The design of Xstrahl Systems is in compliance with internationally Compliance recognised standards for safety. All Xstrahl products have received CE marking approval for sale in Europe, clearance by FDA for sale in the U.S.A., are licensed for sale in Canada and are designed and manufactured in accordance with an ISO13485:2016 certified quality management system.
Not all documents referred to in this document are part of the scope Referenced of delivery for the equipment. Xstrahl reserves the right to determine Documents the documents delivered with the product. Xstrahl X-Ray Therapy Systems must be used only in combination...
Ensure the shipping box for the system and generator are stored Note: upright and the boxes are not stacked at any time. The Xstrahl Radiotherapy systems are suitable for use in a clinical or Electromagnetic hospital environment, within a suitable lead shielded treatment room Compatibility (for relevant X-Ray radiation protection).
All operators must have the required training before attempting to Training operate the Xstrahl X-Ray Therapy System. Because countries have different regulations for training, the operator must be compliant with the local laws and regulations of the jurisdiction in which the equipment is installed.
The Xstrahl 150 defines the exposure in time and does not measure Quality Assurance the output of the X-Ray tube for each exposure conducted. The output of each clinical energy needs to be measured during the acceptance of the unit and schedule in the daily quality assurance conducted on the machine.
Xstrahl® (100, 150, 200 and 300) X-Ray Therapy Systems are Intended Use (of intended to be used for radiation treatment of superficial skin equipment) disorders, boney metastases and diseases of the skin, as determined by a licensed medical practitioner where the system is being used.
Preface This is the Operator Manual for the Xstrahl 150 X-Ray Therapy System. This manual provides the information a medical operator requires to operate the Xstrahl 150. Precautionary Information This section provides an overview of important safety information and cautionary warnings which should be read and thoroughly understood prior to operating the Xstrahl 150.
______________________________________________________ System Description The Xstrahl 150 is a superficial X-Ray therapy system producing X-Rays up to 150kV. The energy of the beam is defined as the half value layer, which is dependent on the kV selected and the filter materials placed within the X-Ray beam.
More information can be found in the Concerto section in this manual. Fisica Fisica is database-driven physics software used to calibrate the system. More information on Fisica can be found in the Xstrahl Technical Manual Note: provided with the equipment.
Each Xstrahl 150 system can have up to ten filters—nine clinical filters and one warm-up filter. The system uses encoding to detect treatment filters within the filter storage unit. The clinical filters can be constructed in accordance with the half value layers defined by the medical department.
CAUTION: Care should be exercised at all times when handling the applicators so as not to damage the applicator. Figure 2-4: Applicator Illustration Standard range of applicators for the Xstrahl 150: Xstrahl 150 Applicators 15 cm FSD Open Applicators 25 cm FSD Open Applicators 1.5 cm diameter...
2.2.2.3. Cleaning the Applicators WARNING: Contact with alcohol can damage the applicators. Xstrahl recommends that only products which are free from alcohol are used to clean the applicators. Due to the ends of the treatment applicators being constructed of Lucite®...
Xstrahl 150 Tube Stand Figure 2-6: Floor/Wall Mounted Xstrahl 150 Illustration The Xstrahl 150 can be supplied with a floor/wall mounted tube stand or on a ceiling track. Electromagnetic brakes lock the tube stand movements. These brakes can be released individually or in combination, by pushing the buttons located on the front panel of the X-Ray tube assembly.
3.1. Power On If the Xstrahl System has been installed with a connection to a mains isolator, the isolator needs to be placed in the on position (this is either a key operation or a mechanical switch). The location of the mains isolator will vary between sites and is established during the installation process.
After the system is powered on, a warm-up should be conducted. To conduct a warm-up: 1. Fit the warm-up filter and a small aperture applicator on the Xstrahl system. The Xstrahl system requires an applicator to be fitted to the sub-tube assembly, but does not require a specific applicator.
3.4. Power Down Procedure The Xstrahl X-Ray Therapy System can be left powered on with the operator pod key switch in Standby (position 2). During the clinical session this leaves the system with mains power. The key should be removed if the machine is left unattended;...
Page 32
Turn the key switch on the operator pod to O (OFF) to shut down mains power from the controller. The Xstrahl system has an internal clock which records the time elapsed since the last exposure was conducted. The cooling system power will be Note: maintained until the shutdown delay value has elapsed.
® Concerto ® Concerto is the clinical interface for the Xstrahl system. Concerto enables the operator to create patients, define treatment parameters and deliver exposures, record values to a database and print hard copy patient reports. The Concerto interface can be used without the X-Ray system being powered on, which can be useful for data housekeeping.
Section 5: System Errors System Errors Xstrahl system errors are displayed on the bottom right side of the exposure screens in Concerto and Fisica. Types of Errors Will persist whilst the problem exists; for example, a Normal Errors filter encoding error will remain until the correct filter is...
Need help?
Do you have a question about the 150 and is the answer not in the manual?
Questions and answers