Page 1
Thank you for your purchase. This manual contains: set-up instructions, unit specs, safety information, controller operation and maintenance steps. K205ULT IMPORTANT: Your unit is preprogrammed Place your unit in the desired location. Plug the unit in and allow it to cool and become stable for a minimum of 24 hours before logging temperature or stocking products.
K205ULT Operating Manual CONTENTS Warnings Product Description Installation and Operation Maintenance Temperature Controller Temperature Logging Parameters Warnings Important operating and/or Hot surface(s) present which may maintenance instructions. Read cause burns to unprotected skin the accompanying text carefully. or to materials which may be damaged by elevate temperatures.
K205ULT Operating Manual Installation and Operation Ventilation Assembly • Remove the appliance from the packaging • All models must have clear and unobstructed and peel off any protective film from all ventilation from the entire surface area of the surfaces.
K205ULT Operating Manual Indicator light status description Symbol Status Meaning Parameter setting Status of temperature measuring and controlling High temperature compressor work High temperature compressor stop FLASH High temperature compresso time delay Low-temperature compressor work in non-forced refrigeration mode Low temperature compressor is closed in non-forced refrigeration mode...
Page 5
K205ULT Operating Manual Setting the Temperature • Press Set key for 3 seconds. it display the code St. • Press Set key again to display the temperature SETPOINT. Modify by pressing the UP or DOWN keys. Changing Parameters - Enter Admin Section •...
Service K2 Scientific want to make sure you are happy with your purchase. There are several ways for you to contact us with questions or service needs. Be sure to include your four digit order number or at least your model number handy to speed up the process.
Page 9
Parameter Description Setting scope Default User menu Set cabinet temperature point LS to US Administrator menu Administrator menu password (The password is 55 and cannot be changed) Cabinet temp. probe calibration when below 0° -15.0~15.0 Use this parameter to match temp display with datalogger Condenser probe calibration (High-temp.
Need help?
Do you have a question about the K205ULT and is the answer not in the manual?
Questions and answers
Alarm keeps alarming and the code is reading EL1