Flow Calibration Procedure
5-14 | Thermox WDG-V
Calibration of the flow meter is only required if the sample flow is not vented to
atmosphere. Performing a calibration will create an offset to compensate for any
pressure effect. An accurate external flow measurement is required to perform the
calibration and is not supplied with the analyzer.
Figure 5-17. Flow Configuration Procedure.
1. From the main menu screen, press the ENTER/MENU key to access the
calibration menu.
2. Use the left and/or right navigation arrows to locate the FLOW SENSOR
CALIBRATION menu and press the ENTER/MENU key.
3. Enter the currently measured flow from an external device using the keypad
and select the appropriate units.
4. When you have completed this screen, navigate to the ACCEPT button and
press the ENTER/MENU key to continue.
5. Flow rate will now be compensated.
Flow calibration is valid only if pressure remains consistent. Any sig-
nificant change in sample pressure will require recalibration.
NOTE