
– Pressure variations are too large
– The D-A converter (IC11) on P.C.B. LP 632 is defective
– The operational amplifier (IC1/IC3) on P.C.B. LP 632 is defective
– Acknowledgement (P_DIAL, X633L/B6)  X632/A29 is defective
– CI signal is missing (P.C.B. LP 632  X632/B22).
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
®
status message within 3 seconds.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 4008 S and is the answer not in the manual?

Questions and answers

Timorshah
January 25, 2025

Hello I have some problem in fresnuse 4008s can you help me?

Cherief
February 21, 2025

Calibration fuite de sang

1 comments:
Mr. Anderson
February 21, 2025

To calibrate blood leak detection on the Fresenius Medical Care 4008 S:

1. Access Calibration Menu: Use the arrow keys to select the "Flow setting" field.
2. Set Blood-Leak Voltage:
- Press the +/– keys to set the blood-leak voltage to 5.0 V.
- Confirm with the Audio paused key.
3. Set Dimness Voltage:
- Press the +/– keys to set the dimness voltage to 5.0 V.
- Confirm with the Audio paused key.
4. Wait for Confirmation:
- After approximately 3 seconds, the system will display "DATA STORED".
5. Verify Tolerances:
- Blood-leak voltage should be 5 V ± 0.2 V.
- Dimness voltage should be 5 V ± 0.3 V.
6. Check for Contamination:
- If values deviate, inspect the glass burette for contamination.
7. Ensure Proper Conditions:
- Close the housing.
- Maintain a temperature of 37°C.
- Avoid external incident light.

After calibration, ensure proper installation and positioning of the cuvette to maintain accurate readings.

This answer is automatically generated

Rehan
March 9, 2025

When i switched ON machine the alarm” internel error ****” shows on the displyay screen. How to solve the problem

Арсен
March 26, 2025

На 800 потоке поднимает поток до 3000