9.8.3.

#WD8-C Voltage Glanded

Connect NEP-5000 Voltage logger:
Note (1): Activate wiper: Short Grey wire to GND.
Further information, refer to 8.3.2.
Figure 9.8.3.A: Voltage Glanded Logger Configuration
1) Description
The NEP-5000 with voltage driver option outputs turbidity is represented as:
to 2.5V or 0 to 1V. The sensor also allows to control built-in optical wiper using a control wire
or periodic control.
3) Voltage selection
2
To select voltage option, please follow the steps below:
ü Select 'Output Stage' option
ü Select the 'Analog voltage' option
ü Then press, selection
memory.
For further information regarding voltage output, please refer to
via Analog Interface"
Observator Instruments
1
Figure 9.8.3.B: Analog Voltage selection
(9.8.3.B1)
(9.8.3.B2)
(9.8.3.B3)
and
(9.8.3.B4)
of this manual.
Page 70
NEP-5000 MANUAL
Note (1)
-
to save settings permanently to the sensor
section 8.2.4 "Measurement
Copyright 2019
V2019-08-29