CSA STD C22.2 No. 60601-1 NOTE: The latest edition of this operating manual can be downloaded from our website at www.maxtec.com This manual describes the function, operation and maintenance of the Maxtec Model MaxO2+ A Never allow an excess length of tubing, lanyard or sensor cable near the patient’s head or ◆...
The Maxtec Max-250 oxygen sensor is a sealed device containing a mild acid electrolyte, Oxygen-specific, galvanic sensor that achieves 90% of final value in approximately 15 • ◆ lead (Pb), and lead acetate. Lead and lead acetate are hazardous waste constituents and seconds at room temperature.
A simple calibration may be made with the sensor open to static ambient air. For optimum After the unit is turned on it will automatically calibrate to room air. The display should be accuracy Maxtec recommends that the Sensor be placed in a closed loop circuit where gas flow stable and reading 20.9%.
If these steps do not correct the error, contact Maxtec for technical support. calibrating as when measuring. NOTE: Use only a Maxtec approved Max-550E sensor called out in Section 9.0 of the Spare Parts 3.4 Humidity Effects List. The Max550E sensor is equipped with an authentication chip to ensure the monitor is used Humidity (non-condensing) has no effect on the performance of the MaxO2+ other than dilut- with an approved sensor.
Clean the sensor with a cloth moistened with isopropyl alcohol (65% alcohol/water ◆ solution). Maxtec does not recommend use of spray disinfectants because they can contain salts, ◆ which can accumulate in the sensor membrane and impair readings. The oxygen sensor is not intended for steam, ethylene oxide or radiation sterilization.
R217P23) frequency communication services. The user might need to take mitigation measures, such PART NUMBER ITEM as relocating or re-orienting the equipment. R206P75 Pole Mount R205P86 Wall Mount R100P10 Rail Mount R206P76 Horizontal Pole Mount WWW.MAXTEC.COM • (800) 748-5355 ENGLISH...
Page 8
Dip >95%, 5 seconds interference. If user requires continued operation during power mains interruptions, ensure that batteries are installed and charged. Ensure that battery life exceeds longest anticipated power outages or provide an additional uninterruptible power source. ENGLISH WWW.MAXTEC.COM • (800) 748-5355...