CF error code is for Calibrator (CDI 540)
Error code
Description
(Recoverable)
CF01
Detected at
calibration start up.
CF02
Detected at
calibration start up.
CF03
Detected at
calibration start up
CF04
Detected at
calibration start up.
CF05
Detected at
calibration start up.
CF06
Detected at
calibration start up.
CF07
Detected at
calibration start up.
CF08
Detected at
calibration start up.
CF09
Detected at
calibration start up.
CF0A
Detected at any
time during
calibration.
CF0B
Detected at
calibration start up.
CF0C
Detected at any
time during
calibration.
CF11
Detected at
calibration start up.
Cause
Calibrator I2C failure.
Calibrator ADC failure.
Calibrator reference voltage
error.
Bottle A valve failure or
Calibrator leakage or bad
transducer for bottle A.
Bottle B valve failure or
Calibrator leakage or bad
transducer for bottle B.
Bad exhaust valve or mix-
line transducer
Mix line transducer failure
or bottle A, B valve failure.
Bad bottle A pressure
regulator or bad mix line
transducer.
Bottle A valve failure, bottle
A pressure regulator failure,
large leak.
Bad exhaust valve.
Bad bottle B pressure
regulator or bad mix line
transducer.
Bad calibrator memory or
un-calibrated calibrator.
Bottle B valve failure, bottle
B pressure regulator failure,
large leak.
Recommendation
Exit calibration and contact Terumo
CVS Technical Service.
Exit calibration and contact Terumo
CVS Technical Service.
Exit calibration and contact Terumo
CVS Technical Service.
Exit calibration and contact Terumo
CVS Technical Service.
Exit calibration and contact Terumo
CVS Technical Service.
Exit calibration and contact Terumo
CVS Technical Service.
Exit calibration and contact Terumo
CVS Technical Service.
Exit calibration and contact Terumo
CVS Technical Service.
Exit calibration and contact Terumo
CVS Technical Service.
Exit calibration and contact Terumo
CVS Technical Service.
Exit calibration and contact Terumo
CVS Technical Service.
Restart calibration process and see
if the error code persists. If the
error code persists, exit calibration
and contact Terumo CVS Technical
Service.
Exit calibration and contact Terumo
CVS Technical Service.
troubleshooting