Fluke Calibration RUSKA 7615 User Manual page 64

Hydraulic pressure controller/calibrator
Table of Contents

Advertisement

RUSKA 7615
Users Manual
Value
-313
-315
-330
-350
-400
-500
-501
-502
-503
-521
-546
-600
-601
-701
-702
-703
-707
-800
B-2
Table B-2. Error Messages, continued.
Description and Corrective Action
Configuration Data Lost. The configuration data has been lost and the unit must be
recalibrated.
Configuration Data Lost. The configuration data has been lost. Check all parameters
to be sure they are correct.
Self-Test Failed. Check the display for the test that failed.
Queue Overflow. The error queue was full and messages were lost.
Query Error. A read request was received when there was nothing to read.
Controller Malfunction. Internal control failure.
High Limit Exceeded. The pressure was greater than the high limit.
Low Limit Exceeded. The pressure was less than the low limit.
Slew Limit Exceeded. The pressure changed faster than the slew limit allowed.
Pressure Over-range. The pressure reading is outside the range of the DPC.
Valve Error (Isolation Valve)
Factory Data Lost. Internal factory constants have been lost. Contact Fluke for more
information.
Calibration Mode. The CAL button must be pressed before SCPI calibration
commands can be executed.
Pump Pressure Out of Range
Pump Position Out of Range
Reservoir Level Low
Pressure Too High (To Switch Between Sensor Ranges)
Solenoid Over-Temperature (Control Valves)